NEMA NU 4-2008 Comparison of Preclinical PET Imaging Systems

The National Electrical Manufacturers Association (NEMA) standard NU 4-2008 for performance measurements of small-animal tomographs was recently published. Before this standard, there were no standard testing procedures for preclinical PET systems, and manufacturers could not provide clear specifications similar to those available for clinical systems under NEMA NU 2-1994 and 2-2001. Consequently, performance evaluation papers used methods that were modified ad hoc from the clinical PET NEMA standard, thus making comparisons between systems difficult. Methods: We acquired NEMA NU 4-2008 performance data for a collection of commercial animal PET systems manufactured since 2000: microPET P4, microPET R4, microPET Focus 120, microPET Focus 220, Inveon, ClearPET, Mosaic HP, Argus (formerly eXplore Vista), VrPET, LabPET 8, and LabPET 12. The data included spatial resolution, counting-rate performance, scatter fraction, sensitivity, and image quality and were acquired using settings for routine PET. Results: The data showed a steady improvement in system performance for newer systems as compared with first-generation systems, with notable improvements in spatial resolution and sensitivity. Conclusion: Variation in system design makes direct comparisons between systems from different vendors difficult. When considering the results from NEMA testing, one must also consider the suitability of the PET system for the specific imaging task at hand.

[1]  Joel S. Karp,et al.  Design evaluation of A-PET: A high sensitivity animal PET camera , 2003 .

[2]  R.A. Mintzer,et al.  Design and performance of a new pixelated- LSO/PSPMT gamma-ray detector for high resolution PET imaging , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.

[3]  R. Mirzoyan,et al.  Studies with a prototype high resolution PET scanner based on LSO-APD modules , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[4]  R. N. Goble,et al.  Performance evaluation of the microPET R4 PET scanner for rodents , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[5]  Martin Gotthardt,et al.  Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner , 2008, Journal of Nuclear Medicine.

[6]  Peter Bruyndonckx,et al.  Performance study of a 3D small animal PET scanner based on BaF2 crystals and a photo sensitive wire chamber , 1997 .

[7]  R. N. Goble,et al.  Performance evaluation of the microPET P4: a PET system dedicated to animal imaging. , 2001, Physics in medicine and biology.

[8]  S. Siegel,et al.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. , 2005, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  T J Spinks,et al.  The design and physical characteristics of a small animal positron emission tomograph. , 1995, Physics in medicine and biology.

[10]  T D Cradduck,et al.  National electrical manufacturers association , 1983, Journal of the A.I.E.E..

[11]  M. Bandstra,et al.  Measurements of Fukushima fallout by the Berkeley Radiological Air and Water Monitoring project , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[12]  S. Cherry,et al.  Observations regarding scatter fraction and NEC measurements for small animal PET , 2006, IEEE Symposium Conference Record Nuclear Science 2004..

[13]  Andrew L. Goertzen,et al.  Variations on the NEMA NU4-2008 testing procedures and effect on the performance measurement results , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[14]  Val J Lowe,et al.  NEMA NU 2-2007 performance measurements of the Siemens Inveon™ preclinical small animal PET system , 2009, Physics in medicine and biology.

[15]  Jurgen Seidel,et al.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[16]  A Rodríguez-Ruano,et al.  Design and performance evaluation of a coplanar multimodality scanner for rodent imaging. , 2009, Physics in medicine and biology.

[17]  D.F. Newport,et al.  Continuously sampled digital pulse processing for inveon small animal PET scanner , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.

[18]  J. S. Lee,et al.  Performance Measurement of the microPET Focus 120 Scanner , 2007, Journal of Nuclear Medicine.

[19]  R. Fontaine,et al.  Performance evaluation of the LabPET12, a large axial FOV APD-based digital PET scanner , 2009, 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC).

[20]  Joel S. Karp,et al.  Imaging performance of a-PET: a small animal PET camera , 2005, IEEE Transactions on Medical Imaging.

[21]  A. Jeavons,et al.  A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals , 1998, 1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255).

[22]  R. Fontaine,et al.  Performance evaluation of the LabPET™ APD-based digital PET scanner , 2009, 2007 IEEE Nuclear Science Symposium Conference Record.

[23]  R. Fontaine,et al.  The Hardware and Signal Processing Architecture of LabPET™, a Small Animal APD-Based Digital PET Scanner , 2009, IEEE Transactions on Nuclear Science.

[24]  James F. Young,et al.  MicroPET: a high resolution PET scanner for imaging small animals , 1996, IEEE Nuclear Science Symposium Conference Record.

[25]  Carole Lartizien,et al.  The ClearPET™ project: development of a 2nd generation high-performance small animal PET scanner , 2005 .

[26]  Qinan Bao,et al.  Performance Evaluation of the Inveon Dedicated PET Preclinical Tomograph Based on the NEMA NU-4 Standards , 2009, Journal of Nuclear Medicine.

[27]  D.F. Newport,et al.  Performance Evaluation of the microPET®—FOCUS-F120 , 2007, IEEE Transactions on Nuclear Science.

[28]  T. Yamashita,et al.  A high resolution PET for animal studies , 1991, Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.

[29]  Roger Lecomte,et al.  Initial results from the Sherbrooke avalanche photodiode positron tomograph , 1996 .

[30]  M E Phelps,et al.  Design features and performance of a PET system for animal research. , 1992, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[31]  Hiroyuki Okada,et al.  A high resolution animal PET scanner using compact PS-PMT detectors , 1996 .