Normal-incidence efficiencies of 4800-grooves/mm-ruled replica gratings with multilayer and gold coatings in the 125-325-A wavelength region.

The normal-incidence efficiencies of two 4800-grooves/mm ruled replica gratings, one with a dual-bandpass molybdenum/silicon multilayer coating and the other with a gold coating, were measured by use of synchrotron radiation in the 125-325-A wavelength region. The peak reflectance of the multilayer coating was 22% in the first Bragg order near 235 A and 28% in the second Bragg order near 126 A. The peak efficiency of the multilayer grating was 2.6% in the first diffraction order near 225 A and 0.3% in the second diffraction order near 125 A. The efficiencies of the multilayer grating were much higher than the corresponding efficiencies of the gold grating. The characterization of the surfaces of the gratings by atomic force microscopy indicated rms microroughness values in the 5-18-microm(-1) frequency range of 12-20 A for the multilayer grating and 22-32 A for the gold grating. Both gratings had bumpy surface features larger than the nominal groove height. The rather large surface roughness and groove irregularities had a detrimental effect on the grating efficiencies.

[1]  W. R. Hunter,et al.  An ultrahigh vacuum reflectometer/goniometer for use with synchrotron radiation , 1986 .

[2]  B. L. Henke,et al.  X-Ray Interactions: Photoabsorption, Scattering, Transmission, and Reflection at E = 50-30,000 eV, Z = 1-92 , 1993 .

[3]  W. R. Hunter,et al.  Upgrades and recent performance of the grating/crystal monochromator , 1989 .

[4]  T W Barbee,et al.  On-blaze operation of a Mo/Si multilayer-coated, concave diffraction grating in the 136-142-A wavelength region and near normal incidence. , 1993, Applied optics.

[5]  Leonid I. Goray Numerical analysis for relief gratings working in the soft x ray and XUV region by the integral equation method , 1994, Optics & Photonics.

[6]  Leonid I. Goray,et al.  Comparison of rigorous methods for x-ray and XUV grating diffraction analysis , 1995, Optics & Photonics.

[7]  H Hara,et al.  Design of a high-resolution extreme-ultraviolet imaging spectrometer with aberration-corrected concave gratings. , 1998, Applied optics.

[8]  T. Watanabe,et al.  UV and X-ray spectroscopy of astrophysical and laboratory plasmas : proceedings of the Eleventh Colloquium on UV and X-Ray Spectroscopy of Astrophysical and Laboratory Plasmas held on May 29-June 2, 1995, Nagoya, Japan , 1996 .

[9]  T W Barbee,et al.  Normal-incidence efficiencies in the 115-340-Å wavelength region of replicas of the Skylab 3600-line/mm grating with multilayer and gold coatings. , 1995, Applied optics.

[10]  Grazing-incidence efficiencies in the 28-42-A wavelength region of replicas of the Skylab 3600-line/mm concave grating with multilayer and gold coatings. , 1997, Applied optics.

[11]  W. Hunter,et al.  Comparison of the calculated and the measured efficiencies of a normal-incidence grating in the 125-225- A wavelength range. , 1997, Applied optics.