A comparison of the conformability of everolimus-eluting bioresorbable vascular scaffolds to metal platform coronary stents.

OBJECTIVES The aim of this study was to assess the differences in terms of curvature and angulation of the treated vessel after the deployment of either a metallic stent or a polymeric scaffold device. BACKGROUND Conformability of metallic platform stents (MPS) is the major determinant of geometric changes in coronary arteries caused by the stent deployment. It is not known how bioresorbable polymeric devices perform in this setting. METHODS This retrospective study compares 102 patients who received an MPS (Multi-link Vision or Xience V, Abbott Vascular, Santa Clara, California) in the SPIRIT FIRST and II trials with 89 patients treated with the Revision 1.1 everolimus-eluting bioresorbable vascular scaffold (BVS) (Abbott Vascular, Santa Clara, California) from cohort B of the ABSORB (A bioabsorbable everolimus-eluting coronary stent system) trial. All patients were treated with a single 3 × 18 mm device. Curvature and angulation were measured with dedicated software by angiography. RESULTS Both the MPS and BVS groups had significant changes in relative region curvature (MPS vs. BVS: 28.7% vs. 7.5%) and angulation (MPS vs. BVS: 25.4% vs. 13.4%) after deployment. The unadjusted comparisons between the 2 groups showed for BVS a nonsignificant trend for less change in region curvature after deployment (MPS vs. BVS: 0.085 cm(-1) vs. 0.056 cm(-1), p = 0.06) and a significantly lower modification of angulation (MPS vs. BVS 6.4° vs. 4.3°, p = 0.03). By multivariate regression analysis, the independent predictors of changes in curvature and angulation were the pre-treatment region curvature, the pre-treatment region angulation, and the used device. CONCLUSIONS Bioresorbable vascular scaffolds have better conformability than conventional MPS. The clinical significance of the observed differences will require further investigation.

[1]  P. Serruys,et al.  A randomised comparison of an everolimus-eluting coronary stent with a paclitaxel-eluting coronary stent:the SPIRIT II trial. , 2006, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[2]  L. Topoleski,et al.  A comparison of the mechanical performance characteristics of seven drug‐eluting stent systems , 2009, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[3]  Patrick W Serruys,et al.  A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial , 2008, The Lancet.

[4]  J. Messenger,et al.  3D coronary reconstruction from routine single-plane coronary angiograms: Clinical validation and quantitative analysis of the right coronary artery in 100 patients , 2000, The International Journal of Cardiac Imaging.

[5]  P. Serruys,et al.  Methodological and clinical implications of the relocation of the minimal luminal diameter after intracoronary radiation therapy. Dose Finding Study Group. , 2000, Journal of the American College of Cardiology.

[6]  P. Serruys,et al.  Usefulness of shear stress pattern in predicting neointima distribution in sirolimus-eluting stents in coronary arteries. , 2003, The American journal of cardiology.

[7]  P. Serruys,et al.  Three-year results of clinical follow-up after a bioresorbable everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB trial. , 2010, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[8]  J. Rumberger,et al.  Coronary artery calcium area by electron-beam computed tomography and coronary atherosclerotic plaque area. A histopathologic correlative study. , 1995, Circulation.

[9]  S. Allaqaband,et al.  Stent fracture, an incidental finding or a significant marker of clinical in‐stent restenosis? , 2008, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[10]  E. Topol,et al.  Results of percutaneous transluminal coronary angioplasty of high-risk angulated stenoses. , 1990, The American journal of cardiology.

[11]  P. Serruys,et al.  Quantitative multi-modality imaging analysis of a bioabsorbable poly-L-lactic acid stent design in the acute phase: a comparison between 2- and 3D-QCA, QCU and QMSCT-CA. , 2008, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[12]  Antonio Colombo,et al.  Selection of coronary stents. , 2002, Journal of the American College of Cardiology.

[13]  Nathan M. Wilson,et al.  Methods for Quantifying Three-Dimensional Deformation of Arteries due to Pulsatile and Nonpulsatile Forces: Implications for the Design of Stents and Stent Grafts , 2008, Annals of Biomedical Engineering.

[14]  T. Ismail,et al.  Restenosis and stent fracture following sirolimus-eluting stent (SES) implantation. , 2007, Circulation journal : official journal of the Japanese Circulation Society.

[15]  Dong-Gu Shin,et al.  Frequency of stent fracture as a cause of coronary restenosis after sirolimus-eluting stent implantation. , 2007, The American journal of cardiology.

[16]  T. Akasaka,et al.  Predictors and prognosis of stent fracture after sirolimus-eluting stent implantation. , 2009, Circulation journal : official journal of the Japanese Circulation Society.

[17]  D Glogar,et al.  Longitudinal straightening effect of stents is an additional predictor for major adverse cardiac events. Austrian Wiktor Stent Study Group and European Paragon Stent Investigators. , 2000, Journal of the American College of Cardiology.

[18]  朝倉 利久,et al.  Flow patterns and spatial distribution of atherosclerotic lesions in human coronary arteries , 1989 .

[19]  J J Wentzel,et al.  Coronary stent implantation changes 3-D vessel geometry and 3-D shear stress distribution. , 2000, Journal of biomechanics.

[20]  Régis Rieu,et al.  Assessment of the trackability, flexibility, and conformability of coronary stents: A comparative analysis , 2003, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[21]  J. Ormiston,et al.  Stent longitudinal flexibility: A comparison of 13 stent designs before and after balloon expansion , 2000, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[22]  D. Baim,et al.  Reliability of the quantitative angiographic measurements in the New Approaches to Coronary Intervention (NACI) registry: a comparison of clinical site and repeated angiographic core laboratory readings. , 1997, The American journal of cardiology.

[23]  J J Wentzel,et al.  Relationship Between Neointimal Thickness and Shear Stress After Wallstent Implantation in Human Coronary Arteries , 2001, Circulation.

[24]  P. Serruys,et al.  Extension of Increased Atherosclerotic Wall Thickness Into High Shear Stress Regions Is Associated With Loss of Compensatory Remodeling , 2003, Circulation.

[25]  J. Messenger,et al.  Three-Dimensional Analysis of in vivo Coronary Stent – Coronary Artery Interactions , 2004, The International Journal of Cardiovascular Imaging.

[26]  Bernard Chevalier,et al.  Comparison of in vivo acute stent recoil between the bioabsorbable everolimus‐eluting coronary stent and the everolimus‐eluting cobalt chromium coronary stent: Insights from the ABSORB and SPIRIT trials , 2007, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.

[27]  Renu Virmani,et al.  Incidence and predictors of drug-eluting stent fracture in human coronary artery a pathologic analysis. , 2009, Journal of the American College of Cardiology.

[28]  B. Gersh,et al.  A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods , 2010 .

[29]  Jan J Piek,et al.  A randomized comparison of a durable polymer Everolimus-eluting stent with a bare metal coronary stent: The SPIRIT first trial. , 2005, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[30]  R. Virmani,et al.  Engineering aspects of stents design and their translation into clinical practice. , 2007, Annali dell'Istituto superiore di sanita.