Performance evaluation of the active-flow personal DataRAM PM2.5 mass monitor (Thermo Anderson pDR-1200) designed for continuous personal exposure measurements

Abstract The need for continuous personal monitoring for exposure to particulate matter has been demonstrated by recent health studies showing effects of PM exposure on time scales of less than a few hours. Filter-based methods cannot measure this short-term variation of PM levels, which can be quite significant considering human activity patterns. The goal of this study was to evaluate the active-flow personal DataRAM for PM2.5 (MIE pDR-1200; Thermo Electron Corp., Franklin, MA) designed as a wearable monitor to continuously measure particle exposure. The instrument precision was found to be good (2.1%) and significantly higher than the passive pDR configuration tested previously. A comparison to other proven continuous monitors resulted in good agreement at low relative humidities. Results at higher humidity followed predictable trends and provided a correction scheme that improved the accuracy of pDR readings. The pDR response to particle size also corresponded to previously observed and theoretical errors. The active flow feature of the pDR allows collection of the sampled particles on a back-up filter. The 24-h mass measured on this filter was found to compare very well with a Federal Reference Method for PM2.5 mass.

[1]  P. Catalano,et al.  Inhaled concentrated ambient particles are associated with hematologic and bronchoalveolar lavage changes in canines. , 2000, Environmental health perspectives.

[2]  H. R. Anderson,et al.  Short-term effects of air pollution on health: a European approach using epidemiological time-series data. The APHEA project: background, objectives, design. , 1995, The European respiratory journal.

[3]  J M Wolfson,et al.  Mechanisms of morbidity and mortality from exposure to ambient air particles. , 2000, Research report.

[4]  J. H. Vincent,et al.  Investigation of the entry characteristics of dust samplers of a type used in the British nuclear industry , 1986 .

[5]  W C Malm,et al.  Aerosol Light Scattering Measurements as a Function of Relative Humidity , 2000, Journal of the Air & Waste Management Association.

[6]  Constantinos Sioutas,et al.  Field evaluation of a modified DataRAM MIE scattering monitor for real-time PM2.5 mass concentration measurements , 2000 .

[7]  W. Bancroft,et al.  The Scattering of Light , 1931 .

[8]  Peter H. McMurry,et al.  Issues in aerosol measurement for optics assessments , 1996 .

[9]  Otto Hänninen,et al.  Fine Particle (PM25) Measurement Methodology, Quality Assurance Procedures, and Pilot Results of the EXPOLIS Study. , 1999, Journal of the Air & Waste Management Association.

[10]  Antonella Zanobetti,et al.  The concentration-response relation between PM(2.5) and daily deaths. , 2002, Environmental health perspectives.

[11]  J D Spengler,et al.  Respiratory health and PM10 pollution. A daily time series analysis. , 1991, The American review of respiratory disease.

[12]  Christine E McLaren,et al.  Association of asthma symptoms with peak particulate air pollution and effect modification by anti-inflammatory medication use. , 2002, Environmental health perspectives.

[13]  P. Koutrakis,et al.  The relationship between airborne acidity and ammonia in indoor environments. , 1994, Journal of exposure analysis and environmental epidemiology.

[14]  R. Michaels Airborne Particle Excursions Contributing to Daily Average Particle Levels may be Managed via a 1 hr Standard, with Possible Public Health Benefits , 1996 .

[15]  T. Larson,et al.  Comparison of light scattering devices and impactors for particulate measurements in indoor, outdoor, and personal environments. , 2002, Environmental science & technology.

[16]  R. Colvile,et al.  Design and validation of a high-flow personal sampler for PM2.5 , 2001, Journal of Exposure Analysis and Environmental Epidemiology.

[17]  J. Ambs,et al.  Development of a Reference Standard for Particulate Matter Mass in Ambient Air , 2001 .

[18]  C. Sioutas,et al.  Development and evaluation of a personal cascade impactor sampler (PCIS) , 2002 .

[19]  M. Kleeman,et al.  Comparison of Real-Time Instruments Used To Monitor Airborne Particulate Matter , 2001, Journal of the Air & Waste Management Association.

[20]  Michael T. Kleinman,et al.  Incidence and Apparent Health Significance of Brief Airborne Particle Excursions , 2000 .

[21]  C. Sloane Optical properties of aerosols of mixed composition , 1984 .

[22]  P. Görner,et al.  Photometer measurement of polydisperse aerosols , 1995 .

[23]  A. Peters,et al.  Particulate air pollution is associated with an acute phase response in men; results from the MONICA-Augsburg Study. , 2001, European heart journal.

[24]  J. Gebhart,et al.  Correlations between gravimetry and light scattering photometry for atmospheric aerosols , 1994 .

[25]  D. Dockery,et al.  Particulate air pollution and daily mortality in Steubenville, Ohio. , 1992, American journal of epidemiology.

[26]  P. Lioy,et al.  Calibration, Intersampler Comparison, and Field Application of a New PM-10 Personal Air-Sampling Impactor , 1991 .

[27]  G. Thurston,et al.  Associations between 1980 U.S. mortality rates and alternative measures of airborne particle concentration. , 1987, Risk analysis : an official publication of the Society for Risk Analysis.

[28]  C A Clayton,et al.  Particle Total Exposure Assessment Methodology (PTEAM) 1990 study: method performance and data quality for personal, indoor, and outdoor monitoring. , 1993, Journal of exposure analysis and environmental epidemiology.

[29]  D. Dockery,et al.  Particulate air pollution as a predictor of mortality in a prospective study of U.S. adults. , 1995, American journal of respiratory and critical care medicine.

[30]  Judith C. Chow,et al.  Sensitivity of estimated light extinction coefficients to model assumptions and measurement errors , 1995 .

[31]  R. Delfino,et al.  Symptoms in pediatric asthmatics and air pollution: differences in effects by symptom severity, anti-inflammatory medication use and particulate averaging time. , 1998, Environmental health perspectives.

[32]  D. Dockery,et al.  An association between air pollution and mortality in six U.S. cities. , 1993, The New England journal of medicine.

[33]  J. Schwartz,et al.  Fine particles are more strongly associated than coarse particles with acute respiratory health effects in schoolchildren. , 2000, Epidemiology.