Visual evoked potential-based acuity assessment in normal vision, artificially degraded vision, and in patients

Aims: To assess visual acuity (VA) objectively using visual evoked potentials (VEPs), avoiding subjective trace evaluation and providing an acuity estimate with associated confidence limits. Methods: 40 normal subjects and 24 patients (with corneal and retinal diseases, decimal VA range 0.15–1.1 ( = 0.8logMAR to −0.04logMAR)) participated in the study. Checkerboard stimuli with six check sizes covering 0.05–0.4° (or 0.09–0.8° for visual acuities below 0.35 ( = 0.46logMAR) were presented in brief-onset mode (40 ms on, 93 ms off) at 7.5 Hz. In normal subjects, the stimuli were also optically degraded by frosted occluders resulting in a decimal VA range of 0.13–2.8 ( = 0.9logMAR to −0.45logMAR). Altogether, 108 steady-state VEPs were recorded with a Laplacian montage (2×Oz−(RO+LO)). Fourier analysis yielded the magnitude (A) at the stimulus frequency, and the average of the two neighboring frequencies as noise estimate (N). A and N determine the significance level p of the response, and from their ratio the non-noise-contaminated response (A*) can be calculated. Tuning curves were obtained by plotting A* vs the dominant spatial frequency of the corresponding checkerboard. A fully automatic algorithm used the significance level (p<5%) and A* to automatically select an appropriate region in the high spatial-frequency range on which a linear regression was performed, yielding a zero-amplitude extrapolated spatial frequency SF0. Subjective VA was obtained with the automated “Freiburg Acuity Test”. Results: The brief-onset presentation evoked high VEP amplitudes; however, many tuning curves displayed the well-known “notch” at intermediate check sizes. The fully automated analysis algorithm succeeded in 107 of 108 cases and effectively ignored the notch, if present. The relation between logVA and log(SF0) was a constant factor throughout the range tested: logVA = log(SF0)/17.6 cpd. In more than 95% of all cases, the acuity predicted from SF0 coincided within a factor of two (up and down, or ±0.3 logMAR) with subjective VA with a coefficient of correlation of 0.90. Conclusion: The fully automated analysis avoided subjective problems in peak-trough assessment. The results provide quantitative limits to assess patients with possible malingering.

[1]  Arden Gb,et al.  The pattern visual evoked response in the assessment of visual acuity. , 1973 .

[2]  K Nakayama,et al.  Rapid assessment of visual function: an electronic sweep technique for the pattern visual evoked potential. , 1979, Investigative ophthalmology & visual science.

[3]  A. Norcia,et al.  Spatial frequency sweep VEP: Visual acuity during the first year of life , 1985, Vision Research.

[4]  Anthony M. Norcia,et al.  Digital Filtering and Robust Regression Techniques for Estimating Sensory Thresholds from the Evoked Potential , 1985, IEEE Engineering in Medicine and Biology Magazine.

[5]  Graham E. Holder,et al.  Assessment of patients with suspected non-organic visual loss using pattern appearance visual evoked potentials , 2007, Graefe's Archive for Clinical and Experimental Ophthalmology.

[6]  W. Cobb,et al.  Cerebral potentials evoked in man by pattern reversal and their suppression in visual rivalry. , 1969, Electroencephalography and clinical neurophysiology.

[7]  C. Teping [Determination of visual acuity by the visually evoked cortical potential (author's transl)]. , 1981, Klinische Monatsblatter fur Augenheilkunde.

[8]  Hans Strasburger,et al.  THE ANALYSIS OF STEADY STATE EVOKED POTENTIALS REVISITED , 1987 .

[9]  N. N. Available World medical association declaration of Helsinki , 2000, Chinese Journal of Integrative Medicine.

[10]  H. Strasburger,et al.  Objective measurement of contrast sensitivity and visual acuity with the steady-state visual evoked potential. , 1996, German journal of ophthalmology.

[11]  D. Regan Assessment of visual acuity by evoked potential recording: Ambiguity caused by temporal dependence of spatial frequency selectivity , 1978, Vision Research.

[12]  J. W. Howe,et al.  Electrophysiological assessment of visual acuity. , 1981, Transactions of the ophthalmological societies of the United Kingdom.

[13]  Michael S. Bradnam,et al.  Faster and more sensitive VEP recording in children , 2003, Documenta Ophthalmologica.

[14]  I. Murray,et al.  Spatio-temporal tuning of VEPs: effect of mode of stimulation , 1999, Vision Research.

[15]  Wolzt,et al.  World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. , 2003, The Journal of the American College of Dentists.

[16]  E. Zrenner,et al.  [Improved objective visual assessment with visual evoked cortical potentials by rapid pattern stimuli sequences of different spatial frequency]. , 1988, Fortschritte der Ophthalmologie : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft.

[17]  A. Herbert,et al.  Stimulus duration, neural adaptation, and sweep visual evoked potential acuity estimates. , 1998, Investigative ophthalmology & visual science.

[18]  G B Arden,et al.  The pattern visual evoked response in the assessment of visual acuity. , 1973, Transactions of the ophthalmological societies of the United Kingdom.

[19]  C. Tyler,et al.  PROPERTIES OF LOCALIZED PATTERN EVOKED POTENTIALS * , 1980, Annals of the New York Academy of Sciences.

[20]  Michael Bach,et al.  Principles and practice of clinical electrophysiology of vision , 1991 .

[21]  Cerebral potentials evoked in man by pattern reversal and their suppresion in visual rivalry. , 1968, The Journal of physiology.

[22]  Anthony M. Norcia,et al.  Measurement of spatial contrast sensitivity with the swept contrast VEP , 1989, Vision Research.

[23]  M. Bach,et al.  Do's and don'ts in Fourier analysis of steady-state potentials , 2004, Documenta Ophthalmologica.

[24]  J. Kelsey,et al.  Visual Evoked Potentials in Man: New Developments , 1978 .

[25]  Comparison of Snellen acuity and objective assessment using the spatial frequency sweep PVER , 1997, Graefe's Archive for Clinical and Experimental Ophthalmology.

[26]  World Medical Association (WMA),et al.  Declaration of Helsinki. Ethical Principles for Medical Research Involving Human Subjects , 2009, Journal of the Indian Medical Association.

[27]  M. Bach,et al.  On the statistical significance of electrophysiological steady-state responses , 2004, Documenta Ophthalmologica.

[28]  [Amblyopia: reading speed in comparison with visual acuity for gratings, single Landolt Cs and series Landolt Cs]. , 1990, Fortschritte der Ophthalmologie : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft.

[29]  K. Mullen The contrast sensitivity of human colour vision to red‐green and blue‐yellow chromatic gratings. , 1985, The Journal of physiology.

[30]  Sheng Chen,et al.  Automatic Kernel Regression Modelling Using Combined Leave-One-Out Test Score and Regularised Orthogonal Least Squares , 2004, Int. J. Neural Syst..

[31]  Spatial frequency sweep pattern reversal VER acuity vs Snellen visual acuity: Effect of optical defocus , 1996, Vision Research.

[32]  William H. Ridder,et al.  Methods of visual acuity determination with the spatial frequency sweep visual evoked potential , 2004, Documenta Ophthalmologica.

[33]  J. Gwiazda,et al.  Repeatability of ETDRS visual acuity in children. , 2003, Investigative ophthalmology & visual science.

[34]  C T White,et al.  Evoked cortical responses to checkerboard patterns: effect of check-size as a function of visual acuity. , 1970, Electroencephalography and clinical neurophysiology.

[35]  M. Bach,et al.  The Freiburg Visual Acuity test--automatic measurement of visual acuity. , 1996, Optometry and vision science : official publication of the American Academy of Optometry.

[36]  M. Bach,et al.  Raster-scan cathode-ray tubes for vision research--limits of resolution in space, time and intensity, and some solutions. , 1997, Spatial vision.

[37]  M Bach,et al.  Contrast adaptation in human retina and cortex. , 2001, Investigative ophthalmology & visual science.

[38]  M. Banks Visual acuity development in human infants: a re-evaluation. , 1977, Investigative ophthalmology & visual science.