Intrasubject reproducibility with repeated determination of vertebral mineral density from a given set of CT images was investigated. The region of interest (ROI) in 10 patient scans was selected by four independent operators either manually or with an automated procedure separating cortical and spongeous bone, the operators being requested to interact in ROI selection. The mean intrasubject variation was found to be much lower with the automated process (0.3 to 0.6%) than with the conventional method (2.5 to 5.2%). In a second study, 10 patients were examined twice to determine the reproducibility of CT slice selection by the operator. The errors were of the same order of magnitude as in ROI selection.