Magnetic measurement of human iron stores

This paper provides a review of the principles and practice of biomagnetic susceptometry, with particular reference to magnetic susceptibility measurements of the human liver. A novel SQUID biomagnetic susceptometer is described consisting of a second-order gradiometer combined with a superconducting field coil assembly. We report the first clinical non-invasive diagnosis of liver iron overload achieved with this susceptometer at the Metropolitan General Hospital in Cleveland. Estimates are provided for the limiting resolution of the method. Ongoing efforts to apply the same principles to the determination of iron overload in the heart are briefly described. Our work represents the first clinical diagnostic procedure to utilize the recently developed superconducting technology of weak field measurement.

[1]  C. D. Coryell,et al.  FERRITIN III. THE MAGNETIC PROPERTIES OF FERRITIN AND SOME OTHER COLLOIDAL FERRIC COMPOUNDS , 1943 .

[2]  M Reite,et al.  Magnetic phenomena of the central nervous system. , 1978, Annual review of biophysics and bioengineering.

[3]  J. Tripp,et al.  Non-invasive in vivo measurement of human iron overload by magnetic susceptibility studies. , 1980, Transactions of the Association of American Physicians.

[4]  L. Mulay,et al.  Magnetic Susceptibility: Recent Aspects of Instrumentation and Applications , 1966 .

[5]  A. Weinfeld STORAGE IRON IN MAN. , 1964, Acta medica Scandinavica. Supplementum.

[6]  D. Cohen Magnetic fields of the human body , 1975 .

[7]  J H Bauman,et al.  Estimation of hepatic iron stores by vivo measurement of magnetic susceptibility. , 1967, The Journal of laboratory and clinical medicine.