Quantitative T1 and proton density mapping with direct calculation of radiofrequency coil transmit and receive profiles from two‐point variable flip angle data

Quantitative T1 mapping of brain tissue is frequently based on the variable flip angle (VFA) method, acquiring spoiled gradient echo (GE) datasets at different excitation angles. However, accurate T1 calculation requires a knowledge of the sensitivity profile B1 of the radiofrequency (RF) transmit coil. For an additional derivation of proton density (PD) maps, the receive coil sensitivity profile (RP) must also be known. Mapping of B1 and RP increases the experiment duration, which may be critical when investigating patients. In this work, a method is presented for the direct calculation of B1 and RP from VFA data. Thus, quantitative maps of T1, PD, B1 and RP can be obtained from only two spoiled GE datasets. The method is based on: (1) the exploitation of the linear relationship between 1/PD and 1/T1 in brain tissue and (2) the assumption of smoothly varying B1 and RP, so that a large number of data points can be fitted across small volume elements where B1 and RP are approximately constant. The method is tested and optimized on healthy subjects. Copyright © 2016 John Wiley & Sons, Ltd.

[1]  M Furuse,et al.  Spin-lattice relaxation times of bound water--its determination and implications for tissue discrimination. , 1987, Magnetic resonance imaging.

[2]  S. Riederer,et al.  Optimizing the precision in T1 relaxation estimation using limited flip angles , 1987, Magnetic resonance in medicine.

[3]  P P Fatouros,et al.  In Vivo Brain Water Determination by T1 Measurements: Effect of Total Water Content, Hydration Fraction, and Field Strength , 1991, Magnetic resonance in medicine.

[4]  F. Franconi,et al.  Radiofrequency map of an NMR coil by imaging. , 1993, Magnetic resonance imaging.

[5]  E M Haacke,et al.  Accurate determination of spin‐density and T1 in the presence of RF‐field inhomogeneities and flip‐angle miscalibration , 1998, Magnetic resonance in medicine.

[6]  A. Marmarou,et al.  Use of magnetic resonance imaging for in vivo measurements of water content in human brain: method and normal values. , 1999, Journal of neurosurgery.

[7]  S. Holland,et al.  NMR relaxation times in the human brain at 3.0 tesla , 1999, Journal of magnetic resonance imaging : JMRI.

[8]  R. Brooks,et al.  T1 and T2 in the brain of healthy subjects, patients with Parkinson disease, and patients with multiple system atrophy: relation to iron content. , 1999, Radiology.

[9]  G. Barker,et al.  Variations in T1 and T2 relaxation times of normal appearing white matter and lesions in multiple sclerosis , 2000, Journal of the Neurological Sciences.

[10]  P M Jakob,et al.  Rapid quantitative lung 1H T1 mapping , 2001, Journal of magnetic resonance imaging : JMRI.

[11]  P. Tofts Quantitative MRI of the Brain , 2003 .

[12]  M. Maier Quantitative MRI of the brain—measuring changes caused by disease , 2004 .

[13]  C. Laule,et al.  Water content and myelin water fraction in multiple sclerosis , 2004, Journal of Neurology.

[14]  J. Pauly,et al.  Saturated double‐angle method for rapid B1+ mapping , 2006, Magnetic resonance in medicine.

[15]  Terry M Peters,et al.  Synthetic T1-weighted brain image generation with incorporated coil intensity correction using DESPOT1. , 2006, Magnetic resonance imaging.

[16]  Nadim Joni Shah,et al.  A new method for fast quantitative mapping of absolute water content in vivo , 2006, NeuroImage.

[17]  Vasily L Yarnykh,et al.  Actual flip‐angle imaging in the pulsed steady state: A method for rapid three‐dimensional mapping of the transmitted radiofrequency field , 2007, Magnetic resonance in medicine.

[18]  S. Deoni,et al.  High‐resolution T1 mapping of the brain at 3T with driven equilibrium single pulse observation of T1 with high‐speed incorporation of RF field inhomogeneities (DESPOT1‐HIFI) , 2007, Journal of magnetic resonance imaging : JMRI.

[19]  Nikolaus Weiskopf,et al.  Rapid Radiofrequency Field Mapping In Vivo Using Single-Shot STEAM MRI , 2008, Magnetic resonance in medicine.

[20]  Glen R Morrell,et al.  A phase‐sensitive method of flip angle mapping , 2008, Magnetic resonance in medicine.

[21]  N.J. Shah,et al.  Quantitative cerebral water content mapping in hepatic encephalopathy , 2008, NeuroImage.

[22]  R. Deichmann,et al.  T1 mapping using spoiled FLASH‐EPI hybrid sequences and varying flip angles , 2009, Magnetic resonance in medicine.

[23]  R. Deichmann,et al.  Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles , 2009, Magnetic resonance in medicine.

[24]  Ralf Deichmann,et al.  A fast B1-mapping method for the correction and normalization of magnetization transfer ratio maps at 3 T , 2010, NeuroImage.

[25]  Tobias Kober,et al.  MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field , 2010, NeuroImage.

[26]  Nikolaus Weiskopf,et al.  Unified segmentation based correction of R1 brain maps for RF transmit field inhomogeneities (UNICORT) , 2011, NeuroImage.

[27]  Peter Börnert,et al.  DREAM—a novel approach for robust, ultrafast, multislice B1 mapping , 2012, Magnetic resonance in medicine.

[28]  Reto Meuli,et al.  MP2RAGE Multiple Sclerosis Magnetic Resonance Imaging at 3 T , 2012, Investigative radiology.

[29]  R. Deichmann,et al.  Correction of systematic errors in quantitative proton density mapping , 2012, Magnetic resonance in medicine.

[30]  Ralf Deichmann,et al.  Quantitative proton density mapping: correcting the receiver sensitivity bias via pseudo proton densities , 2012, NeuroImage.

[31]  B. Hargreaves,et al.  Simultaneous T1 and B1+ Mapping Using Reference Region Variable Flip Angle Imaging , 2013, Magnetic resonance in medicine.

[32]  Ralf Deichmann,et al.  Improved visibility of brain tumors in synthetic MP‐RAGE anatomies with pure T1 weighting , 2015, NMR in biomedicine.