Performance of a 200‐MHz cryogenic RF probe designed for MRI and MRS of the murine brain

A cryogenic 200‐MHz RF transmit/receive probe (CryoProbe) made of copper and designed for murine brain studies is described. The probe operates at 30K and its performance was compared to a home‐made surface coil of equal dimensions and a mouse head volume resonator, both operating at room temperature (RT). Since for small volumes of tissue interrogated (<5 mL) sample noise and thermal noise of the receiver are of comparable magnitude, considerable noise reduction can be achieved by cooling of the receiver system, e.g., using cold helium gas. The effect of temperature lowering on the signal‐to‐noise ratio (SNR) was assessed in phantom and in vivo murine brain studies using conventional MRI sequences (spin and gradient echo) and single voxel MR spectroscopy. Consistent sensitivity increases were observed for the CryoProbe. SNR increases by a factor ≥2 for the phantom experiments and by factors of 1.8 to 1.9 for in vivo experiments when compared to the RT surface coil for both MRI and MRS experiments. These results are in good agreement with estimated SNR gains for a circular surface coil probe cooled to 30K. The use of a cryogenic probe represents an economically attractive option to enhance the sensitivity in small animal MRI/MRS. Magn Reson Med, 2008. © 2008 Wiley‐Liss, Inc.

[1]  J. Johnson Thermal Agitation of Electricity in Conductors , 1927, Nature.

[2]  A. S. Hall,et al.  Use of high temperature superconductor in a receiver coil for magnetic resonance imaging , 1991, Magnetic resonance in medicine.

[3]  J Hennig,et al.  RARE imaging: A fast imaging method for clinical MR , 1986, Magnetic resonance in medicine.

[4]  L Darrasse,et al.  Perspectives with cryogenic RF probes in biomedical MRI. , 2003, Biochimie.

[5]  R. Kirschman,et al.  Potential benefits of a cryogenically cooled NMR probe for room-temperature samples , 1989 .

[6]  A. Haase,et al.  A superconducting probehead applicable for nuclear magnetic resonance microscopy at 7 T , 1998 .

[7]  R. Gruetter,et al.  In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time , 1999, Magnetic resonance in medicine.

[8]  M. C. Cheng,et al.  Performance of large-size Superconducting coil in 0.21T MRI system , 2004, IEEE Transactions on Biomedical Engineering.

[9]  A Mogro-Campero,et al.  A high-temperature superconducting receiver for nuclear magnetic resonance microscopy. , 1993, Science.

[10]  L DelaBarre,et al.  The return of the frequency sweep: designing adiabatic pulses for contemporary NMR. , 2001, Journal of magnetic resonance.

[11]  W. Happer,et al.  Biological magnetic resonance imaging using laser-polarized 129Xe , 1994, Nature.

[12]  J R MacFall,et al.  Performance of a high‐temperature superconducting probe for in vivo microscopy at 2.0 T , 1999, Magnetic resonance in medicine.

[13]  Felix W. Wehrli,et al.  In vivo MR micro imaging with conventional radiofrequency coils cooled to 77°K , 2000 .

[14]  P. Lauterbur,et al.  The sensitivity of the zeugmatographic experiment involving human samples , 1979 .

[15]  W. Edelstein,et al.  The intrinsic signal‐to‐noise ratio in NMR imaging , 1986, Magnetic resonance in medicine.

[16]  L. Hedlund,et al.  MR Imaging with Hyperpolarized 3He Gas , 1995, Magnetic resonance in medicine.

[17]  G. Johnson,et al.  Performance of a high-temperature superconducting resonator for high-field imaging , 1995 .

[18]  P. Bottomley Spatial Localization in NMR Spectroscopy in Vivo , 1987, Annals of the New York Academy of Sciences.

[19]  G A Johnson,et al.  A high‐temperature superconducting Helmholtz probe for microscopy at 9.4 T , 1999, Magnetic resonance in medicine.

[20]  F David Doty,et al.  Radio frequency coil technology for small‐animal MRI , 2007, NMR in biomedicine.

[21]  P Crozat,et al.  High‐temperature superconducting surface coil for in vivo microimaging of the human skin , 2001, Magnetic resonance in medicine.

[22]  S. Crozier,et al.  Progress in high field MRI at the University of Florida , 2001, Magnetic Resonance Materials in Physics, Biology and Medicine.

[23]  P Wach,et al.  Imaging of the active B1 field in vivo , 1996, Magnetic resonance in medicine.

[24]  L. C. Bourne,et al.  Thin film high temperature superconducting RF coils for low field MRI , 1994, Magnetic resonance in medicine.