Measurement of patient setup errors using port films and a computer-aided graphical alignment tool.

Patient orientations were measured for 49 patients treated in the abdomen, chest, and pelvic regions over the course of 20 months. Setup errors were determined using a curve-matching graphical interface to compare digitized port films to digitized simulation films. Data representing both "initial patient setup" and "patient setup at treatment" are presented and compared. Data were sorted by anatomic area and analyzed both at the population level and on a patient-by-patient basis. For each population, setup errors were observed to be primarily random, with population standard deviations of 5-6 mm for each of three translations and 2-3 degrees for each of two rotations. Rotations about the patients' inferior-superior axes were not measured. For each site, correlations between translations and/or rotations were small. The results are consistent with those from previous studies. The data set is among the largest collected to date.

[1]  G T Chen,et al.  Online repositioning during treatment of the prostate: a study of potential limits and gains. , 1993, International journal of radiation oncology, biology, physics.

[2]  G Bentel,et al.  Positioning and immobilization of patients undergoing radiation therapy for Hodgkin's disease. , 1991, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[3]  R Mohan,et al.  The role of uncertainty analysis in treatment planning. , 1991, International journal of radiation oncology, biology, physics.

[4]  W. Press,et al.  Numerical Recipes: The Art of Scientific Computing , 1987 .

[5]  R K Ten Haken,et al.  A head immobilization system for radiation simulation, CT, MRI, and PET imaging. , 1991, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[6]  M van Herk,et al.  Automatic verification of radiation field shape using digital portal images. , 1992, Medical physics.

[7]  D L McShan,et al.  A computer-controlled conformal radiotherapy system. I: Overview. , 1995, International journal of radiation oncology, biology, physics.

[8]  J Bijhold,et al.  Three-dimensional verification of patient placement during radiotherapy using portal images. , 1993, Medical physics.

[9]  P. Munro,et al.  A review of electronic portal imaging devices (EPIDs). , 1992, Medical physics.

[10]  F. V. Heuvel,et al.  Clinical implementation of an objective computer-aided protocol for intervention in intra-treatment correction using electronic portal imaging , 1995 .

[11]  M van Herk,et al.  Automatic on-line inspection of patient setup in radiation therapy using digital portal images. , 1993, Medical physics.

[12]  A G Haus,et al.  The effect of immobilisation on localisation error in the radiotherapy of head and neck cancer. , 1976, Clinical radiology.

[13]  A Fenster,et al.  Daily monitoring and correction of radiation field placement using a video-based portal imaging system: a pilot study. , 1992, International journal of radiation oncology, biology, physics.

[14]  G J Kutcher,et al.  The effect of setup uncertainties on the treatment of nasopharynx cancer. , 1993, International journal of radiation oncology, biology, physics.

[15]  A G Visser,et al.  Accuracy in radiation field alignment in head and neck cancer: a prospective study. , 1988, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[16]  A G Visser,et al.  Quality assurance using portal imaging: the accuracy of patient positioning in irradiation of breast cancer. , 1993, International journal of radiation oncology, biology, physics.

[17]  M. Goitein,et al.  Accuracy of radiation field alignment in clinical practice. , 1985, International journal of radiation oncology, biology, physics.

[18]  G T Chen,et al.  Correlation of projection radiographs in radiation therapy using open curve segments and points. , 1992, Medical physics.

[19]  A Bel,et al.  Time trend of patient setup deviations during pelvic irradiation using electronic portal imaging. , 1993, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.