Fraction of Inspired Oxygen With Low-Flow Versus High-Flow Devices: A Simulation Study

Purpose: The fraction of inspired oxygen while administering oxygen to patients must be measured as it represents the alveolar oxygen concentration, which is important from a respiratory physiology viewpoint. Therefore, the purpose of this study was to compare the fractions of inspired oxygen obtained through different oxygen delivery devices. Methods: A simulation model of spontaneous respiration was used. The fractions of inspired oxygen obtained through low- and high-flow nasal cannulas and a simple oxygen mask were measured. The fraction of inspired air was measured every second for 30 s after 120 s of oxygen administration. This was measured three times under each condition. Results: With a low-flow nasal cannula, airflow reduced both the intratracheal fraction of inspired oxygen and extraoral oxygen concentration, indicating that exhalatory respiration occurred during rebreathing and may be involved in increasing the intratracheal fraction of inspired oxygen. Conclusion: Oxygen administration during expiratory flow may lead to an increased oxygen concentration in the anatomical dead space, which may be involved in the increase in the fraction of inspired oxygen. With a high-flow nasal cannula, a high fraction of inspired oxygen can be achieved even at a flow rate of 10 L/min. When determining the optimum amount of oxygen, it is necessary to set an appropriate flow rate for patients and specific conditions without being bound by the fraction of inspired oxygen values alone. It might be difficult to estimate the fraction of inspired oxygen while using a low-flow nasal cannula and simple oxygen mask in clinical situations.

[1]  B. Bayram,et al.  Impact of High-Flow Nasal Cannula on Arterial Blood Gas Parameters in the Emergency Department , 2020, Cureus.

[2]  S. Rezaie,et al.  High-flow Nasal Cannula: Mechanisms of Action and Adult and Pediatric Indications , 2018, Cureus.

[3]  F. Rapetti,et al.  Reclassifying Acute Respiratory Distress Syndrome , 2018, American journal of respiratory and critical care medicine.

[4]  J. Orr,et al.  Efficiency of oxygen delivery through different oxygen entrainment devices during sedation under low oxygen flow rate: a bench study , 2018, Journal of Clinical Monitoring and Computing.

[5]  J. Soar,et al.  BTS guideline for oxygen use in adults in healthcare and emergency settings , 2017, Thorax.

[6]  C. Pagel,et al.  PaO2/FIO2 Ratio Derived From the SpO2/FIO2 Ratio to Improve Mortality Prediction Using the Pediatric Index of Mortality-3 Score in Transported Intensive Care Admissions* , 2017, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[7]  B. Goudra,et al.  Goudra ventilating bite block to reduce hypoxemia during endoscopic retrograde cholangiopancreatography , 2014, Saudi journal of anaesthesia.

[8]  A. Dahan,et al.  High-inspired oxygen concentration further impairs opioid-induced respiratory depression. , 2013, British journal of anaesthesia.

[9]  I. Cheifetz,et al.  Use of high flow nasal cannula in critically ill infants, children, and adults: a critical review of the literature , 2013, Intensive Care Medicine.

[10]  Arthur S Slutsky,et al.  Acute Respiratory Distress Syndrome The Berlin Definition , 2012 .

[11]  H. Hockey,et al.  Evaluation of a Humidified Nasal High-Flow Oxygen System, Using Oxygraphy, Capnography and Measurement of Upper Airway Pressures , 2011, Anaesthesia and intensive care.

[12]  T. Shaffer,et al.  High‐flow nasal cannula: Impact on oxygenation and ventilation in an acute lung injury model , 2011, Pediatric pulmonology.

[13]  J. Kernick,et al.  What is the evidence for the use of high flow nasal cannula oxygen in adult patients admitted to critical care units? A systematic review. , 2010, Australian critical care : official journal of the Confederation of Australian Critical Care Nurses.

[14]  T. Shaffer,et al.  Research in high flow therapy: mechanisms of action. , 2009, Respiratory medicine.

[15]  R. Beasley,et al.  Is it time to change the approach to oxygen therapy in the breathless patient? , 2007, Thorax.

[16]  N. Soni,et al.  Performance of six types of oxygen delivery devices at varying respiratory rates * , 2007, Anaesthesia.

[17]  Robert A. Smith,et al.  Supplemental oxygen impairs detection of hypoventilation by pulse oximetry. , 2004, Chest.

[18]  John B Downs,et al.  Has oxygen administration delayed appropriate respiratory care? Fallacies regarding oxygen therapy. , 2003, Respiratory care.

[19]  S. M. Kinsella,et al.  100% inspired oxygen from a Hudson mask-a realistic goal? , 2003, Resuscitation.

[20]  T. Waldau,et al.  Evaluation of five oxygen delivery devices in spontaneously breathing subjects by oxygraphy , 1998, Anaesthesia.

[21]  V. Larsen,et al.  The effect of nasal oxygen flow and catheter position on the accuracy of end‐tidal carbon dioxide measurements by a pharyngeal catheter in unintubated, spontaneously breathing subjects , 1995, Anaesthesia.

[22]  J. Leigh Variation in performance of oxygen therapy devices , 1970, Anaesthesia.

[23]  Keith D. Lamb,et al.  Southmedic OxyMaskTM compared with the Hudson RCI® Non-Rebreather MaskTM: Safety and performance comparison , 2016, Canadian journal of respiratory therapy : CJRT = Revue canadienne de la therapie respiratoire : RCTR.

[24]  J. Winck,et al.  Clinical evidence on high flow oxygen therapy and active humidification in adults , 2017 .