Fast color flow mode imaging using plane wave excitation and temporal encoding

In conventional ultrasound color flow mode imaging, a large number (~500) of pulses have to be emitted in order to form a complete velocity map. This lowers the frame-rate and temporal resolution. A method for color flow imaging in which a few (~10) pulses have to be emitted to form a complete velocity image is presented. The method is based on using a plane wave excitation with temporal encoding to compensate for the decreased SNR, resulting from the lack of focusing. The temporal encoding is done with a linear frequency modulated signal. To decrease lateral sidelobes, a Tukey window is used as apodization on the transmitting aperture. The data are beamformed along the direction of the flow, and the velocity is found by 1-D cross correlation of these data. First the method is evaluated in simulations using the Field II program. Secondly, the method is evaluated using the experimental scanner RASMUS and a 7 MHz linear array transducer, which scans a circulating flowrig. The velocity of the blood mimicking fluid in the flowrig is constant and parabolic, and the center of the scanned area is situated at a depth of 40 mm. A CFM image of the blood flow in the flowrig is estimated from two pulse emissions. At the axial center line of the CFM image, the velocity is estimated over the vessel with a mean relative standard deviation of 2.64% and a mean relative bias of 6.91%. At an axial line 5 mm to the right of the center of the CFM image, the velocity is estimated over the vessel with a relative standard deviation of 0.84% and a relative bias of 5.74%. Finally the method is tested on the common carotid artery of a healthy 33-year-old male.

[1]  Jørgen Arendt Jensen,et al.  Experimental ultrasound system for real-time synthetic imaging , 1999, 1999 IEEE Ultrasonics Symposium. Proceedings. International Symposium (Cat. No.99CH37027).

[2]  J. Jensen,et al.  Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[3]  Jørgen Arendt Jensen,et al.  Ultrasound imaging using coded signals , 2001 .

[4]  M. Fink,et al.  Ultrafast compound imaging for 2-D motion vector estimation: application to transient elastography , 2002, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[5]  J. Jensen,et al.  Directional velocity estimation using focusing along the flow direction. II: experimental investigation , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[6]  J. Jensen Estimation of Blood Velocities Using Ultrasound: A Signal Processing Approach , 1996 .

[7]  D. DeBusschere,et al.  Zone-based color flow imaging , 2003, IEEE Symposium on Ultrasonics, 2003.

[8]  J. Arendt Paper presented at the 10th Nordic-Baltic Conference on Biomedical Imaging: Field: A Program for Simulating Ultrasound Systems , 1996 .

[9]  D. Acheson Elementary Fluid Dynamics , 1990 .

[10]  J. Jensen,et al.  An effective coded excitation scheme based on a predistorted FM signal and an optimized digital filter , 1999, 1999 IEEE Ultrasonics Symposium. Proceedings. International Symposium (Cat. No.99CH37027).

[11]  Jorgen Jensen Velocity vector estimation in synthetic aperture flow and B-mode imaging , 2004, 2004 2nd IEEE International Symposium on Biomedical Imaging: Nano to Macro (IEEE Cat No. 04EX821).

[12]  J. Jensen,et al.  Directional velocity estimation using focusing along the flow direction. I: theory and simulation , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.