Representation Method for Spectrally Overlapping Signals in Flow Cytometry Based on Fluorescence Pulse Time-Delay Estimation

Flow cytometry is being applied more extensively because of the outstanding advantages of multicolor fluorescence analysis. However, the intensity measurement is susceptible to the nonlinearity of the detection method. Moreover, in multicolor analysis, it is impossible to discriminate between fluorophores that spectrally overlap; this influences the accuracy of the fluorescence pulse signal representation. Here, we focus on spectral overlap in two-color analysis, and assume that the fluorescence follows the single exponential decay model. We overcome these problems by analyzing the influence of the spectral overlap quantitatively, which enables us to propose a method of fluorescence pulse signal representation based on time-delay estimation (between fluorescence and scattered pulse signals). First, the time delays are estimated using a modified chirp Z-transform (MCZT) algorithm and a fine interpolation of the correlation peak (FICP) algorithm. Second, the influence of hardware is removed via calibration, in order to acquire the original fluorescence lifetimes. Finally, modulated signals containing phase shifts associated with these lifetimes are created artificially, using a digital signal processing method, and reference signals are introduced in order to eliminate the influence of spectral overlap. Time-delay estimation simulation and fluorescence signal representation experiments are conducted on fluorescently labeled cells. With taking the potentially overlap of autofluorescence as part of the observed fluorescence spectrum, rather than distinguishing the individual influence, the results show that the calculated lifetimes with spectral overlap can be rectified from 8.28 and 4.86 ns to 8.51 and 4.63 ns, respectively, using the comprehensive approach presented in this work. These values agree well with the lifetimes (8.48 and 4.67 ns) acquired for cells stained with single-color fluorochrome. Further, these results indicate that the influence of spectral overlap can be eliminated effectively. Moreover, modulation, mixing with reference signals, and low-pass filtering are performed with a digital signal processing method, thereby obviating the need for a high-speed analog device and complex circuit system. Finally, the flexibility of the comprehensive method presented in this work is significantly higher than that of existing methods.

[1]  Mark A Naivar,et al.  Digital analysis and sorting of fluorescence lifetime by flow cytometry , 2010, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[2]  J. Lakowicz Principles of fluorescence spectroscopy , 1983 .

[3]  Ke Zhang,et al.  Phase-selective fluorescence of doped Ge2Sb2Te5 phase-change memory thin films , 2015 .

[4]  Fu-Jen Kao,et al.  Metabolism of HeLa cells revealed through autofluorescence lifetime upon infection with enterohemorrhagic Escherichia coli , 2012, Journal of biomedical optics.

[5]  J. Lakowicz,et al.  Phase-resolved fluorescence lifetime measurements for flow cytometry. , 1993, Cytometry.

[6]  Victor Chernomordik,et al.  In Vivo Fluorescence Lifetime Imaging for Monitoring the Efficacy of the Cancer Treatment , 2014, Clinical Cancer Research.

[7]  Jessica P. Houston,et al.  Cytometric sorting based on the fluorescence lifetime of spectrally overlapping signals. , 2013, Optics express.

[8]  张科,et al.  Phase-selective fluorescence of doped Ge2Sb2Te5 phase-change memory thin films , 2015 .

[9]  Ammasi Periasamy,et al.  Characterization of two‐photon excitation fluorescence lifetime imaging microscopy for protein localization , 2004, Microscopy research and technique.

[10]  Dahe Liu,et al.  Carpet anti-cloak based on transformation optics , 2014 .

[11]  Patrick Jenkins,et al.  Subcellular localization-dependent changes in EGFP fluorescence lifetime measured by time-resolved flow cytometry. , 2013, Biomedical optics express.

[12]  J A Steinkamp,et al.  Phase-sensitive detection methods for resolving fluorescence emission signals and directly quantifying lifetime. , 1994, Methods in cell biology.

[13]  Jimmie D. Parson,et al.  Flow cytometric time-resolved measurements by frequency heterodyning of fluorescence emission signals , 2001 .

[14]  J A Steinkamp,et al.  Fluorescence lifetime measurements in a flow cytometer by amplitude demodulation using digital data acquisition technique. , 1994, Cytometry.

[15]  Hans Georg Breunig,et al.  Two-photon spectral fluorescence lifetime and second-harmonic generation imaging of the porcine cornea with a 12-femtosecond laser microscope , 2016, Journal of biomedical optics.

[16]  N A Zilmer,et al.  Flow cytometric analysis using digital signal processing. , 1995, Cytometry.

[17]  Mark A Naivar,et al.  Expanding the potential of standard flow cytometry by extracting fluorescence lifetimes from cytometric pulse shifts , 2014, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[18]  J A Steinkamp,et al.  Resolution of fluorescence signals from cells labeled with fluorochromes having different lifetimes by phase-sensitive flow cytometry. , 1993, Cytometry.

[19]  J A Steinkamp,et al.  Time-resolved fluorescence-decay measurement and analysis on single cells by flow cytometry. , 1996, Applied optics.

[20]  Ramesh Raskar,et al.  Blind and reference-free fluorescence lifetime estimation via consumer time-of-flight sensors , 2015 .