Experimental measurements and Monte Carlo simulations for dosimetric evaluations of intrafraction motion for gated and ungated intensity modulated arc therapy deliveries

Respiratory gated radiation therapy allows for a smaller margin expansion for the planning target volume (PTV) to account for respiratory induced motion and is emerging as a common method to treat lung and liver tumors. We investigated the dosimetric effect of free motion and gated delivery for intensity modulated arc therapy (IMAT) with experimental measurements and Monte Carlo simulations. The impact of PTV margin and duty cycle for gated delivery is studied with Monte Carlo simulations. A motion phantom is used for this study. Two sets of contours were drawn on the mid-inspiration CT scan of this motion phantom. For each set of contours, an IMAT plan to be delivered with constant dose rate was created. The plans were generated on a CT scan of the phantom in the static condition with 3 mm PTV margin and applied to the motion phantom under four conditions: static, full superior–inferior (SI) motion (A = 1 cm, T = 4 s) and gating conditions (25% and 50% duty cycles) with full SI motion. A 6 by 15 cm piece of radiographic film was placed in the sagittal plane of the phantom and then irradiated under all measurement conditions. Film calibration was performed with a step-wedge method to convert optical density to dose. Gated IMAT delivery was first validated in 2D by comparing static film with that from gating and full motion. A previously verified simulation tool for IMRT that takes the log files from the multileaf collimator (MLC) controller and the gating system were adapted to simulate the delivered IMAT treatment for full 3D dosimetric analysis. The IMAT simulations were validated against the 2D film measurements. The resultant IMAT simulations were evaluated with dose criteria, dose-volume histograms and 3D gamma analysis. We validated gated IMAT deliveries when we compared the static film with the one from gating using 25% duty cycle using 2D gamma analysis. Within experimental and setup uncertainties, film measurements agreed with their corresponding simulated plans using 2D gamma analysis. Finally, when planning with margins designed for gating with 25% duty cycle and applying 50% or no gating during treatment, the dose differences in Dmin, D99% and D95% of the clinical target volume can be up to 27 cGy, 20 cGy and 18 cGy, respectively, for a plan with 200 cGy prescription dose. We have experimentally delivered gated IMAT with constant dose rate to a motion phantom and assessed their accuracies with film dosimetry and Monte Carlo simulations. Film dosimetry demonstrated that 25% gating and static plans are within 2%, 2 mm. The Monte Carlo simulation method was employed to generate dose delivered in 3D to a motion phantom, and the dosimetric results were reported. Since our film measurements agreed well with Monte Carlo simulations, we can reliably use this simulation tool to further study the dosimetric effects of target motion and effectiveness of gating for IMAT deliveries.

[1]  Shuang Luan,et al.  Stochastic versus deterministic kernel-based superposition approaches for dose calculation of intensity-modulated arcs. , 2008, Physics in medicine and biology.

[2]  Eugene Wong,et al.  Monte Carlo dose calculation of segmental IMRT delivery to a moving phantom using dynamic MLC and gating log files. , 2008, Physics in medicine and biology.

[3]  Eugene Wong,et al.  Commissioning a fast Monte Carlo dose calculation algorithm for lung cancer treatment planning , 2008, Journal of applied clinical medical physics.

[4]  Joao Seco,et al.  Dosimetric impact of motion in free-breathing and gated lung radiotherapy: a 4D Monte Carlo study of intrafraction and interfraction effects. , 2007, Medical physics.

[5]  Karl Otto,et al.  Volumetric modulated arc therapy: IMRT in a single gantry arc. , 2007, Medical physics.

[6]  Eugene Wong,et al.  Incorporating geometric ray tracing to generate initial conditions for intensity modulated arc therapy optimization. , 2008, Medical physics.

[7]  Eugene Wong,et al.  Segmentation and leaf sequencing for intensity modulated arc therapy. , 2007, Medical Physics (Lancaster).

[8]  Jan-Jakob Sonke,et al.  A fast algorithm for gamma evaluation in 3D. , 2007, Medical physics.

[9]  Benjamin E Nelms,et al.  On the dose to a moving target while employing different IMRT delivery mechanisms. , 2007, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  H Paganetti,et al.  Effects of organ motion on IMRT treatments with segments of few monitor units. , 2007, Medical physics.

[11]  George Starkschall,et al.  Assessment of lung tumor motion and setup uncertainties using implanted fiducials. , 2007, International journal of radiation oncology, biology, physics.

[12]  Debra H Brinkmann,et al.  Incidence of radiation pneumonitis after thoracic irradiation: Dose-volume correlates. , 2007, International journal of radiation oncology, biology, physics.

[13]  M A Earl,et al.  An arc-sequencing algorithm for intensity modulated arc therapy. , 2007, Medical physics.

[14]  Indrin J Chetty,et al.  How extensive of a 4D dataset is needed to estimate cumulative dose distribution plan evaluation metrics in conformal lung therapy? , 2006, Medical physics.

[15]  Ke Sheng,et al.  A motion phantom study on helical tomotherapy: the dosimetric impacts of delivery technique and motion , 2006, Physics in medicine and biology.

[16]  Lech Papiez,et al.  Excessive toxicity when treating central tumors in a phase II study of stereotactic body radiation therapy for medically inoperable early-stage lung cancer. , 2006, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[17]  Steve B. Jiang Radiotherapy of mobile tumors. , 2006, Seminars in radiation oncology.

[18]  Radhe Mohan,et al.  Audio-visual biofeedback for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy. , 2006, International journal of radiation oncology, biology, physics.

[19]  Steve B. Jiang,et al.  Estimation of the delivered patient dose in lung IMRT treatment based on deformable registration of 4D-CT data and Monte Carlo simulations , 2006, Physics in medicine and biology.

[20]  Steve B. Jiang,et al.  Technical aspects of image-guided respiration-gated radiation therapy. , 2006, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[21]  Jun Duan,et al.  Dosimetric and radiobiological impact of dose fractionation on respiratory motion induced IMRT delivery errors: a volumetric dose measurement study. , 2006, Medical physics.

[22]  Steve B. Jiang,et al.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76. , 2006, Medical physics.

[23]  E. Wong,et al.  SU‐FF‐T‐102: Dosimetric Properties of Respiratory‐Gated Intensity Modulated Arc Treatments , 2005 .

[24]  Fang-Fang Yin,et al.  Time delay measurement for linac based treatment delivery in synchronized respiratory gating radiotherapy. , 2005, Medical physics.

[25]  Lei Dong,et al.  Dosimetric accuracy of Kodak EDR2 film for IMRT verifications. , 2005, Medical physics.

[26]  Margrit Betke,et al.  The correlation between internal and external markers for abdominal tumors: implications for respiratory gating. , 2003, International journal of radiation oncology, biology, physics.

[27]  George Starkschall,et al.  Evaluation of internal lung motion for respiratory-gated radiotherapy using MRI: Part I--correlating internal lung motion with skin fiducial motion. , 2004, International journal of radiation oncology, biology, physics.

[28]  C. Koch,et al.  In regard to Vordermark et al. (Int J Radiat Oncol Biol Phys 2004;58:1242-1250). , 2004, International journal of radiation oncology, biology, physics.

[29]  F. Khuri,et al.  The care of the lung cancer patient in the 21st century: a new age. , 2004, Seminars in oncology.

[30]  G S Bauman,et al.  Tracking the dose distribution in radiation therapy by accounting for variable anatomy , 2004, Physics in medicine and biology.

[31]  Eugene Wong,et al.  Comparison of dose calculation algorithms with Monte Carlo methods for photon arcs. , 2003, Medical physics.

[32]  J. Dempsey,et al.  Evaluation of the gamma dose distribution comparison method. , 2003, Medical physics.

[33]  Jun Duan,et al.  Dosimetric effect of respiration-gated beam on IMRT delivery. , 2003, Medical physics.

[34]  Steve B. Jiang,et al.  An experimental investigation on intra-fractional organ motion effects in lung IMRT treatments. , 2003, Physics in medicine and biology.

[35]  Nzhde Agazaryan,et al.  The effects of tumor motion on planning and delivery of respiratory-gated IMRT. , 2003, Medical physics.

[36]  Joseph O Deasy,et al.  CERR: a computational environment for radiotherapy research. , 2003, Medical physics.

[37]  D L McShan,et al.  Inclusion of organ deformation in dose calculations. , 2003, Medical physics.

[38]  Radhe Mohan,et al.  Patient training in respiratory-gated radiotherapy. , 2003, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[39]  D. G. Lewis,et al.  A DICOM-RT-based toolbox for the evaluation and verification of radiotherapy plans. , 2002, Physics in medicine and biology.

[40]  Nzhde Agazaryan,et al.  An evaluation of gating window size, delivery method, and composite field dosimetry of respiratory-gated IMRT. , 2002, Medical physics.

[41]  A. Olch Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance. , 2002, Medical physics.

[42]  M Saiful Huq,et al.  A dosimetric comparison of various multileaf collimators. , 2002, Physics in medicine and biology.

[43]  R K Ten Haken,et al.  Estimation of tumor control probability model parameters from 3-D dose distributions of non-small cell lung cancer patients. , 1999, Lung cancer.

[44]  Cedric X. Yu,et al.  Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. , 1995, Physics in medicine and biology.