A simulation model of infant-incubator-feedback system with humidification and temperature control

A comprehensive simulation model for the infant-incubator-feedback system was developed in a Matlab/Simulink environment to investigate all heat exchange relationships, variables and factors that have an influence on the overall thermoneutrality of the environment. The model was also used to determine the benefits and limitations of using a convectively heated single-walled incubator in nursing preterm infants with very low birth weight < 1000 grams and low gestational age 28 weeks. The infant was modelled as one-lump with two layers; core and skin. The infant shape was approximated to a cylinder. The model incorporated all compartments of the infantincubator system including core, skin, incubator air space, mattress, incubator walls, aircirculating fan, heating element, added oxygen (for resuscitation purposes), and humidification chamber, which has not previously been considered. The results of the simulation were in terms of the temperature variation over time, of the following parts of the system: core and skin temperatures and incubator air space temperature. Results of the simulation with added humidity showed that the body temperature of a 900 gram infant, with an initial body temperature of 35.5 oC, did not reach the thermo-neutral range between 36.5-37.5 oC in two hours, on air mode. Whereas, on skin mode, both core and skin temperature reached to 36.87 oC and 36.5 oC in two hours, and thus a thermo-neutral environment was achieved. These outcomes are consistent with clinical empirical reports. The simulation model is a closed-loop system with a PID controller for each mode; air servo controlled and skin servo controlled. The controller parameters were virtually estimated by the Zeigler-Nichols Method as real values were not available. Nevertheless, the overall stability of the whole system has been achieved by applying a step input which was verified by the Root Locus Method.

[1]  William Bolton Mechatronics: Electronic Control Systems in Mechanical and Electrical Engineering , 2004 .

[2]  James D. Hardy,et al.  Physiological and behavioral temperature regulation , 1970 .

[3]  S. Baumgart Reduction of oxygen consumption, insensible water loss, and radiant heat demand with use of a plastic blanket for low-birth-weight infants under radiant warmers. , 1984, Pediatrics.

[4]  M. Leblanc Relative efficacy of an incubator and an open warmer in producing thermoneutrality for the small premature infant. , 1982, Pediatrics.

[5]  A. Wheldon,et al.  The heat balance of small babies nursed in incubators and under radiant warmers. , 1982, Early human development.

[6]  W. Oh,et al.  Combined effect of radiant warmer and phototherapy on insensible water loss in low-birth-weight infants. , 1979, The Journal of pediatrics.

[7]  M. Leblanc Thermoregulation: incubators, radiant warmers, artificial skins, and body hoods. , 1991, Clinics in perinatology.

[8]  Katsuhiko Ogata,et al.  Modern Control Engineering , 1970 .

[9]  W. Silverman,et al.  The influence of the thermal environment upon the survival of newly born premature infants. , 1958, Pediatrics.

[10]  S. Vohra,et al.  Effect of polyethylene occlusive skin wrapping on heat loss in very low birth weight infants at delivery: a randomized trial. , 1999, The Journal of pediatrics.

[11]  E. Bell,et al.  A Double-Walled Incubator Alters the Partition of Body Heat Loss of Premature Infants , 1983, Pediatric Research.

[12]  LeBlanc Mh The physics of thermal exchange between infants and their environment. , 1987 .

[13]  Ernest O. Doebelin System Dynamics: Modeling, Analysis, Simulation, Design , 1998 .

[14]  R. Ariagno,et al.  Minimal oxygen consumption in infants cared for under overhead radiant warmers compared with conventional incubators. , 1978, The Journal of pediatrics.

[15]  M. Weinstein,et al.  Heat balance in premature infants: comparative effects of convectively heated incubator and radiant warmer, with without plastic heat shield. , 1980, The Journal of pediatrics.

[16]  W. Fox,et al.  Radiant Warmer Power and Body Size as Determinants of Insensible Water Loss in the Critically Ill Neonate , 1981, Pediatric Research.

[17]  Aural temperature of the newborn infant. , 1977 .

[18]  P. Chessex,et al.  Environmental temperature control in very low birth weight infants (less than 1000 grams) cared for in double-walled incubators. , 1988, The Journal of pediatrics.

[19]  J. Hodgman,et al.  Insensible water loss in preterm infants: changes with postnatal development and non-ionizing radiant energy. , 1974, Pediatrics.

[20]  R. Cooke Neonatal—perinatal medicine: Diseases of the Fetus and Infant, 5th ed. Avroy A. Fanaroff and Richard J. Martin, Eds. St. Louis, MO: Mosby Year Book, 1992, 1450 pp , 1993 .

[21]  N. B. White,et al.  Metabolic aspects of phototherapy. , 1985, Pediatrics.

[22]  W. Fox,et al.  Effect of heat shielding on convective and evaporative heat losses and on radiant heat transfer in the premature infant. , 1981, The Journal of pediatrics.

[23]  G. Merenstein,et al.  Radiant warmers vs incubators for neonatal care. , 1979, American journal of diseases of children.

[24]  D. Miller,et al.  Body temperature in the immediate neonatal period: the effect of reducing thermal losses. , 1966, American journal of obstetrics and gynecology.

[25]  N. McIntosh,et al.  Temperature control in very low birthweight infants during first five days of life , 1997, Archives of disease in childhood. Fetal and neonatal edition.

[26]  J. Lister The optimum thermal environment for naked babies , 1970 .

[27]  M H LeBlanc The physics of thermal exchange between infants and their environment. , 1987, Medical instrumentation.