Robot-assisted biopsy using ultrasound guidance: initial results from in vitro tests

The purpose of this study was to develop a robotic system for ultrasound (US)-guided biopsy and to validate the feasibility, accuracy and efficacy using phantom tests. Twenty peas (mean diameter 9.3±0.1 mm) embedded within a gel-phantom were selected for biopsy. Once the best access was defined, the position of the US transducer was recorded by an optical tracking system. Positional data of the transducer and the corresponding US image were transferred to the roboter planning system (LINUX-based industrial PC equipped with video capture card). Once the appropriate position, angulation and pitch were calculated, the robotic arm moved automatically with seven degrees-of-freedom to the planned insertion path, aiming the needle-positioning unit at the center of the target. Then, the biopsy was performed manually using a coaxial technique. The length of all harvested specimens was measured, and the deviation of the actual needle tract from the center of the target was evaluated sonographically. In all targets, the biopsy specimen (mean length 5±1.2 mm) was harvested with only one needle pass required The mean deviation of the needle tip from the center of the target was 1.1±0.8 mm. Robotic assisted biopsies in-vitro using US-guidance were feasible and provided high accuracy.

[1]  Berthold K. P. Horn,et al.  Closed-form solution of absolute orientation using unit quaternions , 1987 .

[2]  A simple phantom for learning needle placement for sonographically guided biopsy. , 1990, AJR. American journal of roentgenology.

[3]  G Bashein,et al.  3D ultrasonic image feature localization based on magnetic scanhead tracking: in vitro calibration and validation. , 1994, Ultrasound in medicine & biology.

[4]  [Value of image guided neurosurgery in neuro-oncology]. , 1995, Bulletin du cancer.

[5]  Jay B. West,et al.  Predicting error in rigid-body point-based registration , 1998, IEEE Transactions on Medical Imaging.

[6]  Y. Matsuoka,et al.  Robotics for surgery. , 1999, Annual review of biomedical engineering.

[7]  H H Chen,et al.  Development of a robotic navigation system for neurosurgery. , 1999, Studies in health technology and informatics.

[8]  W. Kaiser,et al.  Robotic system for biopsy and therapy of breast lesions in a high-field whole-body magnetic resonance tomography unit. , 2000, Investigative radiology.

[9]  T. Krüger,et al.  ROBODOC - Ein Weg in die Zukunft der Hüftendoprothetik oder eine Fehlinvestition? , 2000 .

[10]  H. Reichel,et al.  [ROBODOC--a path into the future of hip endoprosthetics or an investment error?]. , 2000, Zeitschrift fur Orthopadie und ihre Grenzgebiete.

[11]  K. Cleary,et al.  State of the art in surgical robotics: clinical applications and technology challenges. , 2001, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[12]  R Bruce,et al.  CT-integrated robot for interventional procedures: preliminary experiment and computer-human interfaces. , 2001, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[13]  R H Taylor,et al.  System for robotically assisted percutaneous procedures with computed tomography guidance. , 2001, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[14]  Ken Masamune,et al.  System for robotically assisted prostate biopsy and therapy with intraoperative CT guidance. , 2002, Academic radiology.

[15]  Wolfgang Birkfellner,et al.  Evaluation of a miniature electromagnetic position tracker. , 2002, Medical physics.

[16]  Russell H. Taylor,et al.  Robot-Assisted Stapedotomy: Micropick Fenestration of the Stapes Footplate , 2002, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[17]  K. Cleary,et al.  Assessment of hepatic motion secondary to respiration for computer assisted interventions. , 2002, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[18]  A. Dessl,et al.  A novel vacuum device for extremity immobilisation during digital angiography: preliminary clinical experiences , 2002, European Radiology.

[19]  C. Marx,et al.  Interventionelle MR-Mammographie: Manipulatorgestützte Biopsie und interstitielle Lasertherapie von Tumoren der weiblichen Brust , 2003 .

[20]  [Interventional MR-mammography: manipulator-assisted large core biopsy and interstitial laser therapy of tumors of the female breast]]. , 2003, Zeitschrift fur medizinische Physik.

[21]  Philippe Cinquin,et al.  Robot-based Tele-Echography: the TER system , 2003, MIE.

[22]  Anders Nilsson,et al.  Targeted tumour biopsy under contrast-enhanced ultrasound guidance , 2003, European Radiology.

[23]  T. Helbich,et al.  Stereotactic and ultrasound-guided breast biopsy , 2004, European Radiology.

[24]  J. Bodner,et al.  First experiences with the da Vinci operating robot in thoracic surgery. , 2004, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[25]  B. L. Davies,et al.  The use of ultrasound in automated prostatectomy , 2006, Medical and Biological Engineering and Computing.