In vitro Wear Rate and Co Ion Release of Compositionally and Structurally Graded CoCrMo-Ti6Al4V Structures.

Novel, unitized structures with porous Ti6Al4V alloy on one side and compositionally graded, hard CoCrMo alloy surface on the other side have been fabricated using laser engineered net shaping (LENS™) process. Gradient structures with 50%, 70% and 86% CoCrMo alloy on the top surface showed high hardness in the range of 615 and 957 HV. The gradient structures were evaluated for their in vitro wear rate and Co release up to 3000 m of sliding distance. The wear rate of ultrahigh molecular weight polyethylene and 100% CoCrMo alloy substrates found to depend on the hardness and microstructural features of the counter surface rubbing against them. In general, the wear rate of both the substrates increased with a decrease in the CoCrMo alloy concentration on the top surface of gradient pins. However, the wear rate of gradient pins was lower than 100% CoCrMo alloy pins due to their high hardness. Lowest wear rate in the range of 5.07 to 7.99 × 10(-8) mm(3)/Nm was observed for gradient pins having 86% CoCrMo alloy on the top surface. The amount of Co released, in the range of 0.38 and 0.91 ppm, during in vitro wear testing of gradient structures was comparable to that of 100% CoCrMo alloy (0.25 and 0.77 ppm). Present unitized structures with open porosity on one side and hard, wear resistant surface on the other side can minimize the wear-induced osteolysis and aseptic loosening, and eliminate the need for multiple parts with different compositions for load-bearing implants such as total hip prostheses.

[1]  J. Cuckler The Rationale for Metal-on-Metal Total Hip Arthroplasty , 2005, Clinical orthopaedics and related research.

[2]  A. Lombardi,et al.  Large versus small femoral heads in metal-on-metal total hip arthroplasty. , 2004, The Journal of arthroplasty.

[3]  T. Bauer,et al.  Polyethylene debris-induced osteolysis and loosening in uncemented total hip arthroplasty. A cause of late failure. , 1992, The Journal of arthroplasty.

[4]  Vamsi Krishna Balla,et al.  Design and fabrication of CoCrMo alloy based novel structures for load bearing implants using laser engineered net shaping , 2010 .

[5]  Vamsi Krishna Balla,et al.  Fabrication of compositionally and structurally graded Ti-TiO2 structures using laser engineered net shaping (LENS). , 2009, Acta biomaterialia.

[6]  M. Ries,et al.  Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene , 2001, The Journal of bone and joint surgery. American volume.

[7]  L. J. Yang A test methodology for the determination of wear coefficient , 2005 .

[8]  P. Lachiewicz,et al.  Dislocation of Primary Total Hip Arthroplasty with 36 and 40-mm Femoral Heads , 2006, Clinical orthopaedics and related research.

[9]  R. Sierra,et al.  Dislocation of Primary THA Done through a Posterolateral Approach in the Elderly , 2005, Clinical orthopaedics and related research.

[10]  Richard E. White,et al.  Effect of Posterior Capsular Repair on Early Dislocation in Primary Total Hip Replacement , 2001, Clinical orthopaedics and related research.

[11]  L. Segal John , 2013, The Messianic Secret.

[12]  D. Bartel,et al.  The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. , 1986, The Journal of bone and joint surgery. American volume.

[13]  C. Rieker,et al.  In vivo tribological performance of 231 metal-on-metal hip articulations. , 2002, Hip international : the journal of clinical and experimental research on hip pathology and therapy.

[14]  N. Ravikumar,et al.  Prediction of wear coefficient of Al6061–TiO2 composites , 2005 .

[15]  Amit Bandyopadhyay,et al.  Functionally graded Co-Cr-Mo coating on Ti-6Al-4V alloy structures. , 2008, Acta biomaterialia.

[16]  A. Bandyopadhyay,et al.  Laser-assisted Zr/ZrO(2) coating on Ti for load-bearing implants. , 2009, Acta biomaterialia.

[17]  J. Parvizi,et al.  Metal-on-metal versus metal-on-polyethylene bearings in total hip arthroplasty , 2004 .

[18]  G. G. Stokes "J." , 1890, The New Yale Book of Quotations.

[19]  J. Parvizi,et al.  Metal-on-metal versus metal-on-polyethylene bearings in total hip arthroplasty: a matched case-control study. , 2004, The Journal of arthroplasty.

[20]  Michael Tanzer,et al.  Role of Capsular Repair on Dislocation in Revision Total Hip Arthroplasty , 2006, Clinical orthopaedics and related research.

[21]  Kevin Ong,et al.  The epidemiology of revision total hip arthroplasty in the United States. , 2009, The Journal of bone and joint surgery. American volume.

[22]  Clare M. Rimnac,et al.  Mechanical behavior, wear surface morphology, and clinical performance of UHMWPE acetabular components after 10 years of implantation , 2001 .

[23]  R. Poggie,et al.  Wear evaluation of cobalt-chromium alloy for use in a metal-on-metal hip prosthesis. , 2004, Journal of biomedical materials research. Part B, Applied biomaterials.