The Cleveland Clinic-Nimbus total artificial heart. Design and in vitro function.

We describe the design and in vitro testing of the Cleveland Clinic-Nimbus electrohydraulic permanent total artificial heart as it nears completion of development. The total artificial heart uses an electric motor and hydraulic actuator to drive two diaphragm-type blood pumps. The interventricular space contains the pump control electronics and is vented to an air-filled compliance chamber. Pericardial tissue valves and biolized blood-contacting surfaces potentially eliminate the need for anticoagulation. In vitro studies on a mock circulatory circuit demonstrated preload-sensitive control of pump output over the operating range of the blood pump: 70 to 160 beats/min and 5 to 9.6 L/min at right and left atrial pressures of 1.0 to 7.0 mm Hg and 5.0 to 12.0 mm Hg, respectively. The pump output was found to be insensitive to afterload over a range of 15 to 40 mm Hg mean pulmonary artery pressure and 60 to 130 mm Hg mean systemic pressure. The left master alternate control mode balanced the ventricular outputs during simulated bronchial artery shunting of up to 20% of cardiac output. A 10% to 15% right-pump, stroke-volume limiter balanced ventricular outputs during maximum output of 9.6 L/min. In response to a sustained increase in systemic venous return, the pump increased output by 2 L/min (29%) in 35 seconds. Thus the Cleveland Clinic-Nimbus total artificial heart meets the National Heart, Lung, and Blood Institute hemodynamic performance goals for devices being developed for permanent heart replacement. The biolized blood-contacting surfaces should decrease the risk of thromboembolism associated with circulatory assist devices.

[1]  A. Guyton,et al.  Textbook of Medical Physiology , 1961 .

[2]  Development of a completely implantable total artificial heart. , 1988, ASAIO transactions.

[3]  H. Harasaki,et al.  The Cleveland Clinic-Nimbus total artificial heart. In vivo hemodynamic performance in calves and preclinical studies. , 1994, The Journal of thoracic and cardiovascular surgery.

[4]  H. Harasaki,et al.  "Pass-through" and "inertia" contribution to left-right flow difference (LRFD) in TAH recipients. , 1985, Transactions - American Society for Artificial Internal Organs.

[5]  O. Frazier,et al.  Clinical experience with textured blood contacting surfaces in ventricular assist devices. , 1987, ASAIO transactions.

[6]  J. Duncan,et al.  Clinical Evaluation Gf Right Ventricular Function in Patients with Left Ventricular Assist Device (Lvad) , 1992, International Journal of Artificial Organs.

[7]  W. Pae,et al.  Ventricular assist devices and total artificial hearts: a combined registry experience. , 1993, The Annals of thoracic surgery.

[8]  Malin VanAntwerp,et al.  The Artificial Heart: Prototypes, Policies, and Patients , 1992, Annals of Internal Medicine.

[9]  D Hillegass,et al.  HIGH FLEX RUBBER FOR BLOOD PUMP DIAPHRAGMS , 1977, Transactions - American Society for Artificial Internal Organs.

[10]  P M Portner,et al.  Clinical experience with the Novacor ventricular assist system. Bridge to transplantation and the transition to permanent application. , 1991, The Journal of thoracic and cardiovascular surgery.

[11]  E S Bücherl,et al.  The relationship of cardiac output and venous pressure in long surviving calves with total artificial heart. , 1978, Transactions - American Society for Artificial Internal Organs.

[12]  D. Cosgrove,et al.  The Carpentier-Edwards pericardial aortic valve: intermediate results. , 1992, The Annals of thoracic surgery.

[13]  D. Hosmer,et al.  Maximal oxygen intake and nomographic assessment of functional aerobic impairment in cardiovascular disease. , 1973, American heart journal.

[14]  Kaye Mp The Registry of the International Society for Heart and Lung Transplantation: tenth official report--1993. , 1993, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[15]  B. Griffith,et al.  How much do we need to know before approving a ventricular assist device? , 1993, The Annals of thoracic surgery.

[16]  R. Emery,et al.  Summary of the clinical use of the Symbion total artificial heart: a registry report. , 1992, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[17]  H Harasaki,et al.  Development of the E4T electrohydraulic total artificial heart. , 1990, ASAIO transactions.

[18]  B. Griffith,et al.  Experience with univentricular support in mortally ill cardiac transplant candidates. , 1990, The Annals of thoracic surgery.

[19]  K. Sunagawa,et al.  Effects of left heart bypass on right ventricular performance. Evaluation of the right ventricular end-systolic and end-diastolic pressure-volume relation in the in situ normal canine heart. , 1990, The Journal of thoracic and cardiovascular surgery.

[20]  O. Frazier Chronic left ventricular support with a vented electric assist device. , 1993, The Annals of thoracic surgery.