A neural network-based method for spectral distortion correction in photon counting x-ray CT

Spectral CT using a photon counting x-ray detector (PCXD) shows great potential for measuring material composition based on energy dependent x-ray attenuation. Spectral CT is especially suited for imaging with K-edge contrast agents to address the otherwise limited contrast in soft tissues. We have developed a micro-CT system based on a PCXD. This system enables both 4 energy bins acquisition, as well as full-spectrum mode in which the energy thresholds of the PCXD are swept to sample the full energy spectrum for each detector element and projection angle. Measurements provided by the PCXD, however, are distorted due to undesirable physical effects in the detector and can be very noisy due to photon starvation in narrow energy bins. To address spectral distortions, we propose and demonstrate a novel artificial neural network (ANN)-based spectral distortion correction mechanism, which learns to undo the distortion in spectral CT, resulting in improved material decomposition accuracy. To address noise, post-reconstruction denoising based on bilateral filtration, which jointly enforces intensity gradient sparsity between spectral samples, is used to further improve the robustness of ANN training and material decomposition accuracy. Our ANN-based distortion correction method is calibrated using 3D-printed phantoms and a model of our spectral CT system. To enable realistic simulations and validation of our method, we first modeled the spectral distortions using experimental data acquired from (109)Cd and (133)Ba radioactive sources measured with our PCXD. Next, we trained an ANN to learn the relationship between the distorted spectral CT projections and the ideal, distortion-free projections in a calibration step. This required knowledge of the ground truth, distortion-free spectral CT projections, which were obtained by simulating a spectral CT scan of the digital version of a 3D-printed phantom. Once the training was completed, the trained ANN was used to perform distortion correction on any subsequent scans of the same system with the same parameters. We used joint bilateral filtration to perform noise reduction by jointly enforcing intensity gradient sparsity between the reconstructed images for each energy bin. Following reconstruction and denoising, the CT data was spectrally decomposed using the photoelectric effect, Compton scattering, and a K-edge material (i.e. iodine). The ANN-based distortion correction approach was tested using both simulations and experimental data acquired in phantoms and a mouse with our PCXD-based micro-CT system for 4 bins and full-spectrum acquisition modes. The iodine detectability and decomposition accuracy were assessed using the contrast-to-noise ratio and relative error in iodine concentration estimation metrics in images with and without distortion correction. In simulation, the material decomposition accuracy in the reconstructed data was vastly improved following distortion correction and denoising, with 50% and 20% reductions in material concentration measurement error in full-spectrum and 4 energy bins cases, respectively. Overall, experimental data confirms that full-spectrum mode provides superior results to 4-energy mode when the distortion corrections are applied. The material decomposition accuracy in the reconstructed data was vastly improved following distortion correction and denoising, with as much as a 41% reduction in material concentration measurement error for full-spectrum mode, while also bringing the iodine detectability to 4-6 mg ml(-1). Distortion correction also improved the 4 bins mode data, but to a lesser extent. The results demonstrate the experimental feasibility and potential advantages of ANN-based distortion correction and joint bilateral filtration-based denoising for accurate K-edge imaging with a PCXD. Given the computational efficiency with which the ANN can be applied to projection data, the proposed scheme can be readily integrated into existing CT reconstruction pipelines.

[1]  Wotao Yin,et al.  An Iterative Regularization Method for Total Variation-Based Image Restoration , 2005, Multiscale Model. Simul..

[2]  S M Johnston,et al.  Dual-energy micro-CT of the rodent lung. , 2012, American journal of physiology. Lung cellular and molecular physiology.

[3]  Tuomas Poikela,et al.  Review of hybrid pixel detector readout ASICs for spectroscopic X-ray imaging , 2016 .

[4]  J H Siewerdsen,et al.  Spektr: a computational tool for x-ray spectral analysis and imaging system optimization. , 2004, Medical physics.

[5]  K. Taguchi,et al.  Vision 20/20: Single photon counting x-ray detectors in medical imaging. , 2013, Medical physics.

[6]  David S. Lalush,et al.  Full-Spectrum CT Reconstruction Using a Weighted Least Squares Algorithm With an Energy-Axis Penalty , 2011, IEEE Transactions on Medical Imaging.

[7]  Taly Gilat Schmidt,et al.  Experimental comparison of empirical material decomposition methods for spectral CT , 2015, Physics in medicine and biology.

[8]  Edgar Kraft,et al.  Experimental evaluation of the pile-up trigger method in a revised quantum-counting CT detector , 2012, Medical Imaging.

[9]  A. Rose The sensitivity performance of the human eye on an absolute scale. , 1948, Journal of the Optical Society of America.

[10]  Karl Stierstorfer,et al.  Quantum-counting CT in the regime of count-rate paralysis: introduction of the pile-up trigger method , 2011, Medical Imaging.

[11]  A. Macovski,et al.  Energy-selective reconstructions in X-ray computerised tomography , 1976, Physics in medicine and biology.

[12]  Michael Elad,et al.  On the origin of the bilateral filter and ways to improve it , 2002, IEEE Trans. Image Process..

[13]  G Allan Johnson,et al.  Dual-Energy Computed Tomography Imaging of Atherosclerotic Plaques in a Mouse Model Using a Liposomal-Iodine Nanoparticle Contrast Agent , 2013, Circulation. Cardiovascular imaging.

[14]  Jeffrey R. Ashton,et al.  Dual-Energy Micro-CT Functional Imaging of Primary Lung Cancer in Mice Using Gold and Iodine Nanoparticle Contrast Agents: A Validation Study , 2014, PloS one.

[15]  Katsuyuki Taguchi,et al.  A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT. , 2014, Medical physics.

[16]  E. Nygard,et al.  Photon counting energy dispersive detector arrays for x-ray imaging , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.

[17]  R. Alvarez Estimator for photon counting energy selective x-ray imaging with multibin pulse height analysis. , 2011, Medical physics.

[18]  Michael Campbell,et al.  How spectroscopic x-ray imaging benefits from inter-pixel communication. , 2014, Physics in medicine and biology.

[19]  S. Osher,et al.  Multi-energy CT based on a prior rank, intensity and sparsity model (PRISM) , 2011, Inverse problems.

[20]  Cristian T. Badea,et al.  Simultaneous imaging of multiple contrast agents using full-spectrum micro-CT , 2015, Medical Imaging.

[21]  J. Schlomka,et al.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography , 2008, Physics in medicine and biology.

[22]  L. Hedlund,et al.  A liposomal nanoscale contrast agent for preclinical CT in mice. , 2006, AJR. American journal of roentgenology.

[23]  G Allan Johnson,et al.  Dual-energy micro-computed tomography imaging of radiation-induced vascular changes in primary mouse sarcomas. , 2013, International journal of radiation oncology, biology, physics.

[24]  E. Roessl,et al.  On the influence of noise correlations in measurement data on basis image noise in dual-energylike x-ray imaging. , 2007, Medical physics.

[25]  Mohammad Bagher Menhaj,et al.  Training feedforward networks with the Marquardt algorithm , 1994, IEEE Trans. Neural Networks.

[26]  Michael Campbell,et al.  The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging , 2013 .

[27]  Yi Qi,et al.  Anatomical and functional imaging of myocardial infarction in mice using micro-CT and eXIA 160 contrast agent. , 2014, Contrast media & molecular imaging.

[28]  Woo-Jin Lee,et al.  Material depth reconstruction method of multi-energy X-ray images using neural network , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[29]  C. Badea,et al.  In vivo characterization of tumor vasculature using iodine and gold nanoparticles and dual energy micro-CT , 2013, Physics in medicine and biology.

[30]  Michael Elad,et al.  Fast and robust multiframe super resolution , 2004, IEEE Transactions on Image Processing.

[31]  H Zaidi,et al.  K-edge ratio method for identification of multiple nanoparticulate contrast agents by spectral CT imaging. , 2013, The British journal of radiology.

[32]  Kevin C. Zimmerman,et al.  The effects of extending the spectral information acquired by a photon-counting detector for spectral CT. , 2015, Physics in medicine and biology.

[33]  Axel Thran,et al.  Sensitivity of Photon-Counting Based ${\rm K}$-Edge Imaging in X-ray Computed Tomography , 2011, IEEE Transactions on Medical Imaging.

[34]  Darin P Clark,et al.  Spectral diffusion: an algorithm for robust material decomposition of spectral CT data , 2014, Physics in medicine and biology.