Improved Center Array-Sequensing Phase Unwrapping(ICASPU) method for reconstruction of MR phase image

This study proposed an improved center array-sequencing phase unwrapping (ICASPU) algorithm. 2% agarose phantom dopped with 0.6mM/l MnCl2 was used with clinical 1.5T MRI system and commercial knee coil. Obtained k-space data(raw-data) was transmitted to PC and reconstructed into phase image with MATLAB software. Previous center array-sequence phase unwrapping algorithm wascompared with proposed ICASP algorithm using second order regression analysis. As a result, we found that the amount of error on proposed ICASPU method is less about 5 times than that of previous CASPU method. In this study, we exploit improved Center array-sequence phase unwrapping algorithm and expect to apply to images including phase informations.