Infrared camera NUC and calibration: comparison of advanced methods
暂无分享,去创建一个
Image uniformity and accurate radiometric calibration are key features of state-of-the-art infrared cameras. Over the past years several non-uniformity correction and radiometric calibration techniques have been developed. In this paper we present and compare different techniques: 2-point calibration, CNUC™/multipoint’s calibration and Telops’ Real-Time Image Processing (patent-pending). For each method we assess the performances, the ease of use, the advantages and drawbacks as well as the most important operational limitations considering a broad range of exposure times, ambient and scene temperatures.