A method to estimate mean position, motion magnitude, motion correlation, and trajectory of a tumor from cone-beam CT projections for image-guided radiotherapy.

PURPOSE To develop a probability-based method for estimating the mean position, motion magnitude, and trajectory of a tumor using cone-beam CT (CBCT) projections. METHOD AND MATERIALS CBCT acquisition was simulated for more than 80 hours of patient-measured trajectories for thoracic/abdominal tumors and prostate. The trajectories were divided into 60-second segments for which CBCT was simulated by projecting the tumor position onto a rotating imager. Tumor (surrogate) visibility on all projections was assumed. The mean and standard deviation of the tumor position and motion correlation along the three axes were determined with maximum likelihood estimation based on the projection data, assuming a Gaussian spatial distribution. The unknown position component along the imager axis was approximated by its expectation value, determined by the Gaussian distribution. Transformation of the resulting three-dimensional position to patient coordinates provided the estimated trajectory. Two trajectories were experimentally investigated by CBCT acquisition of a phantom. RESULTS The root-mean-square error of the estimated mean position was 0.05 mm. The root-mean-square error of the trajectories was <1 mm in 99.1% of the thorax/abdomen cases and in 99.7% of the prostate cases. The experimental trajectory estimation agreed with the actual phantom trajectory within 0.44 mm in any direction. Clinical applicability was demonstrated by estimating the tumor trajectory for a pancreas cancer case. CONCLUSIONS A method for estimation of mean position, motion magnitude, and trajectory of a tumor from CBCT projections has been developed. The accuracy was typically much better than 1 mm. The method is applicable to motion-inclusive, respiratory-gated, and tumor-tracking radiotherapy.

[1]  Patrick A Kupelian,et al.  Observations on real-time prostate gland motion using electromagnetic tracking. , 2008, International journal of radiation oncology, biology, physics.

[2]  M. V. van Herk,et al.  Respiratory correlated cone beam CT. , 2005, Medical physics.

[3]  Ali Amer,et al.  Online adaptive radiotherapy of the bladder: small bowel irradiated-volume reduction. , 2006, International journal of radiation oncology, biology, physics.

[4]  Jan-Jakob Sonke,et al.  Automatic prostate localization on cone-beam CT scans for high precision image-guided radiotherapy. , 2005, International journal of radiation oncology, biology, physics.

[5]  Lech Papiez,et al.  On probabilistically defined margins in radiation therapy , 2006, Physics in medicine and biology.

[6]  Parag J. Parikh,et al.  Development of the 4D Phantom for patient-specific end-to-end radiation therapy QA , 2007, SPIE Medical Imaging.

[7]  T W Griffin,et al.  Cone-beam CT for radiotherapy applications. , 1995, Physics in medicine and biology.

[8]  Jan-Jakob Sonke,et al.  3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[9]  Jeffrey C. Lagarias,et al.  Convergence Properties of the Nelder-Mead Simplex Method in Low Dimensions , 1998, SIAM J. Optim..

[10]  T. Marchant,et al.  Measurement of inter and intra fraction organ motion in radiotherapy using cone beam CT projection images , 2008, Physics in medicine and biology.

[11]  Mark Oldham,et al.  Cone-beam-CT guided radiation therapy: A model for on-line application. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[12]  Paul J Keall,et al.  An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patients , 2008, Physics in medicine and biology.

[13]  Paul J Keall,et al.  Three-dimensional prostate position estimation with a single x-ray imager utilizing the spatial probability density , 2008, Physics in medicine and biology.

[14]  David A Jaffray,et al.  Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. , 2007, International journal of radiation oncology, biology, physics.

[15]  Timothy Solberg,et al.  Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy. , 2007, International journal of radiation oncology, biology, physics.