[Comparative investigations of new dental electrosurgical instruments].

11 electrosurgical units of several manufactures were analysed by physical measurements. Conventional units and those with constant output voltage demonstrate nearly identical behavior in their voltage- and power-characteristics at load-impedance occurring during dental electrosurgery. So far there are no advantages of simple voltage control. We did not find technical deficiencies concerning security in the modern units examined.