An evaluation of techniques for dose calculation on cone beam computed tomography.

OBJECTIVE: To assess the accuracy and efficiency of four different techniques, thus determining the optimum method for recalculating dose on cone beam CT (CBCT) images acquired during radiotherapy treatments. METHODS: Four established techniques were investigated and their accuracy assessed via dose calculations: (1) applying a standard planning CT (pCT) calibration curve, (2) applying a CBCT site-specific calibration curve, (3) performing a density override and (4) using deformable registration. Each technique was applied to 15 patients receiving volumetric modulated arc therapy to one of three treatment sites, head and neck, lung and prostate. Differences between pCT and CBCT recalculations were determined with dose volume histogram metrics and 2.0%/0.1 mm gamma analysis using the pCT dose distribution as a reference. RESULTS: Dose volume histogram analysis indicated that all techniques yielded differences from expected results between 0.0 and 2.3% for both target volumes and organs at risk. With volumetric gamma analysis, the dose recalculation on deformed images yielded the highest pass-rates. The median pass-rate ranges at 50% threshold were 99.6-99.9%, 94.6-96.0%, and 94.8.0-96.0% for prostate, head and neck and lung patients, respectively. CONCLUSION: Deformable registration, HU override and site-specific calibration curves were all identified as dosimetrically accurate and efficient methods for dose calculation on CBCT images. ADVANCES IN KNOWLEDGE: With the increasing adoption of CBCT, this study provides clinical radiotherapy departments with invaluable information regarding the comparison of dose reconstruction methods, enabling a more accurate representation of a patient's treatment. It can also integrate studies in which CBCT is used in image-guided radiation therapy and for adaptive radiotherapy planning processes.

[1]  Cai Grau,et al.  Open source deformable image registration system for treatment planning and recurrence CT scans , 2016, Strahlentherapie und Onkologie.

[2]  Peter B Greer,et al.  Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm. , 2015, Medical physics.

[3]  D A Jaffray,et al.  A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets. , 1999, International journal of radiation oncology, biology, physics.

[4]  Ying Xiao,et al.  Investigation into image quality and dose for different patient geometries with multiple cone-beam CT systems. , 2014, Medical physics.

[5]  Markus Stock,et al.  Feasibility of CBCT-based dose calculation: comparative analysis of HU adjustment techniques. , 2012, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[6]  Philippe Lambin,et al.  3D in vivo dosimetry using megavoltage cone-beam CT and EPID dosimetry. , 2009, International journal of radiation oncology, biology, physics.

[7]  K. Brock Results of a multi-institution deformable registration accuracy study (MIDRAS). , 2010, International journal of radiation oncology, biology, physics.

[8]  Craig Cornwall,et al.  Parotid glands dose-effect relationships based on their actually delivered doses: implications for adaptive replanning in radiation therapy of head-and-neck cancer. , 2013, International journal of radiation oncology, biology, physics.

[9]  Uwe Oelfke,et al.  Comparison of CT number calibration techniques for CBCT-based dose calculation , 2015, Strahlentherapie und Onkologie.

[10]  M. Kessler Image registration and data fusion in radiation therapy. , 2006, The British journal of radiology.

[11]  Boyd McCurdy,et al.  Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy , 2009, Physics in medicine and biology.

[12]  Robert J W Louwe,et al.  Validation of Varian’s SmartAdapt® deformable image registration algorithm for clinical application , 2015, Radiation Oncology.

[13]  M. V. van Herk,et al.  Automatic in vivo portal dosimetry of all treatments , 2013, Physics in medicine and biology.

[14]  E. Mok,et al.  Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files , 2010, Physics in medicine and biology.

[15]  P. Greer,et al.  Does the planning dose-volume histogram represent treatment doses in image-guided prostate radiation therapy? Assessment with cone-beam computerised tomography scans. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[16]  Mark A. Edwards,et al.  Simplified material assignment for cone beam computed tomography-based dose calculations of prostate radiotherapy with hip prostheses , 2016 .

[17]  Benoît Macq,et al.  Comparison of 12 deformable registration strategies in adaptive radiation therapy for the treatment of head and neck tumors. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  P. Lambin,et al.  Evaluation of nonrigid registration models for interfraction dose accumulation in radiotherapy. , 2009, Medical physics.

[19]  B Haas,et al.  Automatic segmentation of thoracic and pelvic CT images for radiotherapy planning using implicit anatomic knowledge and organ-specific segmentation strategies , 2008, Physics in medicine and biology.

[20]  A J Cole,et al.  Motion management for radical radiotherapy in non-small cell lung cancer. , 2014, Clinical oncology (Royal College of Radiologists (Great Britain)).

[21]  D. Hallahan,et al.  A study on adaptive IMRT treatment planning using kV cone-beam CT. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[22]  N. Kadoya,et al.  Evaluation of on-board kV cone beam computed tomography-based dose calculation with deformable image registration using Hounsfield unit modifications. , 2014, International journal of radiation oncology, biology, physics.

[23]  Fang-Fang Yin,et al.  Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning. , 2006, International journal of radiation oncology, biology, physics.

[24]  W. D'Souza,et al.  Feasibility of CBCT‐based dose with a patient‐specific stepwise HU‐to‐density curve to determine time of replanning , 2017, Journal of applied clinical medical physics.

[25]  Ben Mijnheer,et al.  In vivo dosimetry in external beam radiotherapy. , 2013, Medical physics.

[26]  S. Cilla,et al.  Quasi real time in vivo dosimetry for VMAT. , 2014, Medical physics.

[27]  W. Tomé,et al.  Dose calculation on kV cone beam CT images: an investigation of the Hu-density conversion stability and dose accuracy using the site-specific calibration. , 2010, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[28]  Sébastien Ourselin,et al.  Toward adaptive radiotherapy for head and neck patients: Feasibility study on using CT-to-CBCT deformable registration for "dose of the day" calculations. , 2014, Medical physics.

[29]  Ben Heijmen,et al.  Correction of conebeam CT values using a planning CT for derivation of the "dose of the day". , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[30]  J. Wong,et al.  Flat-panel cone-beam computed tomography for image-guided radiation therapy. , 2002, International journal of radiation oncology, biology, physics.

[31]  Marcel van Herk,et al.  Online 3D EPID-based dose verification: Proof of concept. , 2016, Medical physics.

[32]  J Fairfoul,et al.  Image-guided radiotherapy of the prostate using daily CBCT: the feasibility and likely benefit of implementing a margin reduction. , 2014, The British journal of radiology.

[33]  Michael Velec,et al.  Accumulated dose in liver stereotactic body radiotherapy: positioning, breathing, and deformation effects. , 2012, International journal of radiation oncology, biology, physics.

[34]  Turki Almatani,et al.  Automated algorithm for CBCT-based dose calculations of prostate radiotherapy with bilateral hip prostheses. , 2016, The British journal of radiology.

[35]  Ergun Ahunbay,et al.  Automated registration of large deformations for adaptive radiation therapy of prostate cancer. , 2009 .

[36]  F. Verhaegen,et al.  Accuracy of dose calculations on kV cone beam CT images of lung cancer patients. , 2016, Medical physics.

[37]  Kavitha Srinivasan,et al.  Cone Beam Computed Tomography for Adaptive Radiotherapy Treatment Planning , 2014 .

[38]  Matthias Guckenberger,et al.  Investigation of the usability of conebeam CT data sets for dose calculation , 2008, Radiation oncology.

[39]  M van Herk,et al.  Catching errors with in vivo EPID dosimetry. , 2010, Medical physics.

[40]  T. Marchant,et al.  Shading correction algorithm for improvement of cone-beam CT images in radiotherapy , 2008, Physics in medicine and biology.