Stable Atrial Sensing on Long-Term Follow Up of VDD Pacemakers

Background The hemodynamic advantages of maintaining AV synchrony through AV synchronous pacing are widely known as compared to single chamber pacing. DDD pacemaker implantation entails higher cost and is technically more challenging than the VDD pacemaker. Methods Seventy one patients underwent VDD lead (Biotronik GmbH, St. Jude Medical and Medtronic Inc.) implantation at KEM hospital, Mumbai during a period of 3 years through subclavian, axillary and cephalic routes for degenerative, post-surgical or congenital high grade atrioventricular or complete heart block. They were followed up regularly for ventricular threshold and P wave amplitude of the floating atrial dipole. Results Follow up data of almost 95% of patients is available for a period of 15.8 ± 6.7 months. P wave amplitude at implant was 2.1 ± 0.7mV and at follow up 1.1 ± 0.6mV with mean ventricular threshold of <0.5V at implant and <1V at follow-up. Conclusion Implantation of a single lead VDD pacemaker is possible in all patients with symptomatic AV block and intact sinus node function without any technical complications. P wave sensing is reliable and consistent with floating atrial lead at an average follow up of 15.8 months, providing an excellent alternative to DDD pacemaker implantation.

[1]  A. Manolis,et al.  Comparative three-year performance of chronic atrial sensing among eight different VDD systems. , 2002, The American journal of cardiology.

[2]  C. Lau,et al.  Limitation of long-term atrial pacing via floating ring electrode using overlapping biphasic impulse stimulation in ambulatory-paced patients. , 1998, The American journal of cardiology.

[3]  A. Liebold,et al.  Clinical results with single lead VDD pacing. , 1997, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[4]  T. Toselli,et al.  Italian Multicenter Study on a Single Lead VDD Pacing System Using a Narrow Atrial Dipole Spacing , 1992, Pacing and clinical electrophysiology : PACE.

[5]  U. Parravicini,et al.  Complications of Single Lead VDD Pacing in 35 Patients with AV Block , 1992, Pacing and clinical electrophysiology : PACE.

[6]  H. Schieffer,et al.  Atrial signal variations and pacemaker malsensing during exercise: a study in the time and frequency domain. , 1988, Journal of the American College of Cardiology.

[7]  H. Schwerdt,et al.  Atrial Sensing Performance of AV Universal Pacemakers During Exercise , 1988, Pacing and clinical electrophysiology : PACE.

[8]  W. Brownlee,et al.  Six Years' Experience with Atrial Leads , 1986, Pacing and clinical electrophysiology : PACE.