Characterization of a 2D ion chamber array for the verification of radiotherapy treatments

In this study we investigated the characteristics of a commercial ion chamber array and its performance in the verification of radiotherapy plans. The device was the 2D Array Seven29 model (PTW, Freiburg, Germany). This is a two-dimensional detector array with 729 ionization chambers uniformly arranged in a 27 x 27 matrix with an active area of 27 x 27 cm(2). The detector short-, medium- and long-term reproducibility have been tested through an extensive set of repeated measurements. Short-term reproducibility was well within 0.2%. Medium- and long-term reproducibility were within 1%, including set-up reproducibility errors and linac output fluctuations. Dose linearity was also assessed. The system response to dose was verified to be linear within the range 2-500 MU. Output factors matched very well pinpoint chamber measurements performed in the same experimental conditions with a maximum local percentage difference of 0.4%. Furthermore, the 2D Array sensitivity to millimetric collimator positional changes and to perturbation effect of irradiated area was tested. The comparison with ion chamber data carried out in water was very satisfying. Finally, measurements of wedge-modulated fields and IMRT beam sequence matched very well ion chamber dose profiles acquired in a water tank. The extensive tests performed in this investigation show that the 2D Array Seven29 is a reliable and accurate dosimeter and that it could be a useful tool for the quality assurance and the verification of radiotherapy plans.

[1]  J. Palta,et al.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. , 1994, Medical physics.

[2]  D. Low,et al.  Phantoms for IMRT dose distribution measurement and treatment verification. , 1998, International journal of radiation oncology, biology, physics.

[3]  J. Tsai,et al.  Dosimetric verification of the dynamic intensity-modulated radiation therapy of 92 patients. , 1998, International journal of radiation oncology, biology, physics.

[4]  G. Starkschall,et al.  American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning. , 1998, Medical physics.

[5]  C. De Wagter,et al.  The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapy. , 2000, Physics in medicine and biology.

[6]  C. De Wagter,et al.  The value of the LA48 linear ion chamber array for characterization of intensity-modulated beams. , 2001, Physics in medicine and biology.

[7]  D Huyskens,et al.  Pre-treatment dosimetric verification by means of a liquid-filled electronic portal imaging device during dynamic delivery of intensity modulated treatment fields. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[8]  Michael J. Zelefsky,et al.  High-dose intensity modulated radiation therapy for prostate cancer: early toxicity and biochemical outcome in 772 patients. , 2002, International journal of radiation oncology, biology, physics.

[9]  B G Fallone,et al.  Dosimetric IMRT verification with a flat-panel EPID. , 2003, Medical physics.

[10]  P. Teo,et al.  Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetric advantage over conventional plans and feasibility of dose escalation. , 2003, International journal of radiation oncology, biology, physics.

[11]  Benedick A Fraass,et al.  Potential gains for irradiation of chest wall and regional nodes with intensity modulated radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[12]  B. Poppe,et al.  [Use of a two-dimensional ionization chamber array for quality assurance in medical linear accelerators]. , 2003, Zeitschrift fur medizinische Physik.

[13]  Ralf Kollhoff,et al.  Einsatz eines zweidimensionalen Ionisationskammer-Arrays zur Qualitätssicherung an medizinischen Linearbeschleunigern , 2003 .

[14]  P. Jursinic,et al.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. , 2003, Medical physics.

[15]  I. Rosen,et al.  Implementing IMRT in clinical practice: a joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in Medicine. , 2004, International journal of radiation oncology, biology, physics.

[16]  M Stasi,et al.  The behavior of several microionization chambers in small intensity modulated radiotherapy fields. , 2004, Medical physics.

[17]  R Tielenburg,et al.  The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes. , 2004, Medical physics.

[18]  Volker Budach,et al.  Clinical and physical quality assurance for intensity modulated radiotherapy of prostate cancer. , 2004, Radiotherapy and Oncology.

[19]  M. Oldham,et al.  Evaluation of a 2D diode array for IMRT quality assurance. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[20]  R. Cirio,et al.  Dosimetric characterization of a large area pixel-segmented ionization chamber. , 2004, Medical physics.

[21]  J J Sonke,et al.  Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device. , 2003, Medical physics.

[22]  Tom Depuydt,et al.  The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[23]  P W Chin,et al.  Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle. , 2004, Physics in medicine and biology.