SWIFTSIN: A High-Resolution Ion Isolation Waveform for the Miniaturized Linear Ion Trap Mass Spectrometer by Coarse to Fine Excitation.

To figure out the reason for the drawback of the stored waveform inverse Fourier transform (SWIFT) waveform and realize the high-resolution ion isolation on the miniaturized linear ion trap mass spectrometer, we studied the efficiency that ions can be excited under different excitation durations and amplitudes at different frequencies and compared the overlap ratios of the effective excitation frequency bandwidths of the adjacent ions. According to this, we proposed a new coarse-to-fine isolation waveform named SWIFTSIN. By superposing one or more sinusoidal waveforms on the SWIFT waveform and modulating the phases of the superposed sinusoidal waveforms, the generated SWIFTSIN waveform can achieve unit mass isolation on the miniaturized linear ion trap mass spectrometer without reducing the intensity of the target ion. The isolation ability of the SWIFTSIN waveform was verified by isolating a single isotope peak in the mixed samples.

[1]  R. Cooks,et al.  Miniature mass spectrometer–based point-of-care assay for cabotegravir and rilpivirine in whole blood , 2022, Analytical and Bioanalytical Chemistry.

[2]  James L. Smith,et al.  Paper spray ionization–high-resolution mass spectrometry (PSI-HRMS) of peroxide explosives in biological matrices , 2021, Analytical and Bioanalytical Chemistry.

[3]  B. P. Harp,et al.  High resolution mass spectral data from the analysis of copper chlorophylls and copper chlorophyll degradation products in bright green table olives , 2020, Data in brief.

[4]  Chad R. Weisbrod,et al.  Ultrahigh Resolution Ion Isolation by Stored Waveform Inverse Fourier Transform 21 T Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. , 2020, Analytical chemistry.

[5]  Quan Yu,et al.  Discontinuous Subatmospheric Pressure Interface Reduces the Gas Flow Effects on Miniature CAPI Mass Spectrometer. , 2020, Analytical chemistry.

[6]  J. Silberring,et al.  MINIATURIZATION IN MASS SPECTROMETRY. , 2019, Mass spectrometry reviews.

[7]  W. Xu,et al.  Pseudo MRM Mode on the "Brick" Mass Spectrometer Using the Grid-SWIFT Waveform. , 2019, Analytical chemistry.

[8]  Zelong Liu,et al.  Structural information of asphaltenes derived from petroleum vacuum residue and its hydrotreated product obtained by FT-ICR mass spectrometry with narrow ion isolation windows , 2018, Fuel.

[9]  W. Xu,et al.  Improving the Performances of a "Brick Mass Spectrometer" by Quadrupole Enhanced Dipolar Resonance Ejection from the Linear Ion Trap. , 2018, Analytical chemistry.

[10]  Jian Sun,et al.  Development of a low power miniature linear ion trap mass spectrometer with extended mass range. , 2017, The Review of scientific instruments.

[11]  Guojun Yao,et al.  Approach for Pesticide Residue Analysis for Metabolite Prothioconazole-desthio in Animal Origin Food. , 2017, Journal of agricultural and food chemistry.

[12]  R. Cooks,et al.  Ion isolation and multigenerational collision-induced dissociation using the inverse Mathieu q scan. , 2017, Rapid communications in mass spectrometry : RCM.

[13]  F. Ahsan,et al.  A highly sensitive LC-MS/MS method for concurrent determination of sildenafil and rosiglitazone in rat plasma. , 2016, Journal of pharmaceutical and biomedical analysis.

[14]  Z. Ouyang,et al.  Direct Analysis of Nonvolatile Chemical Compounds on Surfaces Using a Hand-Held Mass Spectrometer with Synchronized Discharge Ionization Function. , 2016, Analytical chemistry.

[15]  R. Cooks,et al.  Miniature and Fieldable Mass Spectrometers: Recent Advances. , 2016, Analytical chemistry.

[16]  Yuling Liu,et al.  Development and validation of a liquid chromatography-tandem mass spectroscopy method for simultaneous determination of (+)-(13aS)-deoxytylophorinine and its pharmacologically active 3-O-desmethyl metabolite in rat plasma. , 2015, Journal of pharmaceutical and biomedical analysis.

[17]  Scott A. McLuckey,et al.  A method for isolating ions in quadrupole ion traps using an excitation waveform generated by frequency modulation and mixing , 2015 .

[18]  T. Cataldi,et al.  Molecular formula analysis of fragment ions by isotope-selective collision-induced dissociation tandem mass spectrometry of pharmacologically active compounds. , 2014, Journal of mass spectrometry : JMS.

[19]  Brandon L. Barney,et al.  How far can ion trap miniaturization go? Parameter scaling and space-charge limits for very small cylindrical ion traps. , 2014, Journal of mass spectrometry : JMS.

[20]  P. Lescuyer,et al.  Quantitative mass spectrometry analysis of intact hemoglobin A2 by precursor ion isolation and detection. , 2013, Analytical chemistry.

[21]  A. Marshall,et al.  Elemental Composition Validation from Stored Waveform Inverse Fourier Transform (SWIFT) Isolation FT-ICR MS Isotopic Fine Structure , 2013, Journal of The American Society for Mass Spectrometry.

[22]  Y. Komatsu,et al.  Selected reaction monitoring by linear ion-trap mass spectrometry can effectively be applicable to simultaneous quantification of multiple peptides. , 2011, Biological & pharmaceutical bulletin.

[23]  T. Cataldi,et al.  Collision-induced dissociation of the A + 2 isotope ion facilitates glucosinolates structure elucidation by electrospray ionization-tandem mass spectrometry with a linear quadrupole ion trap. , 2010, Analytical chemistry.

[24]  Maw-rong Lee,et al.  Solid-phase extraction coupled with liquid chromatography-tandem mass spectrometry for determination of trace rosiglitazone in urine. , 2005, Journal of chromatography. A.

[25]  B. Thomson,et al.  MS3 using the collision cell of a tandem mass spectrometer system. , 2002, Rapid communications in mass spectrometry : RCM.

[26]  M. V. van Stipdonk,et al.  Multi-stage tandem mass spectrometry of metal cationized leucine enkephalin and leucine enkephalin amide. , 2002, Rapid communications in mass spectrometry : RCM.

[27]  R. Cooks,et al.  Selective Injection and Isolation of Ions in Quadrupole Ion Trap Mass Spectrometry Using Notched Waveforms Created Using the Inverse Fourier Transform , 1994 .

[28]  Scott A. McLuckey,et al.  Filtered noise field signals for mass-selective accumulation of externally formed ions in a quadrupole ion trap , 1994 .

[29]  P. J. Callahan,et al.  Simultaneous isolation of two different m/z ions in an ion‐trap mass spectrometer and their tandem mass spectra using filtered‐noise fields , 1993 .

[30]  R. Graham Cooks,et al.  Reduction of space‐charging in the quadrupole ion trap by sequential injection and simultaneous storage of positively and negatively charged ions , 1993 .

[31]  J. Schwartz,et al.  High resolution parent-ion selection/isolation using a quadrupole ion-trap mass spectrometer , 1992 .

[32]  S. Guan Optimal phase modulation in stored waveform inverse Fourier transform excitation for Fourier transform mass spectrometry. II. Magnitude spectrum smoothing , 1990 .

[33]  S. Guan,et al.  General phase modulation method for stored waveform inverse fourier transform excitation for fourier transform ion cyclotron resonance mass spectrometry , 1989 .