How much can a vibrational diagnostic tool reveal in total hip arthroplasty loosening?

OBJECTIVE To investigate how much information a vibrational technique can reveal regarding the loosening of the femoral component of a total hip arthroplasty. DESIGN Numerical modal analysis using finite element method and computer simulation. BACKGROUND Existing work suggested that this technique is capable of revealing total hip arthroplasty loosening. However, the potentials of this technique are not fully studied yet. There is a need to establish the limitation of this technique. METHODS Numerical modal analysis was conducted to obtain the natural frequencies and mode shapes of femoral models under free vibration. An absent interface portion was hypothesized at specified locations the data from modal analysis were then used in a computer simulation to visualize the vibration diagnosis. RESULTS The effects of interface failure on the femoral component stiffness vary with the locations and sizes of failure. Nominal critical, reliable, and nondetectable levels of interface failure are found. CONCLUSION When the size of failure is more than one-third of the stem length, this tool is able to provide a reliable diagnosis. The critical detectable interface failure size is one-fifth of the stem length. It was found that the motion at higher harmonics is the most sensitive to the interface failure. This tool is unlikely to differentiate the slight differences in natural frequencies that occur when the interface failure length is less than one-sixth of the stem length. It is likely to fail in detecting failures located at the stem central portion regardless of their length. RELEVANCE A numerical study of the vibration tool is important prior to pursuing clinical trials of the tool and may guide the characterizations involved in the trials.

[1]  N B Jones,et al.  Loosening of total hip arthroplasty. Diagnosis by vibration analysis. , 1995, The Journal of bone and joint surgery. British volume.

[2]  G Qi,et al.  Attenuation of acoustic emission body waves in acrylic bone cement and synthetic bone using wavelet time-scale analysis. , 2000, Journal of biomedical materials research.

[3]  G F McCoy,et al.  The Differentiation of Loose and Secure Femoral Implants in Total Hip Replacement Using a Vibrational Technique: An Anatomical and Pilot Clinical Study , 1989, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[4]  P L Li,et al.  Vibration analysis in the detection of total hip prosthetic loosening. , 1996, Medical engineering & physics.

[5]  Leonard Meirovitch,et al.  Elements Of Vibration Analysis , 1986 .

[6]  D L Bartel,et al.  Coulomb frictional interfaces in modeling cemented total hip replacements: a more realistic model. , 1995, Journal of biomechanics.

[7]  S. T. Buckland,et al.  An Introduction to the Bootstrap. , 1994 .

[8]  Ingrid Daubechies,et al.  The wavelet transform, time-frequency localization and signal analysis , 1990, IEEE Trans. Inf. Theory.

[9]  I. Daubechies Orthonormal bases of compactly supported wavelets , 1988 .

[10]  F W Werner,et al.  Tensile strength of the cement-bone interface depends on the amount of bone interdigitated with PMMA cement. , 1997, Journal of biomechanics.

[11]  Gang Qi,et al.  ON PREDICTING THE FRACTURE BEHAVIOR OF CFR AND GFR COMPOSITES USING WAVELET-BASED AE TECHNIQUES , 1997 .

[12]  K A Mann,et al.  Fracture toughness of CoCr alloy-PMMA cement interface. , 1997, Journal of biomedical materials research.

[13]  Gang Qi,et al.  Material fatigue behavior characterization using the wavelet-based AE technique – a case study of acrylic bone cement , 2001 .

[14]  Richard A. Brand Micromovement in Orthopaedics , 1995 .

[15]  Stéphane Mallat,et al.  A Theory for Multiresolution Signal Decomposition: The Wavelet Representation , 1989, IEEE Trans. Pattern Anal. Mach. Intell..

[16]  Robert Puers,et al.  An implantable system for detecting loosening of a hip prosthesis , 1999 .

[17]  Gang Qi,et al.  A new method to evaluate material fatigue behavior , 2001 .

[18]  D L Bartel,et al.  Influence of stem geometry on mechanics of cemented femoral hip components with a proximal bond , 1997, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[19]  G. Qi Wavelet-based AE characterization of composite materials , 2000 .

[20]  HighWire Press,et al.  The journal of bone and joint surgery - British volume , 1948 .

[21]  Gang Qi,et al.  Discrete wavelet decomposition of acoustic emission signals from carbon-fiber-reinforced composites , 1997 .

[22]  H. Amstutz,et al.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening. , 1979, Clinical orthopaedics and related research.

[23]  A. Georgiou,et al.  Accurate diagnosis of hip prosthesis loosening using a vibrational technique. , 2001, Clinical biomechanics.