99 Tc Process Monitoring System In-Lab Performance Characterization

Executive Summary A 99Tc Process Monitoring (Tc-Mon) System has been designed and built for deployment at the recently constructed 200 West Pump & Treat (200W PT therefore, once they have reached their maximum useful capacity, the columns must be disposed of as radioactive waste. The Tc-Mon system’s primary duty will be to periodically sample and analyze the effluents from each of the two primary extraction columns to determine 99Tc breakthrough. The Tc-Mon system willmore » enable the CH2M Hill Plateau Remediation Company (CHPRC) to measure primary extraction column breakthrough on demand. In this manner, CHPRC will be able to utilize each extraction column to its maximum capacity. This will significantly reduce column disposal and replacement costs over the life of the plant. The Tc-Mon system was constructed by Burge Environmental, Inc. and was delivered to PNNL in June 2013 for setup and initial hardware and software performance testing in the 325 Building. By early July, PNNL had initiated an in-laboratory performance characterization study on the system. The objective was to fully calibrate the system and then evaluate the quality of the analytical outputs 1) against a series of clean groundwater samples prepared as 99Tc standards, and 2) on actual 200W P&T Plant grab samples containing 99Tc (and other radioactive and non-radioactive contaminants) at levels reported by the Waste Sampling and Characterization Facility. These grab samples included pre-treated (Pre-Resin) and post-treated (Post-Resin) 200W P&T Plant waters for May through August 2013. This report contains the following information: • The genesis of the 99Tc sensor and the Tc-Mon analytical system. • A description of the Tc-Mon system’s major hardware and software components. • A description of the operational principles behind the 99Tc sensor. • Results from the calibration of three components within the Tc-Mon system. The three systems requiring calibration are: 1. Sampling Chamber 2. Conductivity Sensor 3. 99Tc Sensor • Presentation of analytical results obtained on the fully calibrated Tc-Mon system. This includes a determination of the precision and accuracy of each system defined above. • Estimation of the 99Tc sensor’s minimum detectable activity and limit of quantification. • A brief discussion of potential chemical and radiological influences on the 99Tc sensor based on known contaminants in 200W P&T Plant water. • Observations regarding 99Tc sensor longevity.« less

[1]  J. Grate,et al.  Automated radiochemical separation, analysis, and sensing , 2020, Handbook of Radioactivity Analysis: Volume 2.

[2]  Wooyong Um,et al.  Linearity and reversibility of iodide adsorption on sediments from Hanford, Washington under water saturated conditions. , 2004, Water research.

[3]  J. Grate,et al.  Automated radioanalytical system for the determination of 90Sr in environmental water samples by 90Y Cherenkov radiation counting. , 2009, Analytical chemistry.

[4]  H. Spring,et al.  Determination of technetium-99 by liquid scintillation counting , 1979 .

[5]  Mary J. Hartman,et al.  Automated Ground-Water Sampling and Analysis of Hexavalent Chromium using a “Universal” Sampling/Analytical System , 2005, Sensors (Basel, Switzerland).

[6]  J. Laul Natural radionuclides in groundwaters , 1992 .

[7]  J. Grate,et al.  Quantification of technetium-99 in complex groundwater matrixes using a radiometric preconcentrating minicolumn sensor in an equilibration-based sensing approach. , 2009, Analytical chemistry.

[8]  Lloyd A. Currie,et al.  Limits for qualitative detection and quantitative determination , 1968 .

[9]  Jay W Grate,et al.  Equilibration-based preconcentrating minicolumn sensors for trace level monitoring of radionuclides and metal ions in water without consumable reagents. , 2006, Analytical chemistry.

[10]  L. A. Currie,et al.  LIMITS FOR QUALITATIVE DETECTION AND QUANTITATIVE DETERMINATION. APPLICATION TO RADIOCHEMISTRY. , 1968 .