Multislice and multicoil phase‐sensitive inversion‐recovery imaging

Phase‐sensitive inversion‐recovery (PSIR) imaging may provide enhanced T1 contrast. However, clinical implementation of PSIR imaging is hindered because image reconstruction with this method often lacks robustness and requires manual intervention, particularly for data acquired in multiple slices and with phased‐array coils. In this paper, a new algorithm suitable for automatic PSIR image reconstruction of multislice and multicoil data is presented. This algorithm phase corrects by region‐growing, employing both the magnitude and the phase information of image pixels. Specifically, phase gradients of the original complex image are first calculated and then used to determine the sequence of the region‐growing. The signal direction relating to the phase error for each pixel is then determined during the region‐growing using both the magnitude and the phase of the previously determined pixels that are located within a boxcar neighborhood of the pixel. Finally, the intrinsic intercoil and interslice correlation is exploited to ensure consistency in the global polarity of all of the PSIR images. The results are demonstrated with in vivo human brain images acquired at 3 Tesla with an eight‐channel phased‐array coil. Magn Reson Med 53:904–910, 2005. © 2005 Wiley‐Liss, Inc.

[1]  G M Bydder,et al.  MR Imaging: Clinical Use of the Inversion Recovery Sequence , 1985, Journal of computer assisted tomography.

[2]  Jingfei Ma,et al.  Breath‐hold water and fat imaging using a dual‐echo two‐point dixon technique with an efficient and robust phase‐correction algorithm , 2004, Magnetic resonance in medicine.

[3]  D. Noll,et al.  Homodyne detection in magnetic resonance imaging. , 1991, IEEE transactions on medical imaging.

[4]  M O Leach,et al.  A simple method for the restoration of signal polarity in multi‐image inversion recovery sequences for measuring T1 , 1991, Magnetic resonance in medicine.

[5]  W. Perman,et al.  Improved detectability in low signal-to-noise ratio magnetic resonance images by means of a phase-corrected real reconstruction. , 1989, Medical physics.

[6]  P A Bottomley,et al.  Signal, Noise, and Contrast in Nuclear Magnetic Resonance (NMR) Imaging , 1983, Journal of computer assisted tomography.

[7]  S K Hilal,et al.  An optimized multislice acquisition sequence for the inversion-recovery MR imaging. , 1991, Magnetic resonance imaging.

[8]  A M Aisen,et al.  Regional phase correction of inversion‐recovery MR images , 1990, Magnetic resonance in medicine.

[9]  M. Zhang,et al.  Imaging pulmonary blood flow and perfusion using phase‐sensitive selective inversion recovery , 2000, Magnetic resonance in medicine.

[10]  Christoph Segebarth,et al.  Mapping of normal brain maturation in infants on phase-sensitive inversion-recovery MR images , 2004, Neuroradiology.

[11]  J R MacFall,et al.  Mechanisms of Contrast in NMR Imaging , 1984, Journal of computer assisted tomography.

[12]  E. McVeigh,et al.  Phase‐sensitive inversion recovery for detecting myocardial infarction using gadolinium‐delayed hyperenhancement † , 2002, Magnetic resonance in medicine.

[13]  J Listerud,et al.  OIL FLAIR: Optimized interleaved fluid‐attenuated inversion recovery in 2D fast spin echo , 1996, Magnetic resonance in medicine.

[14]  C. Ahn,et al.  A New Phase Correction Method in NMR Imaging Based on Autocorrelation and Histogram Analysis , 1987, IEEE Transactions on Medical Imaging.

[15]  Q. Xiang,et al.  Inversion recovery image reconstruction with multiseed region‐growing spin reversal , 1996, Journal of magnetic resonance imaging : JMRI.

[16]  P. Roemer,et al.  The NMR phased array , 1990, Magnetic resonance in medicine.

[17]  P. V. van Dijk,et al.  Restoration of Signal Polarity in a Set of Inversion Recovery NMR Images , 1984, IEEE Transactions on Medical Imaging.

[18]  M. Bronskill,et al.  Phase and sensitivity of receiver coils in magnetic resonance imaging. , 1986, Medical physics.

[19]  Z. Cho,et al.  Real‐value representation in inversion‐recovery NMR imaging by use of a phase‐correction method , 1986, Magnetic resonance in medicine.

[20]  Joseph V. Hajnal,et al.  Use of Fluid Attenuated Inversion Recovery (FLAIR) Pulse Sequences in MRI of the Brain , 1992, Journal of computer assisted tomography.