An edge spread technique for measurement of the scatter-to-primary ratio in mammography.

An experimental measurement technique that directly measures the magnitude and spatial distribution of scatter in relation to primary radiation is presented in this work. The technique involves the acquisition of magnified edge spread function (ESF) images with and without scattering material present. The ESFs are normalized and subtracted to yield scatter-to-primary ratios (SPRs), along with the spatial distributions of scatter and primary radiation. Mammography is used as the modality to demonstrate the ESF method, which is applicable to all radiographic environments. Sets of three images were acquired with a modified clinical mammography system employing a flat panel detector for 2, 4, 6, and 8 cm thick breast tissue equivalent material phantoms composed of 0%, 43%, and 100% glandular tissue at four different kV settings. Beam stop measurements of scatter were used to validate the ESF methodology. There was good agreement of the mean SPRs between the beam stop and ESF methods. There was good precision in the ESF-determined SPRs with a coefficient of variation on the order of 5%. SPRs ranged from 0.2 to 2.0 and were effectively independent of energy for clinically realistic kVps. The measured SPRs for 2, 4, and 6 cm 0% glandular phantoms imaged at 28 kV were 0.21+/-0.01, 0.39+/-0.01, and 0.57+/-0.02, respectively. The measured SPRs for 2, 4, and 6 cm 43% glandular phantoms imaged at 28 kV were 0.20+/-0.01, 0.35+/-0.02, and 0.53+/-0.02, respectively. The measured SPRs for 2, 4, and 6 cm 100% glandular phantoms imaged at 28 kV were 0.22+/-0.02, 0.42+/-0.03, and 0.88+/-0.08, respectively.

[1]  D R Dance,et al.  The computation of scatter in mammography by Monte Carlo methods. , 1984, Physics in medicine and biology.

[2]  G. Fagerberg,et al.  Image Quality in Mammography with Special Reference to Anti-Scatter Grids and the Magnification Technique , 1986, Acta radiologica: diagnosis.

[3]  J. Motz,et al.  X-ray scatter background signals in transmission radiography. , 1975, Medical physics.

[4]  C E Ravin,et al.  Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system. , 1991, Medical physics.

[5]  M A King,et al.  Experimental measurements of the scatter reduction obtained in mammography with a scanning multiple slit assembly. , 1981, Medical physics.

[6]  H Bernstein,et al.  Experimental and theoretical energy and angular dependencies of scattered radiation in the mammography energy range. , 1983, Medical physics.

[7]  E. Muntz,et al.  Focal spot size and scatter suppression in magnification mammography. , 1979, AJR. American journal of roentgenology.

[8]  J A Seibert,et al.  Characterization of the veiling glare PSF in x-ray image intensified fluoroscopy. , 1984, Medical physics.

[9]  E. Muntz,et al.  Analysis of the significance of scattered radiation in reduced dose mammography, including magnification effects, scatter suppression, and focal spot and detector blurring. , 1979, Medical physics.

[10]  J A Seibert,et al.  X-ray scatter removal by deconvolution. , 1988, Medical physics.

[11]  H Bernstein,et al.  Experimental verification of a technique for predicting scattered radiation transfer: application to low photon energies. , 1986, Medical physics.

[12]  R Fahrig,et al.  Performance of glass fiber antiscatter devices at mammographic energies. , 1994, Medical physics.

[13]  M Souto,et al.  Scatter in computed radiography. , 1993, Radiology.

[14]  S. C. Prasad,et al.  Scatter reduction in mammography with air gap. , 1996, Medical physics.

[15]  J. Vetter,et al.  Entrance skin exposure and mean glandular dose: effect of scatter and field gradient at mammography. , 1997, Radiology.

[16]  C E Ravin,et al.  Posterior beam-stop method for scatter fraction measurement in digital radiography. , 1992, Investigative radiology.

[17]  O. Korhola,et al.  The effect of scatter reduction on the signal-to-noise ratio in computed radiography. , 1991, European journal of radiology.

[18]  J A Seibert,et al.  Removal of image intensifier veiling glare by mathematical deconvolution techniques. , 1985, Medical physics.

[19]  G T Barnes,et al.  Contrast and scatter in x-ray imaging. , 1991, Radiographics : a review publication of the Radiological Society of North America, Inc.

[20]  I A Brezovich,et al.  The intensity of scattered radiation in mammography. , 1978, Radiology.

[21]  G T Barnes,et al.  Mammography grid performance. , 1999, Radiology.

[22]  W Huda,et al.  Scattered radiation in scanning slot mammography. , 1998, Medical physics.

[23]  S L Fritz,et al.  Scatter/primary ratios for x-ray spectra modified to enhance iodine contrast in screen-film mammography. , 1983, Medical physics.

[24]  P Sprawls,et al.  Grids in mammography. , 1983, Radiology.