Comparing two heat and moisture exchangers with one vaporizing humidifier in patients with minute ventilation greater than 10 L/min.

STUDY OBJECTIVE To evaluate in patients submitted to minute ventilation > 10 L/min the ability to preserve patients' heat and humidity of two heat and moisture exchangers (HMEs) and one vaporizing humidifier (VH). DESIGN Prospective, randomized, comparative, non-blinded study. SETTING Intensive care unit of a university hospital. PATIENTS Nine tracheally intubated, mechanically ventilated patients, sedated and submitted to mechanical ventilation with minute ventilation > 10 L/min. INTERVENTIONS Using the psychrometric method, relative humidity (RH) and absolute humidity (AH) of inspired gas were obtained as well as temperature of inspired gas and tracheal temperatures (maximal and minimal). Following a randomized order, each patient was ventilated for two 24-h periods with a vaporizing humidifier (Bennett Cascade 2, Bennett; France) and one of two HMEs: Pall Ultipor filter BB50 (Pall Biomedical; France) or DAR Hygroster filter (Peters; France). Both were first tested for a 45-min period and then the HME that achieved the best performance in terms of temperature and water preservation was tested for 24 h. MEASUREMENTS AND RESULTS During the 45-min test period, the Pall Ultipor HME achieved a lower performance than the other two systems for any of the studied parameters (p < 0.05 to p < 0.0001). The DAR Hygroster HME achieved lower temperature of inspired gas (29.9 vs 32.0 degrees C, p < 0.005) and lower absolute humidity (29.3 vs 33.2 mg H2O/L, p < 0.005) than the Bennett Cascade 2. After 24 h of use, lower values of temperature of inspired gas (28.5 vs 32.0 degrees C, p < 0.002) and of AH (28.0 vs 33.6 mg H2O/L, p < 0.001) were obtained with the DAR Hygroster HME than with the Bennett Cascade 2. No differences were found between the two systems for the other tested parameters. At that time, no patients had RH lower than 97% and absolute humidity lower than 23 mg H2O/L with the use of the DAR Hygroster HME. CONCLUSIONS In patients with minute ventilation > 10 L/min, the DAR Hygroster HME showed a thermic and humidification capability similar to the reference system, the Bennett Cascade 2 VH. In these patients, the Pall Ultipor HME had a significantly lower capability.

[1]  S INGELSTEDT,et al.  Studies on the conditioning of air in the respiratory tract. , 1956, Acta oto-laryngologica. Supplementum.

[2]  R. Déry,et al.  Humidity in anaesthesiology III. Heat and moisture patterns in the respiratory tract during anaesthesia with the semi-closed system , 1967, Canadian Anaesthetists' Society journal.

[3]  R. A. Epstein Humidification during positive-pressure ventilation of infants. , 1971, Anesthesiology.

[4]  J. Chalon,et al.  Effects of Dry Anesthetic Gases on Tracheobronchial Ciliated Epithelium , 1972, Anesthesiology.

[5]  T. Howells,et al.  Humidification in anaesthesia. A review of the present situation. , 1972, British journal of anaesthesia.

[6]  H. Noguchi,et al.  A study of humidification in tracheostomized dogs. , 1973, British journal of anaesthesia.

[7]  R. Déry The evolution of heat and moisture in the respiratory tract during anaesthesia with a non-rebreathing system , 1973, Canadian Anaesthetists' Society journal.

[8]  A. R. Forbes Humidification and mucus flow in the intubated trachea. , 1973, British journal of anaesthesia.

[9]  A. R. Forbes Temperature, humidity and mucus flow in the intubated trachea. , 1974, British journal of anaesthesia.

[10]  Weeks Db Humidification during anesthesia. , 1975 .

[11]  P. Donahoe,et al.  Effect of dry and humidified gases on the respiratory epithelium in rabbits. , 1975, Journal of pediatric surgery.

[12]  D. Bethune,et al.  A comparative study of condenser humidifiers , 1976, Anaesthesia.

[13]  T. Tsuda,et al.  OPTIMUM HUMIDIFICATION OF AIR ADMINISTERED TO A TRACHEOSTOMY IN DOGS Scanning electron microscopy and surfactant studies , 1977 .

[14]  H. Turndorf,et al.  Humidity and the anesthetized patient. , 1979 .

[15]  J. Chalon,et al.  Humidification of anesthetic gases , 1981 .

[16]  J. Downs,et al.  Adult Body Temperature and Heated Humidification of Anesthetic Gases during General Anesthesia , 1981, Anesthesia and analgesia.

[17]  D. B. Weeks,et al.  Laboratory Investigation of Six Artificial Noses for Use during Endotracheal Anesthesia , 1983, Anesthesia and analgesia.

[18]  C. Membius A Comparative Evaluation of Disposable Humidifiers , 1983 .

[19]  H. Turndorf,et al.  The Pall Ultipor Breathing Circuit Filter—An Efficient Heat and Moisture Exchanger , 1984, Anesthesia and analgesia.

[20]  L. Triner,et al.  Halothane Inhibition of Canine Myocardial Adenylate Cyclase‐Modulation by Endogenous Factors , 1984, Anesthesia and analgesia.

[21]  D. Bethune Hydrophobic versus hygroscopic heat and moisture exchangers , 1985, Anaesthesia.

[22]  H. F. Bowman,et al.  Thermal mapping of the airways in humans. , 1985, Journal of applied physiology.

[23]  R. Latimer,et al.  A comparison of five heat and moisture exchangers , 1986, Anaesthesia.

[24]  M. Turtle,et al.  An Evaluation of Six Disposable Heat and Moisture Exchangers , 1987, Anaesthesia and intensive care.

[25]  I. L. Cohen,et al.  Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit. , 1988, Critical care medicine.

[26]  B. Eckerbom,et al.  Performance evaluation of six heat and moisture exchangers according to the Draft International Standard (ISO/DIS 9360) , 1990, Acta anaesthesiologica Scandinavica.

[27]  B. Eckerbom,et al.  Laboratory evaluation of heat and moisture exchangers. Assessment of the Draft International Standard (ISO/DIS 9360) in practice , 1990, Acta anaesthesiologica Scandinavica.

[28]  G. Perrin,et al.  Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit. , 1990, Chest.