Quantitative enzyme-linked immunosorbent assay for determination of polychlorinated biphenyls in environmental soil and sediment samples.
暂无分享,去创建一个
An enzyme-linked immunosorbent assay (ELISA) for the quantitative determination of Aroclors 1242, 1248, 1254, and 1260 in soil and sediments was developed and its performance compared with that of gas chromatography (GC). The detection limits for Aroclors 1242 and 1248 in soil are 10.5 and 9 ng/g, respectively. The assay linear dynamic range is 50-1333 ng/g. Cross-reactivity of the assay with 37 structurally related potential cocontaminants in environmental soil samples was examined; none of the chlorinated anisoles, benzenes, or phenols exhibited >3% cross-reactivity, with <0.1% cross-reactivity being the norm. Soil spike recoveries of 107% and 104% were obtained for Aroclors 1242 and 1248, respectively, for a spike level of 5 mg/kg, with corresponding relative standard deviations of 14% and 17%. One hundred forty-eight environmental soil, sediment, and paper pulp samples, obtained from two EPA listed Superfund sites, were analyzed by ELISA and standard GC methods. Samples were extracted for ELISA analysis by shaking with methanol. Additional extractions of the same samples were performed either with supercritical carbon dioxide or by Soxhlet extraction with methanol. ELISA results for both the supercritical fluid and the Soxhlet extracts were in close agreement with the GC results, while the ELISA results for the methanol shake extracts were not. The data for the environmental samples demonstrated the capability of the ELISA to provide accurate results and reinforced the dependence of any detection method, including ELISA, on appropriate extraction procedures.