Small field output factors evaluation with a microDiamond detector over 30 Italian centers.

PURPOSE The aim of the study was a multicenter evaluation of MLC&jaws-defined small field output factors (OF) for different linear accelerator manufacturers and for different beam energies using the latest synthetic single crystal diamond detector commercially available. The feasibility of providing an experimental OF data set, useful for on-site measurements validation, was also evaluated. METHODS This work was performed in the framework of the Italian Association of Medical Physics (AIFM) SBRT working group. The project was subdivided in two phases: in the first phase each center measured OFs using their own routine detector for nominal field sizes ranging from 10×10cm2 to 0.6×0.6cm2. In the second phase, the measurements were repeated in all centers using the PTW 60019 microDiamond detector. RESULTS The project enrolled 30 Italian centers. Micro-ion chambers and silicon diodes were used for OF measurements in 24 and 6 centers respectively. Gafchromic films and TLDs were used for very small field OFs in 3 and 1 centers. Regarding the measurements performed with the user's detectors, OF standard deviations (SD) for field sizes down to 2×2cm2 were in all cases <2.7%. In the second phase, a reduction of around 50% of the SD was obtained using the microDiamond detector. CONCLUSIONS The measured values presented in this multicenter study provide a consistent dataset for OFs that could be a useful tool for improving dosimetric procedures in centers. The microDiamond data present a small variation among the centers confirming that this detector can contribute to improve overall accuracy in radiotherapy.

[1]  J Seuntjens,et al.  On the correction, perturbation and modification of small field detectors in relative dosimetry , 2014, Physics in medicine and biology.

[2]  R Svensson,et al.  Design and characterization of a tissue-equivalent CVD-diamond detector for clinical dosimetry in high-energy photon beams. , 2008, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[3]  R Ferrand,et al.  Application of the Exradin W1 scintillator to determine Ediode 60017 and microDiamond 60019 correction factors for relative dosimetry within small MV and FFF fields , 2015, Physics in medicine and biology.

[4]  Hiroki Shirato,et al.  Stereotactic body radiotherapy (SBRT) for operable stage I non-small-cell lung cancer: can SBRT be comparable to surgery? , 2011, International journal of radiation oncology, biology, physics.

[5]  S Cora,et al.  Calculation of k(Q(clin),Q(msr) ) (f(clin),f(msr) ) for several small detectors and for two linear accelerators using Monte Carlo simulations. , 2011, Medical physics.

[6]  Natalka Suchowerska,et al.  Small field diode correction factors derived using an air core fibre optic scintillation dosimeter and EBT2 film , 2012, Physics in medicine and biology.

[7]  David I Thwaites,et al.  Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations. , 2011, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[8]  S Derreumaux,et al.  Lessons from recent accidents in radiation therapy in France. , 2008, Radiation protection dosimetry.

[9]  John D Fenwick,et al.  Monte Carlo modeling of small photon fields: Quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small-field measurements. , 2009, Medical physics.

[10]  M. Aspradakis,et al.  Small Field MV Photon Dosimetry , 2009 .

[11]  J. Seuntjens,et al.  A new formalism for reference dosimetry of small and nonstandard fields. , 2008, Medical physics.

[12]  J. Lárraga-Gutiérrez,et al.  Properties of a commercial PTW-60019 synthetic diamond detector for the dosimetry of small radiotherapy beams , 2015, Physics in medicine and biology.

[13]  Abraham Philip,et al.  Stereotactic Body Radiation Therapy (SBRT) for lung metastases , 2006, Acta oncologica.

[14]  Ahmed Meghzifene,et al.  Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams. , 2014, Medical physics.

[15]  Elena Villaggi,et al.  A feasibility dosimetric study on prostate cancer , 2015, Strahlentherapie und Onkologie.

[16]  Scott B Crowe,et al.  Dosimetry of cone-defined stereotactic radiosurgery fields with a commercial synthetic diamond detector. , 2014, Medical physics.

[17]  Pietro Mancosu,et al.  Evaluation of a synthetic single-crystal diamond detector for relative dosimetry on the Leksell Gamma Knife Perfexion radiosurgery system. , 2015, Medical physics.

[18]  Jamie V Trapp,et al.  A methodological approach to reporting corrected small field relative outputs. , 2013, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[19]  Pietro Mancosu,et al.  SBRT for prostate cancer: Challenges and features from a physicist prospective. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[20]  A. Chalkley,et al.  Evaluation of a synthetic single-crystal diamond detector for relative dosimetry measurements on a CyberKnife. , 2014, The British journal of radiology.

[21]  Carlo Cavedon,et al.  Total scatter factors of small beams: a multidetector and Monte Carlo study. , 2008, Medical physics.

[22]  D I Thwaites,et al.  Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors. , 2011, Medical physics.

[23]  S Derreumaux,et al.  Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones. , 2013, Medical physics.

[24]  Elisabetta Cagni,et al.  Technical Note: Multicenter study of TrueBeam FFF beams with a new stereotactic diode: Can a common small field signal ratio curve be defined? , 2016, Medical physics.

[25]  Lei Xing,et al.  Stereotactic body radiation therapy in multiple organ sites. , 2007, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[26]  T Kairn,et al.  A practical and theoretical definition of very small field size for radiotherapy output factor measurements. , 2014, Medical physics.

[27]  Marco Marinelli,et al.  Dosimetric characterization of a synthetic single crystal diamond detector in clinical radiation therapy small photon beams. , 2012, Medical physics.

[28]  Serenella Russo,et al.  Role of the technical aspects of hypofractionated radiation therapy treatment of prostate cancer: a review. , 2015, International journal of radiation oncology, biology, physics.

[29]  D I Thwaites,et al.  Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterization , 2012, Physics in medicine and biology.

[30]  Pietro Mancosu,et al.  Stereotactic body radiation therapy for liver tumours using flattening filter free beam: dosimetric and technical considerations , 2012, Radiation oncology.

[31]  M. Bucciolini,et al.  Diamond detector versus silicon diode and ion chamber in photon beams of different energy and field size. , 2003, Medical physics.

[32]  Elisabetta Cagni,et al.  Multicentre treatment planning inter-comparison in a national context: The liver stereotactic ablative radiotherapy case. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[33]  S. Rustgi Evaluation of the dosimetric characteristics of a diamond detector for photon beam measurements. , 1995, Medical physics.

[34]  Sandro Onori,et al.  Clinical studies of optimised single crystal and polycrystalline diamonds for radiotherapy dosimetry , 2008 .

[35]  J. Debus,et al.  Correction of patient positioning errors based on in-line cone beam CTs: clinical implementation and first experiences , 2005, Radiation oncology.

[36]  Serenella Russo,et al.  CyberKnife beam output factor measurements: A multi-site and multi-detector study. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[37]  M. Schell,et al.  Hypofractionated stereotactic body radiation therapy (SBRT) for limited hepatic metastases. , 2007, International journal of radiation oncology, biology, physics.

[38]  B McClean,et al.  Small field detector correction factors kQclin,Qmsr (fclin,fmsr) for silicon-diode and diamond detectors with circular 6 MV fields derived using both empirical and numerical methods. , 2015, Medical physics.

[39]  Andrea Molineu,et al.  The Radiological Physics Center's standard dataset for small field size output factors , 2012, Journal of applied clinical medical physics.

[40]  Serenella Russo,et al.  Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[41]  Josep Sempau,et al.  Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: a PENELOPE Monte Carlo study. , 2014, Medical physics.

[42]  A. Brahme,et al.  Comparative dosimetry in narrow high-energy photon beams. , 2000, Physics in medicine and biology.

[43]  K. Leszczynski,et al.  Estimation of the focal spot size and shape for a medical linear accelerator by Monte Carlo simulation. , 2007, Medical physics.

[44]  G. Prestopino,et al.  Synthetic single crystal diamond dosimeters for Intensity Modulated Radiation Therapy applications , 2009 .

[45]  Otto A Sauer,et al.  Measurement of output factors for small photon beams. , 2007, Medical physics.

[46]  Paolo Francescon,et al.  Qclin;Qmsr for several small detectors and for two linear accelerators using Monte Carlo simulations , 2011 .

[47]  N Suchowerska,et al.  Small field dosimetric characterization of a new 160-leaf MLC. , 2013, Physics in medicine and biology.

[48]  Elisabetta Cagni,et al.  Lung stereotactic ablative body radiotherapy: A large scale multi-institutional planning comparison for interpreting results of multi-institutional studies. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.