Dosing Accuracy of Insulin Aspart FlexPens after Transport through the Pneumatic Tube System

Purpose The purpose of this study was to evaluate whether transporting insulin aspart FlexPens via a pneumatic tube system affects the dosing accuracy of the pens. Methods A total of 115 Novo Nordisk FlexPens containing insulin aspart were randomly assigned to be transported via a pneumatic tube system (n = 92) or to serve as the control (n = 23). Each pen was then randomized to 10 international unit (IU) doses (n = 25) or 30 IU doses (n = 67), providing 600 and 603 doses, respectively, for the pneumatic tube group. The control group also received random assignment to 10 IU doses (n = 6) or 30 IU doses (n = 17), providing 144 and 153 doses, respectively. Each dose was expelled using manufacturer instructions. Weights were recorded, corrected for specific gravity, and evaluated based on acceptable International Organization for Standardization (ISO) dosing limits. Results In the group of pens transported through the pneumatic tube system, none of the 600 doses of 10 IU (0.0%; 95% CI, 0.0 to 0.6) and none of the 603 doses of 30 IU (0.0%; 95% CI, 0.0 to 0.6) fell outside of the range of acceptable weights. Correspondingly, in the control group, none of the 144 doses at 10 IU (0.0%; 95% CI, 0.0 to 2.5) and none of the 153 doses at 30 IU (0.0%; 95% CI, 0.0 to 2.4) were outside of acceptable ISO limits. Conclusion Transportation via pneumatic tube system does not appear to compromise dosing accuracy. Hospital pharmacies may rely on the pneumatic tube system for timely and accurate transport of insulin aspart FlexPens.

[1]  C. Destache,et al.  Patient satisfaction and costs associated with insulin administered by pen device or syringe during hospitalization. , 2008, American journal of health-system pharmacy : AJHP : official journal of the American Society of Health-System Pharmacists.

[2]  A. Pfützner,et al.  Differences in the Dose Accuracy of Insulin Pens , 2008, Journal of diabetes science and technology.

[3]  Michael Safinowski,et al.  Comparison of the Dose Accuracy of Prefilled Insulin Pens , 2009, Journal of diabetes science and technology.

[4]  M. Korytkowski,et al.  A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitus. , 2003, Clinical therapeutics.

[5]  W. Schwenk,et al.  Accuracy of pen injectors versus insulin syringes in children with type 1 diabetes. , 1999, Diabetes care.

[6]  Leah G Ward,et al.  Impact of an interchange program to support use of insulin pens. , 2011, American journal of health-system pharmacy : AJHP : official journal of the American Society of Health-System Pharmacists.

[7]  P. Newland,et al.  Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes , 1998, Archives of disease in childhood.

[8]  Katherine Keith,et al.  Accuracy and Precision of Low-Dose Insulin Administration Using Syringes, Pen Injectors, and a Pump , 2004, Clinical pediatrics.