A MCM-based microsystem for biological fluids analysis by optical absorption
暂无分享,去创建一个
This paper presents a MCM (multi-chip-module) microsystem for biological fluids analysis. It. is composed of three parts: (1) a CMOS silicon optical detection microsystem which includes a photodetector and a light-to-frequency converter for readout; (2) a dielectric thin-film based optical filter on top of the photodetector, fabricated using IC-compatible post-processing; (3) a glass die on top containing the microchannels fabricated using SU-8 techniques. The application is in the low-cost measurement of concentration of biomolecules in biological fluids, by the optical absorption in a part of the visible spectrum defined by the specific molecule. Signals proportional to the intensity of the light transmitted through the biological fluid are available at the output in the form of bit streams, which allows simple computer interfacing. The quantitative measurement of total protein in urine is successfully demonstrated. The photodiode responsivity is 224 mA/W with a FWHM (full-width-half-maximum) of 10 nm at /spl lambda/=600 nm. The optical system sensitivity is 1 kHz/Wm/sup -2/ at /spl lambda/=670 nm.
[1] Reinoud F. Wolffenbuttel,et al. Optical CMOS sensor system for detection of light sources , 2004 .
[2] José Higino Correia,et al. 2-D modeling and simulation of fluidic microsystems for biological fluids analysis , 2003 .
[3] B. Fowler,et al. A CMOS area image sensor with pixel-level A/D conversion , 1994, Proceedings of IEEE International Solid-State Circuits Conference - ISSCC '94.
[4] P. Connolly. Clinical diagnostics opportunities for biosensors and bioelectronics. , 1995, Biosensors & bioelectronics.