Onset of nanoscale wear of metallic implant materials: Influence of surface residual stresses and contact loads

Abstract Nanoscale wear response of cobalt–chromium alloy (ASTM F-75) was investigated as a function of contact load and surface residual stress in order to identify mechanism governing onset of surface damage in modular implants. A unique loading configuration was utilized to apply range of known in-plane stress states to the specimen surface in order to simulate different residual stress levels. Using the tip of a scanning force microscope as a well characterized “asperity,” controlled contact forces were applied on the specimen to mechanically stimulate the loaded surface. Volume of material removed was measured to characterize the wear rate as a function of the contact loads and surface stress state. Experimental measurements of material removal indicate that a critical level of contact pressure is required to initiate wear of the cobalt-chromium surface and as expected higher contact pressures accelerate the wear process. At a constant contact pressure, wear rate is accelerated by compressive in-plane stress while tensile in-plane stresses tend to suppress the surface wear. A surface damage mechanism based on successive damage/delamination of native oxide covered surface due to single asperity contact and repassivation of exposed surface is proposed to elucidate the experimental observations.

[1]  George M. Pharr,et al.  Influences of stress on the measurement of mechanical properties using nanoindentation: Part I. Experimental studies in an aluminum alloy , 1996 .

[2]  M. Metikoš-huković,et al.  Passive film on orthopaedic TiAlV alloy formed in physiological solution investigated by X-ray photoelectron spectroscopy. , 2000, Biomaterials.

[3]  K. Komvopoulos,et al.  Effect of Residual Stress in Surface Layer on Contact Deformation of Elastic-Plastic Layered Media , 2003 .

[4]  John W. Hutchinson,et al.  Crack deflection at an interface between dissimilar elastic-materials , 1989 .

[5]  Z. Suo,et al.  Stress-dependent surface reactions and implications for a stress measurement technique , 2000 .

[6]  Jack R. Worrall,et al.  Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles. , 1998, Journal of biomedical materials research.

[7]  J. Celis,et al.  Oxidational wear of TiN coatings on tool steel and nitrided tool steel in unlubricated fretting , 1995 .

[8]  J. Jacobs,et al.  Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacements , 2004, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[9]  Donald W. Brenner,et al.  Atomistic simulation of the influence of pre-existing stress on the interpretation of nanoindentation data , 2004 .

[10]  A. Batchelor,et al.  Some considerations on the mitigation of fretting damage by the application of surface-modification technologies , 2000 .

[11]  A. Iwabuchi The role of oxide particles in the fretting wear of mild steel , 1991 .

[12]  D. Zukor,et al.  Size, shape, and composition of wear particles from metal-metal hip simulator testing: effects of alloy and number of loading cycles. , 2003, Journal of biomedical materials research. Part A.

[13]  Studies on fretting wear: influence of rubbing surface materials and some considerations , 2002 .

[14]  G. Webster,et al.  Measurement of Residual Stress Through a Shot Peened Surface Subjected to Successive Material Removal , 2002 .

[15]  J. Jacobs,et al.  Orthopedic Implant Fretting Corrosion , 2003 .

[16]  J Fisher,et al.  Wear of surface engineered metal-on-metal hip prostheses , 2004, Journal of materials science. Materials in medicine.

[17]  Marco Tortonese,et al.  Characterization of application-specific probes for SPMs , 1997, Photonics West.

[18]  R. Pilliar,et al.  The fretting corrosion resistance of PVD surface-modified orthopedic implant alloys. , 2001, Journal of biomedical materials research.

[19]  C. Langlade,et al.  Residual stresses and fretting fatigue , 1999 .

[20]  J. Gilbert,et al.  Electrochemical response of CoCrMo to high-speed fracture of its metal oxide using an electrochemical scratch test method. , 1997, Journal of biomedical materials research.

[21]  D. Landolt,et al.  A tribo-electrochemical apparatus for in vitro investigation of fretting–corrosion of metallic implant materials , 2002 .

[22]  H. Tan,et al.  Evolution of a Surface-Roughness Spectrum Caused by Stress in Nanometer-Scale Chemical Etching , 1999 .

[23]  Ingrid Milošev,et al.  The composition of the surface passive film formed on CoCrMo alloy in simulated physiological solution , 2003 .

[24]  T. Boland,et al.  Nanotribology of CoCr–UHMWPE TJR prosthesis using atomic force microscopy , 2002 .