Design of an implanted compact antenna for an artificial cardiac pacemaker system
暂无分享,去创建一个
An implanted compact antenna for an artificial cardiac pacemaker is proposed. The dimension of the pacemaker system, including the antenna element, is 30mm × 35mm × 7mm. When the antenna is embedded in a semi-solid flat phantom with equivalent electrical properties as the human body, S11 value is -19.2dB at 403.5MHz and the -10dB impedance bandwidth of the antenna is 10MHz (399∼409MHz). The proposed antenna in the phantom has a peak gain of -24.61dBi at 403.5MHz. The measured specific absorption ratio (SAR) value of the proposed antenna is 0.0079W/Kg (1g tissue). Moreover, to estimate the communication performance of the proposed antenna operated in the real environment, a link budget analysis is performed.
[1] M. Takahashi,et al. Implanted Planar Inverted F-Antenna for Cardiac Pacemaker System , 2008, 2008 International Workshop on Antenna Technology: Small Antennas and Novel Metamaterials.
[2] C.M. Furse,et al. Design of implantable microstrip antenna for communication with medical implants , 2004, IEEE Transactions on Microwave Theory and Techniques.