Breath alcohol concentration determined with a new analyzer using free exhalation predicts almost precisely the arterial blood alcohol concentration.

A new breath alcohol (ethanol) analyzer has been developed, which allows free exhalation, standardizes measured exhaled alcohol concentration to fully saturated water vapor at a body temperature of 37 degrees C (43.95 mg/L) and includes a built-in self-calibration system. We evaluated the performance of this instrument by comparing standardized alcohol concentration in freely expired breath (BrAC) with arterial (ABAC) and venous (VBAC) blood alcohol concentrations in fifteen healthy volunteers who drank 0.6 g of alcohol per kg body weight. The precision (coefficient of variation, CV) of the analyzer based on in vivo duplicate measurements in all phases of the alcohol metabolism was 1.7%. The ABAC/BrAC ratio was 2251+/-46 (mean+/-S.D.) in the post-absorptive phase and the mean bias between ABAC and BrAC x 2251 was 0.0035 g/L with 95% limits of agreement of 0.033 and -0.026. The ABAC and BrAC x 2251 were highly correlated (r=0.998, p<0.001) and the regression relationship was ABAC = 0.00045 + 1.0069 x (BrAC x 2251) indicating excellent agreement and no fixed or proportional bias. In the absorption phase, ABAC exceeded BrAC x 2251 by at most 0.04+/-0.03 g/L when tests were made at 10 min post-dosing (p<0.05). The VBAC/BrAC ratio never stabilized and varied continuously between 1834 and 3259. There was a proportional bias between VBAC and BrAC x 2251 (ABAC) in the post-absorptive phase (p<0.001). The pharmacokinetic analysis of the elimination rates of alcohol and times to zero BAC confirmed that BrAC x 2251 and ABAC agreed very well with each other, but not with VBAC (p<0.001). We conclude that this new breath analyzer using free exhalation has a high precision for in vivo testing. The BrAC reflects very accurately ABAC in the post-absorption phase and substantially well in the absorption phase and thereby reflects the concentration of alcohol reaching the brain. Our findings highlight the magnitude of arterio-venous differences in alcohol concentration and support the use of breath alcohol analyzers as a stand-alone test for medical and legal purposes.

[1]  K. Dubowski,et al.  Breath-alcohol analysis: uses, methods, and some forensic problems--review and opinion. , 1976, Journal of forensic sciences.

[2]  E. Aharonson,et al.  Effect of respiratory airflow rate on removal of soluble vapors by the nose. , 1974, Journal of applied physiology.

[3]  E. Martin,et al.  The pharmacokinetics of alcohol in human breath, venous and arterial blood after oral ingestion , 2004, European Journal of Clinical Pharmacology.

[4]  W. Neuteboom,et al.  The effects of hypo- and hyperventilation on breath alcohol measurements. , 1987, Blutalkohol.

[5]  K. Dubowski Breath-Alcohol Simulators: Scientific Basis and Actual Performance , 1979 .

[6]  M P Hlastala,et al.  The alcohol breath test--a review. , 1998, Journal of applied physiology.

[7]  Alcohol Breath Tests: Gross Errors in Current Methods of Measuring Alveolar Gas Concentrations , 1971, Science.

[8]  G. Eisenhofer,et al.  Effect of ethanol ingestion on plasma vasopressin and water balance in humans. , 1982, The American journal of physiology.

[9]  R. Forney,et al.  Estimation of the level of blood alcohol from analysis of breath. , 1950, The Journal of laboratory and clinical medicine.

[10]  A. Jones,et al.  Variability of the blood/breath alcohol ratio in drinking drivers. , 1996, Journal of forensic sciences.

[11]  E. R. Mcfadden,et al.  Respiratory heat and water exchange: physiological and clinical implications. , 1983, Journal of applied physiology: respiratory, environmental and exercise physiology.

[12]  J. Hayward,et al.  Effect of hyperthermia on breath-alcohol analysis. , 1987, Journal of forensic sciences.

[13]  L. Glasser,et al.  Serum osmolality and its applicability to drug overdose. , 1973, American journal of clinical pathology.

[14]  A. Jones,et al.  Determination of liquid/air partition coefficients for dilute solutions of ethanol in water, whole blood, and plasma. , 1983, Journal of analytical toxicology.

[15]  A. Jones,et al.  Comparison of ethanol concentrations in venous blood and end-expired breath during a controlled drinking study. , 2003, Forensic science international.

[16]  A. Zwart,et al.  Uptake of highly soluble gases in the epithelium of the conducting airways , 1985, Pflügers Archiv.

[17]  A. Jones,et al.  How Breathing Technique Can Influence the Results of Breath-Alcohol Analysis , 1982, Medicine, science, and the law.

[18]  A. Jones,et al.  Computer-aided headspace gas chromatography applied to blood-alcohol analysis: importance of online process control. , 1989, Journal of forensic sciences.

[19]  W. L. Chiou,et al.  The Phenomenon and Rationale of Marked Dependence of Drug Concentration on Blood Sampling Site , 1989, Clinical pharmacokinetics.

[20]  A. Jones,et al.  Laboratory Evaluation of a New Evidential Breath-Alcohol Analyser Designed for Mobile Testing – the Evidenzer , 2005, Medicine, science, and the law.

[21]  J J Opdam,et al.  Alveolar sampling and fast kinetics of tetrachloroethene in man. I. Alveolar sampling. , 1986, British journal of industrial medicine.

[22]  Robert Q. Thompson The Thermodynamics of Drunk Driving , 1997 .

[23]  G. Schoknecht,et al.  [Influence of breath volume on alcohol analysis]. , 1990, Blutalkohol.

[24]  A. Jones Effects of temperature and humidity of inhaled air on the concentration of ethanol in a man's exhaled breath. , 1982, Clinical science.

[25]  J. Wagner,et al.  Concentrations of ethanol in two segments of the vascular system. , 1976, Journal of forensic sciences.

[26]  R. Forney,et al.  Estimation of the level of blood alcohol from analysis of breath. II. Use of rebreathed air. , 1956, Quarterly journal of studies on alcohol.

[27]  D. Altman,et al.  Measuring agreement in method comparison studies , 1999, Statistical methods in medical research.