Quantitating Fluorescence Intensity From Fluorophores: Practical Use of MESF Values

The present work uses fluorescein as the model fluorophore and points out critical steps in the use of MESF (Molecules of Equivalent Soluble Fluorophores) values for quantitative flow cytometric measurements. It has been found that emission spectrum matching between a reference solution and an analyte and normalization by the corresponding extinction coefficient are required for quantifying fluorescence signals using flow cytometers. Because of the use of fluorescein, the pH value of the medium is also critical for accurate MESF assignments. Given that the emission spectrum shapes of microbead suspensions and stained biological cells are not significantly different, the percentage of error due to spectrum mismatch is estimated. We have also found that the emission spectrum of a microbead with a seven-methylene linker between the fluorescein and the bead surface (bead7) provides the best match with the spectra from biological cells. Therefore, bead7 is potentially a better calibration standard for flow cytometers than the existing one that is commercially available and used in the present study.

[1]  William H. Sawyer,et al.  Spectral properties of the prototropic forms of fluorescein in aqueous solution , 1996, Journal of Fluorescence.

[2]  Edward Early,et al.  Quantitating Fluorescence Intensity from Fluorophore: The Definition of MESF Assignment , 2002, Journal of research of the National Institute of Standards and Technology.

[3]  J V Giorgi,et al.  Quantitation of CD38 expression using QuantiBRITE beads. , 1998, Cytometry.

[4]  J V Giorgi,et al.  Quantitation of CD38 activation antigen expression on CD8+ T cells in HIV-1 infection using CD4 expression on CD4+ T lymphocytes as a biological calibrator. , 1998, Cytometry.

[5]  L O Henderson,et al.  Interlaboratory study of cellular fluorescence intensity measurements with fluorescein-labeled microbead standards. , 1991, Cytometry.

[6]  L O Henderson,et al.  Terminology and nomenclature for standardization in quantitative fluorescence cytometry. , 1998, Cytometry.

[7]  S L Jacques,et al.  Experimental tests of a simple diffusion model for the estimation of scattering and absorption coefficients of turbid media from time-resolved diffuse reflectance measurements. , 1992, Applied optics.

[8]  O. C. Blair,et al.  Practical Flow Cytometry , 1985, The Yale Journal of Biology and Medicine.

[9]  P. Renzi,et al.  Analysis of T cell subsets in normal adults. Comparison of whole blood lysis technique to Ficoll-Hypaque separation by flow cytometry. , 1987, Journal of immunological methods.

[10]  D. Seigneurin [Cytometry]. , 2020, Annales de Pathologie.

[11]  M. Kubista,et al.  Absorption and fluorescence properties of fluorescein , 1995 .

[12]  N. Micali,et al.  Separation of scattering and absorption contributions in UV/visible spectra of resonant systems. , 2001, Analytical chemistry.

[13]  E. Gibbs,et al.  Resonance Light Scattering and Its Application in Determining the Size, Shape, and Aggregation Number for Supramolecular Assemblies of Chromophores , 1999 .

[14]  H. Micklem,et al.  Right angle light scatter: a necessary parameter in flow cytofluorimetric analysis of human peripheral blood mononuclear cells. , 1983, Journal of immunological methods.

[15]  D. Dibbern-brunelli,et al.  Exciton splitting of fluorescein disodium in polyethylene glycol 600 , 1991 .

[16]  J. Fowler,et al.  Journal of Research of the National Institute of Standards and Technology INFORMATION TECHNOLOGY FOR ENGINEERING AND MANUFACTURING Gaithersburg , MD June 12-13 , 2000 , 2000 .

[17]  A Schwartz,et al.  Standardizing flow cytometry: construction of a standardized fluorescence calibration plot using matching spectral calibrators. , 1996, Cytometry.