Quantitative measurement of normal and hydrocephalic cerebrospinal fluid flow using phase contrast cine MR imaging.

Measurements of the cerebrospinal fluid (CSF) flow using phase contrast cine magnetic resonance (MR) imaging were performed on a phantom, 12 normal subjects and 20 patients with normal pressure hydrocephalus (NPH). The phantom study demonstrated the applicability of phase contrast in quantitative measurement of the slow flow. The CSF flows of the normal subjects showed a consistent pattern with a to-and-fro movement of the flow in the anterior subarachnoid space at the C2/3 level, and they were dependent on the cardiac cycle in all subjects. However, the patients with NPH showed variable patterns of the CSF pulsatile flow and these patterns could be divided into four types according to velocity and amplitude. The amplitudes of each type were as follows: type 0 (n = 1), 87.6mm; type I (n = 2), 58.2mm (mean); type II (n = 6), 48.0 +/- 5.0mm (mean +/- SEM); and type III (n = 11), 19.9 +/- 1.8mm (mean +/- SEM). The decrease of the amplitudes correlated to a worsening of the clinical symptoms. After the shunting operation, the amplitude of to-and-fro movement of the CSF increased again in the patients with NPH who improved clinically. Some of the type III cases were reclassified type II, I and 0 and also one of the type II cases changed type I after the shunting operation. We conclude that the phase contrast cine MR imaging is a practically and clinically applicable technique for the quantitative measurement of the CSF flow.