A simple fully integrated contact-coupled wear prediction for ultra-high molecular weight polyethylene hip implants

Abstract A fully coupled contact and wear model was developed in the present study for hip implants employing an ultra-high molecular weight polyethylene (UHMWPE) cup in combination with a metallic or ceramic femoral head. A simple elasticity equation based on the concept of constrained column model was employed to solve the contact mechanics between the acetabular cup and the femoral head under the three-dimensional physiological loading condition. The wear model was based on the classical Archard—Lancaster equation in common with all other studies reported in the literature. The fully coupled contact and wear model was applied to both conventional and cross-linked UHMWPE cups under a wide range of design parameters such as the clearance and the femoral head radius. The predicted linear and volumetric wear as well as their rates for conventional UHMWPE cups were found to be in good agreement with those obtained from a similar analysis by Maxian but using the finite element method for the contact mechanics analysis. The predicted maximum contact pressure was found to decrease rapidly within the first 106 cycles, and below the limit to cause plastic deformation within the UHMWPE cup with a nominal radial clearance of 0.2 mm. The effect of the clearance between the head and the cup on the predicted wear was found to be negligible. For the cross-linked UHMWPE cup with relatively large diameters up to 48 mm and a fixed outside diameter of 50 mm, the predicted wear, which was found to increase with increasing femoral head radius, remained small owing to the small wear factor associated with these materials. Furthermore, if the head diameter increases beyond 42 mm, a rapid increase in the contact pressure was predicted, owing to the decrease in the wall thickness of the cross-linked UHMWPE cup.

[1]  R Pietrabissa,et al.  Wear of polyethylene cups in total hip arthroplasty: a parametric mathematical model. , 1998, Medical engineering & physics.

[2]  D Dowson,et al.  Comparative study of the activity of total hip arthroplasty patients and normal subjects. , 2001, The Journal of arthroplasty.

[3]  Olof Calonius,et al.  Slide track analysis of the relative motion between femoral head and acetabular cup in walking and in hip simulators. , 2002, Journal of biomechanics.

[4]  D Dowson,et al.  A full numerical analysis of hydrodynamic lubrication in artificial hip joint replacements constructed from hard materials , 1999 .

[5]  Julia C. Shelton,et al.  A hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions , 2001 .

[6]  James Shih-Shyn Wu,et al.  The computer simulation of wear behavior appearing in total hip prosthesis , 2003, Comput. Methods Programs Biomed..

[7]  G. Smidt,et al.  Measurement of hip-joint motion during walking. Evaluation of an electrogoniometric method. , 1969, The Journal of bone and joint surgery. American volume.

[8]  Mauricio Silva,et al.  Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity. , 2002, The Journal of arthroplasty.

[9]  T D Brown,et al.  The Frank Stinchfield Award. 3-Dimensional sliding/contact computational simulation of total hip wear. , 1996, Clinical orthopaedics and related research.

[10]  John J Callaghan,et al.  Local head roughening as a factor contributing to variability of total hip wear: a finite element analysis. , 2002, Journal of biomechanical engineering.

[11]  John J. Craig,et al.  Introduction to Robotics Mechanics and Control , 1986 .

[12]  G Lewis,et al.  Properties of crosslinked ultra-high-molecular-weight polyethylene. , 2001, Biomaterials.

[13]  R Pietrabissa,et al.  Improved mathematical model of the wear of the cup articular surface in hip joint prostheses and comparison with retrieved components , 2001, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[14]  W. Harris,et al.  Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. , 2001, The Journal of arthroplasty.

[15]  G. Smidt,et al.  Measurement of hip joint motion during walking , 1969 .

[16]  R. Brand,et al.  Comparison of hip force calculations and measurements in the same patient. , 1994, The Journal of arthroplasty.

[17]  T D Brown,et al.  Finite element analysis of acetabular wear. Validation, and backing and fixation effects. , 1997, Clinical orthopaedics and related research.

[18]  D Dowson,et al.  The walking activity of patients with artificial hip joints. , 1982, Engineering in medicine.

[19]  Z. Jin,et al.  Experimental and Theoretical Study of the Contact Mechanics of Five Total Knee Joint Replacements , 1995, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[20]  S H Teoh,et al.  An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. , 2002, Journal of biomechanics.

[21]  A Unsworth,et al.  Is the wear factor in total joint replacements dependent on the nominal contact stress in ultra-high molecular weight polyethylene contacts? , 2004, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[22]  T D Brown,et al.  Temporal and spatial distributions of directional counterface motion at the acetabular bearing surface in total hip arthroplasty. , 1998, The Iowa orthopaedic journal.

[23]  J Fisher,et al.  Microscopic asperity contact and deformation of ultrahigh molecular weight polyethylene bearing surfaces , 2003, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[24]  Olof Calonius,et al.  Force track analysis of contemporary hip simulators. , 2003, Journal of biomechanics.

[25]  Aiguo Wang,et al.  Quantification of the effect of cross-path motion on the wear rate of ultra-high molecular weight polyethylene , 2003 .

[26]  C. V. White,et al.  Backside nonconformity and locking restraints affect liner/shell load transfer mechanisms and relative motion in modular acetabular components for total hip replacement. , 1998, Journal of biomechanics.

[27]  J. Fisher,et al.  Biological reactions to wear debris in total joint replacement , 2000, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[28]  T. Brown,et al.  A sliding-distance-coupled finite element formulation for polyethylene wear in total hip arthroplasty. , 1996, Journal of biomechanics.

[29]  J Fisher,et al.  Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. , 2000, The Journal of arthroplasty.

[30]  S. Kurtz,et al.  An augmented hybrid constitutive model for simulation of unloading and cyclic loading behavior of conventional and highly crosslinked UHMWPE. , 2004, Biomaterials.

[31]  Jaakko Keränen,et al.  Effect of slide track shape on the wear of ultra-high molecular weight polyethylene in a pin-on-disk wear simulation of total hip prosthesis. , 2004, Journal of biomedical materials research. Part B, Applied biomaterials.

[32]  D D Auger,et al.  An axisymmetric contact model of ultra high molecular weight polyethylene cups against metallic femoral heads for artificial hip joint replacements , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[33]  V. Saikko,et al.  Effect of counterface roughness on the wear of conventional and crosslinked ultrahigh molecular weight polyethylene studied with a multi-directional motion pin-on-disk device. , 2001, Journal of biomedical materials research.

[34]  P S Walker,et al.  The effect of contact area on wear in relation to fixed bearing and mobile bearing knee replacements. , 2001, Journal of biomedical materials research.

[35]  A. Wang,et al.  Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement , 2001, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[36]  D L Bartel,et al.  The effect of conformity and plastic thickness on contact stresses in metal-backed plastic implants. , 1985, Journal of biomechanical engineering.

[37]  John Fisher,et al.  The influence of contact stress on the wear of UHMWPE for total replacement hip prostheses , 1995 .