A navigation system for percutaneous needle interventions based on PET/CT images: Design, workflow and error analysis of soft tissue and bone punctures

Percutaneous needle intervention based on PET/CT images is effective, but exposes the patient to unnecessary radiation due to the increased number of CT scans required. Computer assisted intervention can reduce the number of scans, but requires handling, matching and visualization of two different datasets. While one dataset is used for target definition according to metabolism, the other is used for instrument guidance according to anatomical structures. No navigation systems capable of handling such data and performing PET/CT image-based procedures while following clinically approved protocols for oncologic percutaneous interventions are available. The need for such systems is emphasized in scenarios where the target can be located in different types of tissue such as bone and soft tissue. These two tissues require different clinical protocols for puncturing and may therefore give rise to different problems during the navigated intervention. Studies comparing the performance of navigated needle interventions targeting lesions located in these two types of tissue are not often found in the literature. Hence, this paper presents an optical navigation system for percutaneous needle interventions based on PET/CT images. The system provides viewers for guiding the physician to the target with real-time visualization of PET/CT datasets, and is able to handle targets located in both bone and soft tissue. The navigation system and the required clinical workflow were designed taking into consideration clinical protocols and requirements, and the system is thus operable by a single person, even during transition to the sterile phase. Both the system and the workflow were evaluated in an initial set of experiments simulating 41 lesions (23 located in bone tissue and 18 in soft tissue) in swine cadavers. We also measured and decomposed the overall system error into distinct error sources, which allowed for the identification of particularities involved in the process as well as highlighting the differences between bone and soft tissue punctures. An overall average error of 4.23 mm and 3.07 mm for bone and soft tissue punctures, respectively, demonstrated the feasibility of using this system for such interventions. The proposed system workflow was shown to be effective in separating the preparation from the sterile phase, as well as in keeping the system manageable by a single operator. Among the distinct sources of error, the user error based on the system accuracy (defined as the distance from the planned target to the actual needle tip) appeared to be the most significant. Bone punctures showed higher user error, whereas soft tissue punctures showed higher tissue deformation error.

[1]  Rajni V. Patel,et al.  Needle insertion into soft tissue: a survey. , 2007, Medical engineering & physics.

[2]  Barry A Siegel,et al.  Impact of positron emission tomography/computed tomography and positron emission tomography (PET) alone on expected management of patients with cancer: initial results from the National Oncologic PET Registry. , 2008, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[3]  T. Vogl,et al.  Navigation-Based Needle Puncture of a Cadaver Using a Hybrid Tracking Navigational System , 2006, Investigative radiology.

[4]  Jay B. West,et al.  Predicting error in rigid-body point-based registration , 1998, IEEE Transactions on Medical Imaging.

[5]  Willi A. Kalender,et al.  A Navigation System for Minimally Invasive CT-Guided Interventions , 2005, MICCAI.

[6]  S. Heiba,et al.  Impact of PET/CT in Comparison With Same Day Contrast Enhanced CT in Breast Cancer Management , 2007, Clinical nuclear medicine.

[7]  J. Wiskirchen,et al.  PET-CT-guided interventions in the management of FDG-positive lesions in patients suffering from solid malignancies: initial experiences , 2009, European Radiology.

[8]  Lena Maier-Hein,et al.  In vivo accuracy assessment of a needle-based navigation system for CT-guided radiofrequency ablation of the liver. , 2008, Medical physics.

[9]  Kevin Cleary,et al.  Implementation of an electromagnetic tracking system for accurate intrahepatic puncture needle guidance: accuracy results in an in vitro model. , 2007, Academic radiology.

[10]  N. Glossop,et al.  Advantages of optical compared with electromagnetic tracking. , 2009, The Journal of bone and joint surgery. American volume.

[11]  J. Krücker,et al.  Electromagnetic tracking for thermal ablation and biopsy guidance: clinical evaluation of spatial accuracy. , 2007, Journal of vascular and interventional radiology : JVIR.

[12]  K. Cleary,et al.  Navigation with electromagnetic tracking for interventional radiology procedures: a feasibility study. , 2005, Journal of vascular and interventional radiology : JVIR.

[13]  E.,et al.  A least-squares algorithm for the equiform transformation from spatial marker co-ordinates. , 1988, Journal of biomechanics.

[14]  Luc Soler,et al.  An augmented reality system for liver thermal ablation: Design and evaluation on clinical cases , 2009, Medical Image Anal..

[15]  K. Cleary,et al.  Image-guided interventions: technology review and clinical applications. , 2010, Annual review of biomedical engineering.