Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source.
暂无分享,去创建一个
This work provides full dosimetric data for a new high-strength 192Ir source currently launched by Varian Oncology Systems for use in their high dose rate remote afterloading systems. The active core length of the new source is reduced to 5 mm compared to a value of 10 mm for the existing VariSource source design, with all other geometric source and encapsulation details being similar. Dose-rate constant, radial dose functions, geometry factors, and anisotropy functions, utilized in the AAPM Task Group 43 dose calculation formalism, were calculated using Monte Carlo simulation. Results are compared with corresponding data published for the existing VariSource and microSelectron high dose rate sources. The dose-rate constant for the new Varian source was found to be equal to 1.101 +/- 0.006cGyh(-1) U(-1), compared to values of 1.043 +/- 0.005 and 1.116 +/- 0.006 cGyh(-1) U(-1) calculated for the existing VariSource and microSelectron sources, respectively. The radial dose functions between the three sources are similar with the exception of their values at radial distances very close to the source (r approximately 2 mm) where differences of approximately 3% are observed. The new Varian source demonstrates a smaller anisotropy relative to the existing VariSource source design for polar angles close to the source longitudinal axis, due to its smaller active core length.