Acousto-electrical speckle pattern in Lorentz force electrical impedance tomography.

Ultrasound speckle is a granular texture pattern appearing in ultrasound imaging. It can be used to distinguish tissues and identify pathologies. Lorentz force electrical impedance tomography is an ultrasound-based medical imaging technique of the tissue electrical conductivity. It is based on the application of an ultrasound wave in a medium placed in a magnetic field and on the measurement of the induced electric current due to Lorentz force. Similarly to ultrasound imaging, we hypothesized that a speckle could be observed with Lorentz force electrical impedance tomography imaging. In this study, we first assessed the theoretical similarity between the measured signals in Lorentz force electrical impedance tomography and in ultrasound imaging modalities. We then compared experimentally the signal measured in both methods using an acoustic and electrical impedance interface. Finally, a bovine muscle sample was imaged using the two methods. Similar speckle patterns were observed. This indicates the existence of an 'acousto-electrical speckle' in the Lorentz force electrical impedance tomography with spatial characteristics driven by the acoustic parameters but due to electrical impedance inhomogeneities instead of acoustic ones as is the case of ultrasound imaging.

[1]  M. Unser,et al.  A bulk modulus dependent linear model for acoustical imaging. , 2009, Journal of the Acoustical Society of America.

[2]  G. Cloutier,et al.  A critical review and uniformized representation of statistical distributions modeling the ultrasound echo envelope. , 2010, Ultrasound in medicine & biology.

[3]  E. Feleppa,et al.  Relationship of Ultrasonic Spectral Parameters to Features of Tissue Microstructure , 1987, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[4]  Bradley J Roth,et al.  The role of magnetic forces in biology and medicine , 2011, Experimental biology and medicine.

[5]  Tadashi Yamaguchi,et al.  The Quantitative Ultrasound Diagnosis of Liver Fibrosis Using Statistical Analysis of the Echo Envelope , 2013 .

[6]  Y. Xu,et al.  Magneto-acousto-electrical tomography: a potential method for imaging current density and electrical impedance. , 2008, Physiological measurement.

[7]  Bin He,et al.  Magnetoacoustic tomography with magnetic induction (MAT-MI) , 2005, Physics in medicine and biology.

[8]  Jeffrey C. Bamber,et al.  Physical Principles of Medical Ultrasonics: Hill/Physical Principles of Medical Ultrasonics, Second Edition , 2005 .

[9]  J Jossinet,et al.  Scanning Electric Conductivity Gradients with Ultrasonically-Induced Lorentz Force , 2001, Ultrasonic imaging.

[10]  R. Cobbold Foundations of Biomedical Ultrasound , 2006 .

[11]  S. K. Jespersen,et al.  Multi-Angle Compound Imaging , 1998, Ultrasonic imaging.

[12]  B.A. Baertlein,et al.  Authors' reply [to comments on "Theoretical model for an MRI radio frequency resonator"] , 1994, IEEE Transactions on Biomedical Engineering.

[13]  R. F. Wagner,et al.  Statistics of Speckle in Ultrasound B-Scans , 1983, IEEE Transactions on Sonics and Ultrasonics.

[14]  J. Shah,et al.  Hall effect imaging , 1998, IEEE Transactions on Biomedical Engineering.

[15]  Jeffrey C. Bamber,et al.  Physical principles of medical ultrasonics , 2004 .

[16]  Martin S. Roden Analog and Digital Communication Systems , 1979 .

[17]  Habib Ammari,et al.  Mathematical models and reconstruction methods in magneto-acoustic imaging , 2009, European Journal of Applied Mathematics.

[18]  M. Oelze,et al.  State of the Art Methods for Estimating Backscatter Coefficients , 2013 .

[19]  Jean-Yves Chapelon,et al.  Lorentz force electrical impedance tomography , 2013, 1402.2573.

[20]  R. F. Wagner,et al.  Describing small-scale structure in random media using pulse-echo ultrasound. , 1990, The Journal of the Acoustical Society of America.

[21]  J. Alison Noble,et al.  Ultrasound image segmentation: a survey , 2006, IEEE Transactions on Medical Imaging.

[22]  J. Ophir,et al.  Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues , 1991, Ultrasonic imaging.

[23]  F. L. Thurstone,et al.  Acoustic Speckle: Theory and Experimental Analysis , 1979 .

[24]  J C Bamber,et al.  Ultrasonic B-scanning: a computer simulation , 1980, Physics in medicine and biology.

[25]  Guy Cloutier,et al.  Review of Envelope Statistics Models for Quantitative Ultrasound Imaging and Tissue Characterization. , 2023, Advances in experimental medicine and biology.

[26]  E. Feleppa,et al.  Theoretical framework for spectrum analysis in ultrasonic tissue characterization. , 1983, The Journal of the Acoustical Society of America.

[27]  W. O’Brien Single-element transducers. , 1993, Radiographics : a review publication of the Radiological Society of North America, Inc.

[28]  Peter Basser,et al.  A theoretical model for magneto-acoustic imaging of bioelectric currents , 1994, IEEE Transactions on Biomedical Engineering.

[29]  C. Burckhardt Speckle in ultrasound B-mode scans , 1978, IEEE Transactions on Sonics and Ultrasonics.

[30]  Jonathan Mamou,et al.  Quantitative Ultrasound in Soft Tissues , 2013, Springer Netherlands.

[31]  W.D. O'Brien,et al.  Current time-domain methods for assessing tissue motion by analysis from reflected ultrasound echoes-a review , 1993, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[32]  F L Thurstone,et al.  Acoustic speckle: theory and experimental analysis. , 1979, Ultrasonic imaging.

[33]  M. O’Donnell,et al.  Internal displacement and strain imaging using ultrasonic speckle tracking , 1994, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[34]  M.A. Lubinski,et al.  Speckle tracking methods for ultrasonic elasticity imaging using short-time correlation , 1999, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[35]  Laurent Seppecher,et al.  A mathematical and numerical framework for ultrasonically-induced Lorentz force electrical impedance tomography☆ , 2014 .

[36]  R. W. Lau,et al.  The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. , 1996, Physics in medicine and biology.

[37]  Bin He,et al.  Magnetoacoustic Tomography With Magnetic Induction: Bioimepedance Reconstruction Through Vector Source Imaging , 2013, IEEE Transactions on Medical Imaging.