Optimization of the Injected Activity in Dynamic 3D PET: A Generalized Approach Using Patient-Specific NECs as Demonstrated by a Series of 15O-H2O Scans

The magnitude of the injected activity (A0) has a direct impact on the statistical quality of PET images. This study aimed to develop a generalized method for maximizing the statistical quality of dynamic PET images by optimizing A0. Methods: Patient-specific noise-equivalent counts (PS-NECs) were used as a metric of the statistical quality of each time frame of a dynamic PET image. Previous methodology developed to extrapolate the NEC as a function of A0 was extended to dynamic PET, enabling the NEC to be extrapolated as a function of both A0 and the time after injection. This method allowed A0 to be optimized after a single scan (at a single A0), by maximizing the NEC within the time interval for which the parameter estimation is most sensitive. The extrapolation method was validated by a series of 15O-H2O scans of the body acquired in 3-dimensional mode. Each patient (n = 6) underwent between 3 and 6 scans at 1 bed position. The injected activities were varied over a wide range (140–840 MBq). Noise-equivalent counting rate (NECR) versus A0 curves and the optimal injected activities were calculated from each injection. Results: PS-NECR versus A0 curves as extrapolated from different injected activities were consistent (coefficient of variation, typically <5%). The optimal injected activities for an individual, as derived from these curves, were also consistent (maximum coefficient of variation, 4.3%). For abdominal (n = 4) and chest (n = 1) scans, we found optimal injected activities of 15O-H2O in the range of 220–350 MBq for estimating blood perfusion (F) and 660–1,070 MBq for estimating the volume of distribution (VT). Higher optimal injected activities were found in the case of a pelvic scan (n = 1; 570 MBq for F and 1,530 MBq for VT). Conclusion: PS-NECs are a valid and generic method for optimizing the injected activity in PET, allowing scanning protocols to be improved after the collection of an initial, single dynamic dataset. This generic method can be used to estimate the optimal injected activity, which is specific to the patient, tracer, PET scanner, and body region being scanned.

[1]  C. Watson,et al.  Count rate dependence of local signal-to-noise ratio in positron emission tomography , 2003, IEEE Transactions on Nuclear Science.

[2]  S. Kety The theory and applications of the exchange of inert gas at the lungs and tissues. , 1951, Pharmacological reviews.

[3]  C.C. Watson,et al.  Advances in scatter correction for 3D PET/CT , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[4]  D. Comar,et al.  Radiation dosimetry for bolus administration of oxygen-15-water. , 1995, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[5]  R. Nutt,et al.  The NEC dependence of different scintillators for positron emission tomography , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[6]  E. Hoffman,et al.  Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts , 1990 .

[7]  Richard E. Carson,et al.  Noise characteristics of 3-D and 2-D PET images , 1998, IEEE Transactions on Medical Imaging.

[8]  Stefan Eberl,et al.  Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  F. Fazio,et al.  Generation of the Acquisition-Specific NEC (AS-NEC) Curves to Optimize the Injected Dose in 3D$^18$F-FDG Whole Body PET Studies , 2006, IEEE Transactions on Nuclear Science.

[10]  T D Cradduck,et al.  National electrical manufacturers association , 1983, Journal of the A.I.E.E..

[11]  R. Boellaard,et al.  Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[12]  R E Carson,et al.  Noise reduction in oncology FDG PET images by iterative reconstruction: a quantitative assessment. , 2001, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[13]  Carole Lartizien,et al.  Optimization of injected dose based on noise equivalent count rates for 2- and 3-dimensional whole-body PET. , 2002, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[14]  Roberta Matheoud,et al.  Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[15]  T. Beyer,et al.  Evaluation of clinical PET count rate performance , 2002 .

[16]  M N Maisey,et al.  Optimization of noise-equivalent count rates in 3D PET. , 1996, Physics in medicine and biology.

[17]  Suleman Surti,et al.  Imaging characteristics of a 3-dimensional GSO whole-body PET camera. , 2004, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[18]  P. Price,et al.  Patient-specific noise-equivalent-counts from repeated, dose varying [O-15]H2O PET scans , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.

[19]  J. John Mann,et al.  OS-EM and FBP reconstructions at low count rates: effect on 3D PET studies of [11C] WAY-100635 , 2004, NeuroImage.

[20]  W. Heiss,et al.  The ECAT HRRT: an example of NEMA scatter estimation issues for LSO-based PET systems , 2005, IEEE Transactions on Nuclear Science.