Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique.

PURPOSE This technical note quantifies the dose and image quality performance of a clinically available organ-dose-based tube current modulation (ODM) technique, using experimental and simulation phantom studies. The investigated ODM implementation reduces the tube current for the anterior source positions, without increasing current for posterior positions, although such an approach was also evaluated for comparison. METHODS Axial CT scans at 120 kV were performed on head and chest phantoms on an ODM-equipped scanner (Optima CT660, GE Healthcare, Chalfont St. Giles, England). Dosimeters quantified dose to breast, lung, heart, spine, eye lens, and brain regions for ODM and 3D-modulation (SmartmA) settings. Monte Carlo simulations, validated with experimental data, were performed on 28 voxelized head phantoms and 10 chest phantoms to quantify organ dose and noise standard deviation. The dose and noise effects of increasing the posterior tube current were also investigated. RESULTS ODM reduced the dose for all experimental dosimeters with respect to SmartmA, with average dose reductions across dosimeters of 31% (breast), 21% (lung), 24% (heart), 6% (spine), 19% (eye lens), and 11% (brain), with similar results for the simulation validation study. In the phantom library study, the average dose reduction across all phantoms was 34% (breast), 20% (lung), 8% (spine), 20% (eye lens), and 8% (brain). ODM increased the noise standard deviation in reconstructed images by 6%-20%, with generally greater noise increases in anterior regions. Increasing the posterior tube current provided similar dose reduction as ODM for breast and eye lens, increased dose to the spine, with noise effects ranging from 2% noise reduction to 16% noise increase. At noise equal to SmartmA, ODM increased the estimated effective dose by 4% and 8% for chest and head scans, respectively. Increasing the posterior tube current further increased the effective dose by 15% (chest) and 18% (head) relative to SmartmA. CONCLUSIONS ODM reduced dose in all experimental and simulation studies over a range of phantoms, while increasing noise. The results suggest a net dose/noise benefit for breast and eye lens for all studied phantoms, negligible lung dose effects for two phantoms, increased lung dose and/or noise for eight phantoms, and increased dose and/or noise for brain and spine for all studied phantoms compared to the reference protocol.

[1]  J. H. Hubbell,et al.  XCOM : Photon Cross Sections Database , 2005 .

[2]  J. Damilakis,et al.  The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. , 2005, Medical physics.

[3]  J. Winslow,et al.  Organ dose and inherent uncertainty in helical CT dosimetry due to quasiperiodic dose distributions. , 2011, Medical physics.

[4]  Shuai Leng,et al.  Radiation dose reduction to the breast in thoracic CT: comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current. , 2011, Medical physics.

[5]  W P Segars,et al.  Population of anatomically variable 4D XCAT adult phantoms for imaging research and optimization. , 2013, Medical physics.

[6]  M. Lungren,et al.  Radiation dose estimations to the thorax using organ-based dose modulation. , 2012, AJR. American journal of roentgenology.

[7]  J. Boone,et al.  Experimental validation of a method characterizing bow tie filters in CT scanners using a real-time dose probe. , 2011, Medical physics.

[8]  Hyung Sik Kim,et al.  Reduced radiation exposure of the female breast during low-dose chest CT using organ-based tube current modulation and a bismuth shield: comparison of image quality and radiation dose. , 2013, AJR. American journal of roentgenology.

[9]  S. Evans Catalogue of Diagnostic X-Ray Spectra and Other Data , 1998 .

[10]  M. Kalra,et al.  Techniques and applications of automatic tube current modulation for CT. , 2004, Radiology.

[11]  C. Claussen,et al.  Automated Computed Tomography Dose-Saving Algorithm to Protect Radiosensitive Tissues: Estimation of Radiation Exposure and Image Quality Considerations , 2012, Investigative radiology.

[12]  Cynthia H McCollough,et al.  Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms. , 2009, Medical physics.

[13]  Maria Zankl,et al.  Reducing radiation dose to selected organs by selecting the tube start angle in MDCT helical scans: a Monte Carlo based study. , 2009, Medical physics.

[14]  Shuai Leng,et al.  Dose reduction to anterior surfaces with organ-based tube-current modulation: evaluation of performance in a phantom study. , 2011, AJR. American journal of roentgenology.

[15]  K. Hopper,et al.  The breast: in-plane x-ray protection during diagnostic thoracic CT--shielding with bismuth radioprotective garments. , 1997, Radiology.

[16]  A database for estimating organ dose for coronary angiography and brain perfusion CT scans for arbitrary spectra and angular tube current modulation. , 2012, Medical physics.

[17]  M. Körner,et al.  Filtered back projection, adaptive statistical iterative reconstruction, and a model-based iterative reconstruction in abdominal CT: an experimental clinical study. , 2013, Radiology.

[18]  Willi A. Kalender,et al.  Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current-modulated CT examinations , 2008, European Radiology.

[19]  Adrian K. Dixon,et al.  Benefits and costs, an eternal balance , 2007 .

[20]  K. F. Eckerman,et al.  Specific absorbed fractions of energy at various ages from internal photon sources: 6, Newborn , 1987 .

[21]  Taly Gilat Schmidt,et al.  Reducing radiation dose to the female breast during CT coronary angiography: a simulation study comparing breast shielding, angular tube current modulation, reduced kV, and partial angle protocols using an unknown-location signal-detectability metric. , 2013, Medical physics.

[22]  Shuai Leng,et al.  Bismuth shielding, organ-based tube current modulation, and global reduction of tube current for dose reduction to the eye at head CT. , 2012, Radiology.

[23]  W A Kalender,et al.  Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. , 1999, Medical physics.