Performance of StatSensor Point-of-Care Device for Measuring Creatinine in Patients With Chronic Kidney Disease and Postkidney Transplantation

Background: The StatSensor is a point-of-care device which measures creatinine in capillary whole blood. Previous studies reported an underestimation of the creatinine measurements at high creatinine concentrations and were performed in the prestandardization era for creatinine. Objective: This accuracy-based study evaluates the use of this device in kidney-transplanted patients and those with chronic kidney disease (CKD). Design: Cross-sectional diagnostic accuracy study. Setting: Nephrology outpatient clinic in an urban tertiary center. Participants: Adults with CKD or a functioning kidney transplant. Measurements: Duplicate StatSensor creatinine measurements were performed on capillary whole blood samples collected by direct fingerstick and SAFE-T-FILL collection device. Results were compared with simultaneous venous blood sampling for serum and plasma creatinine measured by an enzymatic method on the Roche Integra 400 mainframe analyzer with traceability to the ID-GC-MS (isotope dilution gas chromatography mass spectrometry) reference method. Methods: Deming regression, Pearson correlation coefficient, and Bland-Altman analysis were used to assess accuracy and comparability between capillary whole blood measured by StatSensor and plasma creatinine measured by routine analyzer with traceability to the reference method. Estimated glomerular filtration (eGFR) rates were calculated using the Chronic Kidney Disease Epidemiology Collaboration (CKD-EPI) equation and concordance with Kidney Disease Improving Global Outcomes (KDIGO) CKD stage classification was evaluated. Results: There were 60 participants (mean age = 61.9 ± 15.0 years, 55% men, 33% transplant, mean plasma creatinine = 137 ± 59 µmol/L). Bland-Altman analysis indicated a positive mean bias of 12.7 µmol/L between StatSensor fingerstick creatinine measurement and plasma creatinine. Comparison of eGFR (CKD-EPI) calculated from the StatSensor fingerstick creatinine versus plasma creatinine showed misclassification across all KDIGO CKD stages. Postanalytical correction of the bias did not improve misclassifications. The use of mean of duplicate StatSensor creatinine results did not improve performance compared with the use of singlet results. Limitations: Single center, limited participant numbers. Conclusions: The results of our study suggest that the limiting characteristics of the StatSensor device are not only bias, but also imprecision. The level of imprecision observed may influence clinical decision-making and limit the usefulness of StatSensor as a CKD screening tool. If choosing to utilize it for either screening for or monitoring CKD, it is essential that clinicians understand the limitations of point-of-care devices and apply this knowledge to test interpretation.

[1]  E. Cavalier,et al.  Estimated glomerular filtration rate using a point of care measure of creatinine in patients with iohexol determinate GFR. , 2019, Clinica chimica acta; international journal of clinical chemistry.

[2]  M. Woodward,et al.  Past Decline Versus Current eGFR and Subsequent Mortality Risk. , 2016, Journal of the American Society of Nephrology : JASN.

[3]  A. Milovic,et al.  Evaluation of the Nova StatSensor® XpressTM Creatinine Point-Of-Care Handheld Analyzer , 2015, PloS one.

[4]  Daniel T Holmes,et al.  cp-R, an interface the R programming language for clinical laboratory method comparisons. , 2015, Clinical biochemistry.

[5]  M. Tonelli,et al.  Short‐Term Change in eGFR and Risk of Cardiovascular Events , 2014, Journal of the American Heart Association.

[6]  J. Coresh,et al.  Change in the estimated glomerular filtration rate over time and risk of all-cause mortality. , 2013, Kidney international.

[7]  S. Schoenberg,et al.  Evaluation of a handheld creatinine measurement device for real-time determination of serum creatinine in radiology departments. , 2012, World journal of radiology.

[8]  A. Lyon,et al.  Investigating interferences of a whole-blood point-of-care creatinine analyzer: comparison to plasma enzymatic and definitive creatinine methods in an acute-care setting. , 2011, Clinical chemistry.

[9]  Bertram L Kasiske,et al.  The definition, classification, and prognosis of chronic kidney disease: a KDIGO Controversies Conference report. , 2011, Kidney international.

[10]  Olivia Corso,et al.  Assessment of the Nova StatSensor whole blood point-of-care creatinine analyzer for the measurement of kidney function in screening for chronic kidney disease , 2010, Clinical chemistry and laboratory medicine.

[11]  J. Dubois,et al.  Evaluation of the analytical performance of the Nova StatSensor creatinine meter and reagent strip technology for whole blood testing. , 2010, Clinical biochemistry.

[12]  E. Williamson,et al.  Comparison of three whole blood creatinine methods for estimation of glomerular filtration rate before radiographic contrast administration. , 2009, American journal of clinical pathology.

[13]  C. Schmid,et al.  A new equation to estimate glomerular filtration rate. , 2009, Annals of internal medicine.

[14]  R. Jansen,et al.  Trueness verification of actual creatinine assays in the European market demonstrates a disappointing variability that needs substantial improvement. An international study in the framework of the EC4 creatinine standardization working group , 2008, Clinical chemistry and laboratory medicine.

[15]  A. Levin,et al.  Regional implementation of creatinine measurement standardization. , 2008, Journal of the American Society of Nephrology : JASN.

[16]  M. Welch,et al.  Certification of creatinine in a human serum reference material by GC-MS and LC-MS. , 2007, Clinical chemistry.

[17]  Tom Greene,et al.  Using Standardized Serum Creatinine Values in the Modification of Diet in Renal Disease Study Equation for Estimating Glomerular Filtration Rate , 2006, Annals of Internal Medicine.

[18]  W Greg Miller,et al.  Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program. , 2006, Clinical chemistry.

[19]  G. Eknoyan,et al.  Definition and classification of chronic kidney disease: a position statement from Kidney Disease: Improving Global Outcomes (KDIGO). , 2005, Kidney international.

[20]  A. Killeen,et al.  Creatinine measurement: state of the art in accuracy and interlaboratory harmonization. , 2005, Archives of pathology & laboratory medicine.