Measurements of Surgeons' Exposure to Ionizing Radiation Dose During Intraoperative Use of C-Arm Fluoroscopy

Study Design. Measurement of radiation dose from C-arm fluoroscopy during a simulated intraoperative use in spine surgery. Objective. To investigate scatter radiation doses to specific organs of surgeons during intraoperative use of C-arm fluoroscopy in spine surgery and to provide practical intraoperative guidelines. Summary of Background Data. There have been studies that reported the radiation dose of C-arm fluoroscopy in various procedures. However, radiation doses to surgeons' specific organs during spine surgery have not been sufficiently examined, and the practical intraoperative radioprotective guidelines have not been suggested. Methods. Scatter radiation dose (air kerma rate) was measured during the use of a C-arm on an anthropomorphic chest phantom on an operating table. Then, a whole body anthropomorphic phantom was located besides the chest phantom to simulate a surgeon, and scatter radiation doses to specific organs (eye, thyroid, breast, and gonads) and direct radiation dose to the surgeon's hand were measured using 4 C-arm configurations (standard, inverted, translateral, and tube translateral). The effects of rotating the surgeon's head away from the patient and of a thyroid shield were also evaluated. Results. Scatter radiation doses decreased as distance from the patient increased during C-arm fluoroscopy use. The standard and translateral C-arm configurations caused lower scatter doses to sensitive organs than inverted and tube translateral configurations. Scatter doses were highest for breast and lowest for gonads. The use of a thyroid shield and rotating the surgeon's head away from the patient reduced scatter radiation dose to the surgeon's thyroid and eyes. The direct radiation dose was at least 20 times greater than scatter doses to sensitive organs. Conclusion. The following factors could reduce radiation exposure during intraoperative use of C-arm; (1) distance from the patient, (2) C-arm configuration, (3) radioprotective equipments, (4) rotating the surgeons' eyes away from the patient, and (5) avoiding direct exposure of surgeons' hands.

[1]  Glenn R Rechtine,et al.  Patient and surgeon radiation exposure: comparison of standard and mini-C-arm fluoroscopy. , 2009, The Journal of bone and joint surgery. American volume.

[2]  N. Theocharopoulos,et al.  Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. , 2003, The Journal of bone and joint surgery. American volume.

[3]  R. Hsu,et al.  Minimally invasive vertebroplasty managed by a two C‐arm fluoroscopic technique , 2007, Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy.

[4]  M. J. Dennis,et al.  Radiation Exposure With Use Of the Inverted-C-Arm Technique in Upper-Extremity Surgery , 2001, The Journal of bone and joint surgery. American volume.

[5]  E. Kotter,et al.  Computer-navigated Iliosacral Screw Insertion Reduces Malposition Rate and Radiation Exposure , 2009, Clinical orthopaedics and related research.

[6]  D. Lalonde,et al.  Measurement of radiation exposure over a one-year period from Fluoroscan mini c-arm imaging unit. , 2007, Plastic and reconstructive surgery.

[7]  L. Manchikanti,et al.  Radiation exposure to the physician in interventional pain management. , 2002, Pain physician.

[8]  Brian D Giordano,et al.  Exposure to direct and scatter radiation with use of mini-c-arm fluoroscopy. , 2007, The Journal of bone and joint surgery. American volume.

[9]  S. Wolfe,et al.  Radiation exposure in hand surgery: mini versus standard C-arm. , 2005, The Journal of hand surgery.

[10]  Lynn Rill,et al.  Radiation exposure with use of the mini-C-arm for routine orthopaedic imaging procedures. , 2005, The Journal of bone and joint surgery. American volume.

[11]  Eleanor A. Blakely,et al.  Biological Effects of Cosmic Radiation: Deterministic and Stochastic , 2000, Health physics.

[12]  Jonathan N Grauer,et al.  Radiation exposure from musculoskeletal computerized tomographic scans. , 2009, The Journal of bone and joint surgery. American volume.

[13]  R. Bansal,et al.  A Comparison of Radiation Exposure with the Conventional versus Mini C Arm in Orthopedic Extremity Surgery , 2008, Foot & ankle international.

[14]  G. Rechtine,et al.  Minimally invasive surgery and radiation exposure. , 2009, Journal of neurosurgery. Spine.

[15]  Glenn R Rechtine,et al.  Cervical Spine Imaging Using Standard C-Arm Fluoroscopy: Patient and Surgeon Exposure to Ionizing Radiation , 2008, Spine.

[16]  J. Baumhauer,et al.  Cervical Spine Imaging Using Mini–C-arm Fluoroscopy: Patient and Surgeon Exposure to Direct and Scatter Radiation , 2009, Journal of spinal disorders & techniques.