Fluid Dynamics of Small Diameter Tubes Used in Membrane-tube Type Glaucoma Shunt Devices

Purpose To investigate the outflow characteristics of silicone tubes with intraluminal stents used in membrane-tube (MT) type glaucoma shunt devices. Methods The silicone tubes used in MicroMT (internal diameter of 100 µm with a 7-0 nylon intraluminal stent) and Finetube MT (internal diameter of 200 µm with a 5-0 nylon intraluminal stent) were connected to a syringe-pump that delivered a continuous flow of distilled water at flow rates of 2, 5, 10, and 25 µL/min. The pressures and resistances of tubes were measured at a steady flow rate with full-length, half-length, and absence of intraluminal stents. Results The mean outflow resistance of the two types of tubes ranged from 3.0 ± 1.9 to 3.8 ± 1.7 mmHg/µL/min with a full-length intraluminal stent, 1.8 ± 1.1 to 2.2 ± 1.1 mmHg/µL/min with a half-length intraluminal stent, and 0.1 ± 0.0 to 0.2 ± 0.0 mmHg/µL/min without an intraluminal stent. Theoretically, for a physiologic state with a flow rate of 2 µL/min and episcleral venous pressure of 6 mmHg, the mean pressures of tubes were expected to be 13.2 ± 3.0, 10.5 ± 2.4, and 6.4 ± 0.2 mmHg in MicroMT with full-length, half-length, and absence of intraluminal stents, respectively, and 12.5 ± 3.9, 9.6 ± 2.4, and 6.2 ± 0.2 mmHg in Finetube MT with full-length, half-length, and absence of intraluminal stents, respectively. The pressure variance also decreased with intraluminal stent retraction (p < 0.01). Conclusions The small diameter tubes of 100 and 200 µm internal diameters, with 7-0 and 5-0 nylon intraluminal stents, respectively, used in the MT-type glaucoma shunt device showed safe and effective outflow characteristics.

[1]  Y. Hwang,et al.  Implantation of a Modified Baerveldt Glaucoma Implant with a Smaller Tube and Intraluminal Stent , 2017, Korean journal of ophthalmology : KJO.

[2]  Y. Hwang,et al.  Safety and Efficacy of a Membrane-Tube-Type Glaucoma Shunt Device: An Animal Trial , 2017, Current eye research.

[3]  Y. Hwang,et al.  Novel membrane‐tube type glaucoma shunt device for glaucoma surgery , 2016, Clinical & experimental ophthalmology.

[4]  M. Noh,et al.  Performance of and Pressure Elevation Formed by Small-diameter Microtubes Used in Constant-flow Sets , 2016, Korean journal of ophthalmology : KJO.

[5]  Y. Hwang,et al.  Membrane-tube-type glaucoma shunt device for refractory glaucoma surgery , 2016, Graefe's Archive for Clinical and Experimental Ophthalmology.

[6]  I. Ahmed,et al.  Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma. , 2015, Investigative ophthalmology & visual science.

[7]  W. Feuer,et al.  Five-year treatment outcomes in the Ahmed Baerveldt comparison study. , 2014, Ophthalmology.

[8]  S. Sarkisian,et al.  Complications of tube implants and their management , 2014, Current opinion in ophthalmology.

[9]  W. Feuer,et al.  Three-year treatment outcomes in the Ahmed Baerveldt comparison study. , 2014, Ophthalmology.

[10]  Dong Myung Kim,et al.  Clinical Outcomes of Ahmed Glaucoma Valve Implantation Using Tube Ligation and Removable External Stents , 2009, Korean journal of ophthalmology : KJO.

[11]  A. J. van der Veen,et al.  Postoperative Pressure Regulation in Glaucoma Shunt Surgery: Focal Tube Constriction is Not the Answer , 2004, Journal of glaucoma.

[12]  You-Jin Choi,et al.  A comparison of the clinical effect between e-PTFE membrane-tube implant and Ahmed glaucoma valve implant for the treatment of refractory glaucoma. , 2003, Korean journal of ophthalmology : KJO.

[13]  L. Labree,et al.  Intraocular pressure control and complications with two-stage insertion of the Baerveldt implant. , 2003, Ophthalmology.

[14]  C. Kim,et al.  Clinical experience of e-PTFE membrane implant surgery for refractory glaucoma , 2003, The British journal of ophthalmology.

[15]  C. Kee Prevention of Early Postoperative Hypotony by Partial Ligation of Silicone Tube in Ahmed Glaucoma Valve Implantation , 2001, Journal of glaucoma.

[16]  J. R. Trible,et al.  Occlusive ligature and standardized fenestration of a Baerveldt tube with and without antimetabolites for early postoperative intraocular pressure control. , 1998, Ophthalmology.

[17]  B. Francis,et al.  Characteristics of glaucoma drainage implants during dynamic and steady-state flow conditions. , 1998, Ophthalmology.

[18]  M. F. Smith,et al.  Prevention of early hypotony associated with Molteno implants by a new occluding stent technique. , 1993, Ophthalmology.

[19]  B. Ahn,et al.  A membranous drainage implant in glaucoma filtering surgery: animal trial. , 1988, Korean journal of ophthalmology : KJO.

[20]  J. S. Jacob,et al.  Single stage insertion of the Molteno tube for glaucoma and modifications to reduce postoperative hypotony. , 1988, The British journal of ophthalmology.